| DICOM PS3.3 2022a - Information Object Definitions |
|---|
The RT Brachy Application Setup Delivery Instruction Module contains additional information required by a Treatment Delivery System (TDS) when specifying delivery of a brachytherapy treatment.
This Module references the RT Plan to be delivered and specifies the treatment tasks to be performed for the Application Setups of the RT Plan.
The RT Brachy Application Setup Delivery Instruction Module consists of a Sequence of one or more tasks.
Table C.8.8.30-1. RT Brachy Application Setup Delivery Instruction Module Attributes
|
Reference to a single RT Plan SOP Instance containing all the brachy application setups and the fraction group referenced in this SOP Instance. |
|||
|
>Include Table C.17-3 “Hierarchical SOP Instance Reference Macro Attributes” |
|||
|
Indicates which fraction group of the referenced RT Plan in the Referenced RT Plan Sequence (300C,0002) is to be treated in the treatment session. |
|||
|
The index of the fraction that is to be delivered or completed in this treatment session. See Section C.8.8.30.3. |
|||
|
The index of the pulse that is to be delivered or completed in this treatment session. Required if Brachy Treatment Type (300A,0202) of the referenced RT Plan in the Referenced RT Plan Sequence (300C,0002) is PDR and if Treatment Delivery Type (300A,00CE) is CONTINUATION in any of the Items in the Brachy Task Sequence (0074,1401). See Section C.8.8.30.4. |
|||
|
Specification of Brachy Application Setups to be delivered. One or more Items shall be included in this Sequence. See Section C.8.8.30.1. |
|||
|
Enumerated Values: See Section C.8.8.30.2. |
|||
|
Total Reference Air Kerma within the Application Setup referenced by Referenced Brachy Application Setup Number (300C,000C) at which treatment delivery starts. |
|||
|
Total Reference Air Kerma within the Application Setup referenced by Referenced Brachy Application Setup Number (300C,000C) at which treatment delivery ends. |
|||
|
Brachy Application Setup Number (300A,0234) within Brachy Application Setup Sequence (300A,0230) in the referenced RT Plan. |
|||
|
The channel to be delivered, specified by the value of Channel Number (300A,0282) in referenced RT Plan. |
|||
|
The order of delivery of this channel, monotonically increasing by 1, starting from 1, where number 1 is the first channel to be delivered. |
|||
|
The treatment time weights for channels to be delivered to complete a previously interrupted fraction. |
|||
|
The channel to be delivered, specified by the value of Channel Number (300A,0282) in referenced RT Plan. |
|||
|
Cumulative Time Weight where to start/continue the delivery of the referenced channel. See Section C.8.8.15.6. |
|||
|
Cumulative Time Weight where to stop the delivery of the referenced channel. See Section C.8.8.15.6. |
|||
|
Application Setups of the current RT Plan containing channels not to be delivered. Required if Delivery Type (300A,00CE) is CONTINUATION and one or more channels of any Application Setup are omitted. |
|||
|
Brachy Application Setup Number (300A,0234) within Brachy Application Setup Sequence (300A,0230) in referenced RT Plan. |
|||
|
Channels defined for this Application Setup not to be delivered. |
|||
|
Channel to be omitted, specified by the value of Channel Number (300A,0282) in the referenced RT Plan. |
|||
The Brachy Task Sequence (0074,1401) contains instructions for the delivery of a single treatment session. A treatment session is the time that the patient is connected to the treatment delivery system. During a treatment session, one fraction of the referenced plan is delivered. If the value of Brachy Treatment Type (300A,0202) in the reference RT Plan is HDR, the entire fraction is delivered at one time. If the value is PDR, the fraction is divided into a number of pulses which are delivered periodically during the treatment session.
The Application Setup groups together the channels to be delivered. The task Sequence can also contain the delivery order of the channels. The Omitted Application Setup Sequence (0074,140E) contains the channels to skip in the case of continuation of an interrupted treatment.
The following scenarios demonstrate the usage of the RT Brachy Application Setup Delivery Instruction Module for the Treatment Management System (TMS) to instruct the Treatment Delivery System (TDS) to initiate brachytherapy treatment and continue an interrupted delivery.
It has been decided to proceed to fraction 2 to deliver the remainder of the interrupted plan as instructed by the following task transmitted inside the delivery instruction:
The remainder is defined as to skip 25 sec of the first dwell position of channel 2 and continue with the second dwell position.
| DICOM PS3.3 2022a - Information Object Definitions |
|---|