| DICOM PS3.7 2022a - Message Exchange |
|---|
The N-SET-RSP Message contains all fields as defined in Table 10.3-6 and in Annex C. Each field shall conform to DICOM encoding and Value Representation as defined in PS3.5. Fields are required as specified in the N-SET service definition unless otherwise noted. Fields not specified in the N-SET service definition but present in Table 10.3-6 are required by the DIMSE-N protocol.
Table 10.3-6. N-SET-RSP Message Fields
|
The even number of bytes from the end of the value field to the beginning of the next group. |
||||
|
SOP Class UID of the SOP Instance for which Attribute Values were modified. |
||||
|
This field distinguishes the DIMSE-N operation conveyed by this Message. The value of this field shall be set to 8120H for the N-SET-RSP Message. |
||||
|
Shall be set to the value of the Message ID (0000,0110) field used in associated N-SET-RQ Message. |
||||
|
This field indicates if a Data Set is present in the Message. This field shall be set to the value of 0101H if no Data Set is present; any other value indicates a Data Set is included in the Message. |
||||
|
The value of this field depends upon the status type. Annex C defines the encoding of the status types defined in the service definition. |
||||
|
Contains the UID of the SOP Instance for which Attribute Values were modified. |
||||
|
This field is encoded as a Data Set. One Data Element is encoded for each Attribute and Attribute Value applicable to the operation. |
| DICOM PS3.7 2022a - Message Exchange |
|---|