DICOM PS3.4 2022a - Service Class Specifications

EE.3 Conformance

EE.3.1 Conformance Statement

The implementation conformance statement of this SOP Class shall follow PS3.2.

The SCU Conformance Statement shall specify the following items:

  • Maximum number of associations to be supported at the same time

  • List of SOP Classes supported

  • For each of the supported SOP Classes:

    • List of supported SOP Class Attributes and DICOM Message Service Elements

    • For each supported Attribute (mandatory and optional), a valid value range

The SCP Conformance Statement shall specify the following items:

  • Maximum number of associations to be supported at the same time

  • List of SOP Classes supported

  • For each of the supported SOP Classes:

    • List of supported SOP Class Attributes and DICOM Message Service Elements

    • For each supported Attribute (mandatory and optional)

      • Valid value range

      • Default value if no value is supplied by the SCU

      • Status code (Failure or Warning) if the SCU supplies a value that is out of range

  • For each supported DIMSE service

    • SCP behavior for all specific status codes

DICOM PS3.4 2022a - Service Class Specifications