| DICOM PS3.3 2022a - Information Object Definitions |
|---|
This section describes the Breast Tomosynthesis Acquisition Module.
Table C.8.21.3.4-1. Breast Tomosynthesis Acquisition Module Attributes
|
Each Item represents an acquisition context related to one or more reconstructions. The values of Acquisition Index (0020,9518) may be used as index in this Sequence. |
|||
|
>Include Table C.8.21.3.1.1-1 “X-Ray 3D General Shared Acquisition Macro Attributes” |
|||
|
>Include Table C.8.21.3.1.3-1 “X-Ray 3D General Positioner Movement Macro Attributes” |
|||
|
Distance in mm from source to detector center on the chest wall line. Note
|
|||
|
Distance in mm from source to the breast support side that is closest to the Imaging Subject, as measured along the X-Ray beam vector. Note
|
|||
|
Ratio of Source Image Receptor Distance (SID) over Source Object Distance (SOD). |
|||
|
The average thickness in mm of the body part examined when compressed, if compression has been applied during exposure. |
|||
|
Text description of the mechanism of exposure control. May describe the number and type of exposure sensors or position of the sensitive area of the imaging detector. |
|||
|
The thickness of Aluminum in mm required to reduce the X-Ray Output (0040,0312) by a factor of two. |
|||
|
Organ dose value measured in dGy representing the collective total for all acquired frames described in this Sequence Item. |
|||
|
Entrance dose value measured in mGy at the surface of the patient representing the collective total for all acquired frames described in this Sequence Item. |
|||
|
Describes what type of dose is represented by the values of Entrance Dose in mGy (0040,8302). Enumerated Values:
Only meaningful if Entrance Dose in mGy (0040,8302) is present. |
|||
|
Nominal focal spot size in mm used to acquire the projection images. |
|||
|
Number of active detectors used to generate a single pixel. Specified as number of row detectors per pixel then column. Required if detector binning was applied to the projection images. |
|||
|
Type of filter(s) inserted into the X-Ray beam (e.g., wedges). See Attribute Description in Section C.8.7.10 for Defined Terms. |
|||
|
The X-Ray absorbing material used in the filter. May be multi-valued. See Attribute Description in Section C.8.7.10 for Defined Terms. |
|||
|
See Attribute Description in Section C.8.7.10. |
|||
|
See Attribute Description in Section C.8.7.10. |
|||
|
See Attribute Description in Section C.8.7.10. |
|||
|
See Attribute Description in Section C.8.7.10. |
|||
|
The compression force applied to the body part during exposure, measured in Newtons. |
|||
|
The average compression pressure applied to the body part during exposure, calculated in kPa. |
|||
|
The area of the body part to which compression has been applied during exposure, measured in mm2. |
|||
|
Description of the compression paddle, if compression was applied to the body part during exposure. |
|||
|
Sequence containing detailed acquisition context of each individual projection used in this acquisition context. |
|||
|
>>Include Table C.8.21.3.1.2-1 “X-Ray 3D General Per Projection Acquisition Macro Attributes” |
|||
|
Signed position in degrees of the X-Ray beam vector in the coronal anatomical plane as if the patient were standing facing the equipment where vertical is zero. |
|||
|
The interpretation of the sign of the Positioner Primary Angle (0018,1510). Enumerated Values:
See Section C.8.31.2.1. |
|||
|
Position in degrees of the X-Ray beam vector in the sagittal anatomical plane as if the patient were standing where movement of the X-Ray source from anterior to posterior is positive, and vertical is zero. Required if secondary positioner was used during acquisition. |
|||
|
Duration of X-Ray exposure in milliseconds. See Section C.8.7.2.1.1. |
|||
|
The exposure expressed in milliampereseconds, for example calculated from Exposure Time and X-Ray Tube Current. |
|||
|
Indication of the applied dose, in manufacturer specific units. Note
|
|||
|
Average entrance dose value measured in mGy at the surface of the patient during the acquisition of this projection image. |
|||
|
Describes what type of dose is represented by the values of Entrance Dose in mGy (0040,8302). Enumerated Values:
Only meaningful if Entrance Dose in mGy (0040,8302) is present. |
|||
|
Unique identification of the irradiation event(s) associated with the acquisition of this image. |
|||
| DICOM PS3.3 2022a - Information Object Definitions |
|---|