DICOM PS3.3 2022a - Information Object Definitions

A.74.3 Breast Projection X-Ray Image IOD Module Table

Table A.74-1 specifies the Modules of the Breast Projection X-Ray Image IOD.

Table A.74-1. Breast Projection X-Ray Image IOD Modules

IE

Module

Reference

Usage

Patient

Patient

C.7.1.1

M

Clinical Trial Subject

C.7.1.3

U

Study

General Study

C.7.2.1

M

Patient Study

C.7.2.2

U

Clinical Trial Study

C.7.2.3

U

Series

General Series

C.7.3.1

M

Clinical Trial Series

C.7.3.2

U

DX Series

C.8.11.1

M

Enhanced Mammography Series

C.8.11.10

M

Frame of Reference

Frame of Reference

C.7.4.1

M

Synchronization

C.7.4.2

C - Required if time synchronization was applied.

Equipment

General Equipment

C.7.5.1

M

Enhanced General Equipment

C.7.5.2

M

Image

Enhanced Mammography Image

C.8.31.1

M

Breast View

C.8.21.6

M

Image Pixel

C.7.6.3

M

Enhanced Contrast/Bolus

C.7.6.4b

C - Required if contrast media was applied.

Device

C.7.6.12

U

Intervention

C.7.6.13

U

Acquisition Context

C.7.6.14

M

Multi-frame Functional Groups

C.7.6.16

M

Multi-frame Dimension

C.7.6.17

U

Patient Orientation

C.7.6.20

M

Specimen

C.7.6.22

U

SOP Common

C.12.1

M

Common Instance Reference

C.12.2

U

Frame Extraction

C.12.3

C - Required if the SOP Instance was created in response to a Frame-Level retrieve request


A.74.3.1 Breast Projection X-Ray Image IOD Content Constraints

A.74.3.1.1 Modality Attribute

Modality (0008,0060) shall have the value MG.

A.74.3.1.2 Overlay Plane Module, Curve Module and VOI LUT Module

The Overlay Plane Module, Curve Module (Retired), VOI LUT Module and Softcopy Presentation LUT Module shall not be used in a Standard Extended SOP Class of the Breast Projection X-Ray Image.

Note

The VOI LUT function is provided by a Frame VOI LUT Functional Group.

DICOM PS3.3 2022a - Information Object Definitions