DICOM PS3.3 2022a - Information Object Definitions

C.13 Print Management Specific Modules

The following Sections specify Modules used for Print Management.

C.13.1 Basic Film Session Presentation Module

Table C.13-1. Basic Film Session Presentation Module Attributes

Attribute name

Tag

Attribute Description

Number of Copies

(2000,0010)

Number of copies to be printed for each film of the film session.

Print Priority

(2000,0020)

Specifies the priority of the print job.

Enumerated Values:

HIGH

MED

LOW

Medium Type

(2000,0030)

Type of medium on which the print job will be printed.

Defined Terms:

PAPER

CLEAR FILM

BLUE FILM

MAMMO CLEAR FILM

MAMMO BLUE FILM

Film Destination

(2000,0040)

Film destination.

Defined Terms:

MAGAZINE

the exposed film is stored in film magazine

PROCESSOR

the exposed film is developed in film processor

BIN_i

the exposed film is deposited in a sorter bin where "I" represents the bin number. Film sorter BINs shall be numbered sequentially starting from 1 and no maximum is placed on the number of BINs. The encoding of the BIN number shall not contain leading zeros.

Film Session Label

(2000,0050)

Human readable label that identifies the film session

Memory Allocation

(2000,0060)

Amount of memory allocated for the film session. Value is expressed in KB

Owner ID

(2100,0160)

Identification of the owner of the film session


DICOM PS3.3 2022a - Information Object Definitions