| DICOM PS3.3 2022a - Information Object Definitions |
|---|
This Module records the latest results of QA activities, such as calibration, performed on the Display Subsystems.
Table C.32.3-1. QA Results Module Attributes
|
Exactly one Item shall be included for each Display Subsystem present in the Display System. |
|||
|
The Display Subsystem ID of the Item in the Display Subsystem Sequence (0028,7023) to which these results apply. |
|||
|
Results of QA activities for the identified Display Subsystem. |
|||
|
The Configuration ID of the Item in the Display Subsystem Configuration Sequence (0028,700A) under which the QA was performed. A given Configuration ID value shall not appear more than once in this Sequence. |
|||
|
Results of QA activities for the identified Display Subsystem in the identified configuration |
|||
|
>>>>Include Table C.32.3.1-1 “Display QA Result Context Macro Attributes” |
|||
|
The Luminance Characteristics ID of the Item in the Target Luminance Characteristics Sequence (0028,7008) that represents the target values used during this calibration. |
|||
|
>>>>Include Table C.32.3.1-1 “Display QA Result Context Macro Attributes” |
|||
|
>>>>Include Table C.32.3.2-1 “Visual Evaluation Result Macro Attributes” |
|||
|
>>>>Include Table C.32.3.1-1 “Display QA Result Context Macro Attributes” |
|||
|
>>>>Include Table C.32.3.3-1 “Luminance Uniformity Result Macro Attributes” |
|||
|
>>>>Include Table C.32.3.1-1 “Display QA Result Context Macro Attributes” |
|||
|
>>>>Include Table C.32.3.4-1 “Luminance Result Macro Attributes” |
|||
Several Result Sequences in the above table are permitted to contain zero Items. This means that no valid "most recent result" exists. This may be because such results have never been generated, or because the system considers that the most recent result is no longer valid (perhaps due to age or intervening equipment changes). Specifying such policy is outside the scope of DICOM.
This Macro records the context for a set of collected display QA results: when they were produced, the person/device that produced them.
Table C.32.3.1-1. Display QA Result Context Macro Attributes
|
Date and Time at which the QA job that generated the results started. |
||
|
Date and Time at which the QA job that generated the results ended. |
||
|
Organization to which the human performer is accountable for the results. |
||
|
Describes the measurement equipment (if any) used in deriving the results. This may describe standalone equipment or may re-iterate details of equipment built into the Display System. |
||
|
Measurement functions of the measurement equipment. Enumerated Values: This multi-valued Attribute shall contain at most one of each enumerated value. When multiple values are present, the equipment is capable of performing multiple functions. |
||
|
Characteristic(s) which are measured by the measurement equipment. Enumerated Values: This multi-valued Attribute shall contain at most one of each enumerated value. When multiple values are present, the equipment is capable of measuring multiple characteristics. |
||
|
Type of measurement equipment. See Section C.32.1.2 for Enumerated Values. |
||
|
Date and time of the last calibration of the measurement equipment. |
||
| DICOM PS3.3 2022a - Information Object Definitions |
|---|