| DICOM PS3.15 2022a - Security and System Management Profiles |
|---|
This message describes the event of exporting data from a system, meaning that the data is leaving control of the system's security domain. Examples of exporting include printing to paper, recording on film, conversion to another format for storage in an EHR, writing to removable media, or sending via e-mail. Multiple patients may be described in one event message.
Table A.5.3.4-1. Audit Message for Data Export
|
The identity of the remote user or process receiving the data |
|||
|
The identity of the local user or process exporting the data. If both are known, then two active participants shall be included (both the person and the process). |
|||
|
See Section A.5.2.3 |
|||
|
See Section A.5.2.4 |
|||
|
Required if being exported to other than physical media, e.g., to a network destination rather than to film, paper or CD. May be present otherwise. |
|||
|
Required if Net Access Point Type Code is present. May be present otherwise. |
|||
|
Values selected from DCID 405 “Media Type Code” |
|||
|
See Table A.5.2-1 |
|||
| DICOM PS3.15 2022a - Security and System Management Profiles |
|---|