DICOM PS3.3 2022a - Information Object Definitions

C.23.4.2 Hanging Protocol Selector Attribute Value Macro

Table C.23.4-2 specifies the Attributes that identify the value(s) for a Data Element Tag that is used as a Selector Attribute (0072,0026) in the Image Set Selector Sequence (0072,0022) or Filter Operations Sequence (0072,0400).

Table C.23.4-2. Hanging Protocol Selector Attribute Value Macro Attributes

Attribute Name

Tag

Type

Attribute Description

Selector AT Value

(0072,0060)

1C

The value(s) of the Attribute identified by Selector Attribute (0072,0026).

Required if Selector Attribute VR (0072,0050) is present and the value is AT.

Selector CS Value

(0072,0062)

1C

The value(s) of the Attribute identified by Selector Attribute (0072,0026).

Required if Selector Attribute VR (0072,0050) is present and the value is CS.

Selector IS Value

(0072,0064)

1C

The value(s) of the Attribute identified by Selector Attribute (0072,0026).

Required if Selector Attribute VR (0072,0050) is present and the value is IS. Some leniency in precision and format (including padding with spaces and leading zeros, e.g., "001" = " 1 " = "1") will be required.

Selector LO Value

(0072,0066)

1C

The value(s) of the Attribute identified by Selector Attribute (0072,0026).

Required if Selector Attribute VR (0072,0050) is present and the value is LO.

Selector LT Value

(0072,0068)

1C

The value of the Attribute identified by Selector Attribute (0072,0026).

Required if Selector Attribute VR (0072,0050) is present and the value is LT.

Selector PN Value

(0072,006A)

1C

The value(s) of the Attribute identified by Selector Attribute (0072,0026).

Required if Selector Attribute VR (0072,0050) is present and the value is PN.

Selector SH Value

(0072,006C)

1C

The value(s) of the Attribute identified by Selector Attribute (0072,0026).

Required if Selector Attribute VR (0072,0050) is present and the value is SH.

Selector ST Value

(0072,006E)

1C

The value of the Attribute identified by Selector Attribute (0072,0026).

Required if Selector Attribute VR (0072,0050) is present and the value is ST.

Selector UT Value

(0072,0070)

1C

The value of the Attribute identified by Selector Attribute (0072,0026).

Required if Selector Attribute VR (0072,0050) is present and the value is UT.

Selector DS Value

(0072,0072)

1C

The value(s) of the Attribute identified by Selector Attribute (0072,0026).

Required if Selector Attribute VR (0072,0050) is present and the value is DS. Some leniency in precision and format (including padding and scientific notation) will be required.

Selector FD Value

(0072,0074)

1C

The value(s) of the Attribute identified by Selector Attribute (0072,0026).

Required if Selector Attribute VR (0072,0050) is present and the value is FD.

Selector FL Value

(0072,0076)

1C

The value(s) of the Attribute identified by Selector Attribute (0072,0026).

Required if Selector Attribute VR (0072,0050) is present and the value is FL.

Selector UL Value

(0072,0078)

1C

The value(s) of the Attribute identified by Selector Attribute (0072,0026).

Required if Selector Attribute VR (0072,0050) is present and the value is UL.

Selector US Value

(0072,007A)

1C

The value(s) of the Attribute identified by Selector Attribute (0072,0026).

Required if Selector Attribute VR (0072,0050) is present and the value is US.

Selector SL Value

(0072,007C)

1C

The value(s) of the Attribute identified by Selector Attribute (0072,0026).

Required if Selector Attribute VR (0072,0050) is present and the value is SL.

Selector SS Value

(0072,007E)

1C

The value(s) of the Attribute identified by Selector Attribute (0072,0026).

Required if Selector Attribute VR (0072,0050) is present and the value is SS.

Selector Code Sequence Value

(0072,0080)

1C

The value(s) of the Attribute identified by Selector Attribute (0072,0026). One or more Items shall be included in this Sequence. See Section C.23.4.2.1.2.

Required if Selector Attribute VR (0072,0050) is present and the value is SQ, and Selector Attribute (0072,0026) is a Code Sequence.

>Include Table 8.8-1 “Code Sequence Macro Attributes”

No Baseline CID is defined.

Selector UI Value

(0072,007F)

1C

The value of the Attribute identified by Selector Attribute (0072,0026).

Required if Selector Attribute VR (0072,0050) is present and the value is UI.


C.23.4.2.1 Hanging Protocol Selector Attribute Value Macro Attribute Descriptions
C.23.4.2.1.1 Selector Attribute Value Matching

The value of Specific Character Set (0008,0005) in the Image and the Hanging Protocol Instance may differ and shall be taken into account for matching.

The value of Specific Character Set (0008,0005) may influence how matching of text Attributes is performed, in an implementation dependent manner. No requirements are specified for case sensitive or accent sensitive matching, or for ignoring padding.

C.23.4.2.1.2 Selector Code Sequence Value

The matching shall be performed on Coding Scheme Designator (0008,0102) and Code Value (0008,0100) (or Long Code Value (0008,0119) or URN Code Value (0008,0120), if present). Code Meaning (0008,0104) is required to be present, but shall be ignored for matching purposes. The matching is case sensitive, and leading and trailing spaces are not significant. The Coding Scheme Version (0008,0103) shall be ignored unless Coding Scheme Designator (0008,0102) is not sufficient to identify Code Value (0008,0100) unambiguously.

DICOM PS3.3 2022a - Information Object Definitions