| DICOM PS3.7 2022a - Message Exchange |
|---|
The information necessary for the C-MOVE request and indication DIMSE-C primitives are conveyed in the C-MOVE-RQ Message. The information necessary for the C-MOVE response and confirmation DIMSE-C primitives are conveyed in the C-MOVE-RSP Message. The information necessary for the C-MOVE Cancel request and Cancel indication primitives are conveyed in the C-CANCEL-MOVE-RQ Message.
The C-MOVE-RQ Message contains fields as defined in Table 9.3-9. Each field shall conform to DICOM encoding and Value Representation as defined in PS3.5. Fields are required as specified in the C-MOVE service definition unless otherwise noted in Table 9.3-9. Fields not specified in the C-GET-MOVE service definition but present in Table 9.3-9 are required by the DIMSE-C protocol.
Table 9.3-9. C-MOVE-RQ Message Fields
|
The even number of bytes from the end of the value field to the beginning of the next group. |
||||
|
This field distinguishes the DIMSE-C operation conveyed by this Message. The value of this field shall be set to 0021H for the C-MOVE-RQ Message. |
||||
|
Implementation-specific value that distinguishes this Message from other Messages. |
||||
|
This field indicates that a Data Set is present in the Message. It shall be set to any value other than 0101H (Null). |
||||
|
Shall be set to the DICOM AE Title of the destination DICOM AE to which the C-STORE sub-operations are being performed. |
||||
|
A Data Set that encodes the Identifier to be matched. See Section 9.1.4.1.6. |
| DICOM PS3.7 2022a - Message Exchange |
|---|