| DICOM PS3.3 2022a - Information Object Definitions |
|---|
Table C.31-1 specifies the Attributes used to convey the parameters used in external verification of both conventional radiotherapy (photon or electron) and ion treatment deliveries.
Table C.31-1. RT General Machine Verification Module Attributes
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
||
|
Conditions under which treatment was verified by a verification system. |
||
|
List of parameters that failed verification by the MPV. Zero or more Items shall be included in this Sequence. See Section C.31.1.1. |
||
|
List of parameters that were overridden by the user. Zero or more Items shall be included in this Sequence. See Section C.31.1.1. |
||
|
Name of operator who authorized override of overridden parameter. |
||
|
Identification of the operator who authorized override. Only a single Item is permitted in this Sequence. |
||
|
>>Include Table 10-1 “Person Identification Macro Attributes” |
||
|
User-defined description of reason for override of overridden parameter. |
||
|
Sequence containing general machine verification parameters. |
||
|
Desired machine setting of primary Meterset. The units shall match those specified by Primary Dosimeter Unit (300A,00B3) in the plan specified in the Referenced RT Plan Sequence (300C,0002). |
||
|
Desired machine setting of secondary Meterset. The units shall match those specified for Specified Secondary Meterset (3008,0033) in the plan specified in the Referenced RT Plan Sequence (300C,0002). |
||
|
Beam limiting device (collimator) jaw or leaf (element) leaf pair values. |
||
|
Type of beam limiting device (collimator). Enumerated Values: |
||
|
Number of leaf (element) or jaw pairs (equal to 1 for standard beam limiting device jaws). |
||
|
Identification number of the Wedge. The value of Wedge Number (300A,00D2) shall be unique within the Wedge Sequence. |
||
|
Orientation of wedge, i.e., orientation of the IEC WEDGE FILTER coordinate system with respect to the IEC BEAM LIMITING DEVICE coordinate system, in degrees. |
||
|
An identifier for the accessory intended to be read by a device such as a bar-code reader. |
||
|
An identifier for the accessory intended to be read by a device such as a bar-code reader. |
||
|
Uniquely identifies compensator specified by Compensator Number (300A,00E4) within Beam referenced by Referenced Beam Number (300C,0006). |
||
|
An identifier for the accessory intended to be read by a device such as a bar-code reader. |
||
|
Uniquely identifies block specified by Block Number (300A,00FC) within Beam referenced by Referenced Beam Number (300C,0006). |
||
|
User-defined name identifying treatment machine to be used for beam delivery. See Section C.8.8.25.2. |
||
|
Number of shielding blocks or Electron Inserts associated with Beam. |
||
|
An identifier for the accessory intended to be read by a device such as a bar-code reader. |
||
|
Enumerated Values:
|
||
|
An identifier for the accessory intended to be read by a device such as a bar-code reader. |
||
|
References Beam specified by Beam Number (300A,00C0) in Beam Sequence (300A,00B0) in RT Beams Module of referenced RT Plan or in Ion Beam Sequence (300A,03A2) in RT Ion Beams Module within the referenced RT Ion Plan. |
||
|
Uniquely identifies ROI representing the Bolus specified by ROI Number (3006,0022) in Structure Set ROI Sequence (3006,0020) in Structure Set Module within RT Structure Set in Referenced Structure Set Sequence (300C,0060) in RT General Plan Module. |
||
|
An accessory identifier to be read by a device such as a bar code reader. |
||
Each Verify Failed Parameter or Overridden Parameter is used to specify a verification Attribute that was either unverified (out of range), or overridden explicitly by the user. To uniquely identify both the context and value item of the Attribute occurrence, a special notation is used, described in Section 10.17.
If a Verify Failed Parameter or Overridden Parameter refers to an Attribute occurrence within a Control Point Sequence, then that Attribute occurrence was failed or overridden during the administration of the beam segment immediately prior to the specified control point.
| DICOM PS3.3 2022a - Information Object Definitions |
|---|