| DICOM PS3.3 2022a - Information Object Definitions |
|---|
The RT Second Generation IODs use the General Series Module described in Section C.7.3.1, specialized by the Enhanced RT Series Module.
Table C.36.3-1 specifies the Attributes that identify and describe general information about the Enhanced RT Series.
Table C.36.3-1. Enhanced RT Series Module Attributes
|
Type of device, process or method that originally acquired the data used to create the Instances in this Series. See Section C.36.3.1.1. |
|||
|
Uniquely identifies the Performed Procedure Step SOP Instance that resulted in creation of the Series (e.g. a Modality or Unified Procedure Step SOP Instance). Only a single Item shall be included in this Sequence. Required if the series has been created as a result of a single procedure step request and the Instance-Level Referenced Performed Procedure Step Sequence (3010,0044) is not present. If different Instances in the series are created as a result of a procedure step, the Instance-Level Referenced Performed Procedure Step Sequence (3010,0044) in the Radiotherapy Common Instance Module shall be used. |
|||
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
| DICOM PS3.3 2022a - Information Object Definitions |
|---|