DICOM PS3.3 2022a - Information Object Definitions

A.7.4 US Multi-frame Image IOD Module Table

Table A.7-1 specifies the Modules of the US Multi-frame Image IOD.

Table A.7-1. US Multi-frame Image IOD Modules

IE

Module

Reference

Usage

Patient

Patient

C.7.1.1

M

Clinical Trial Subject

C.7.1.3

U

Study

General Study

C.7.2.1

M

Patient Study

C.7.2.2

U

Clinical Trial Study

C.7.2.3

U

Series

General Series

C.7.3.1

M

Clinical Trial Series

C.7.3.2

U

Frame of Reference

Frame of Reference

C.7.4.1

U

Synchronization

C.7.4.2

C - Required if Modality (0008,0060) = IVUS.

May be present otherwise.

Equipment

General Equipment

C.7.5.1

M

Image

General Image

C.7.6.1

M

General Reference

C.12.4

U

Image Pixel

C.7.6.3

M

Contrast/Bolus

C.7.6.4

C - Required if contrast media was used in this image.

Cine

C.7.6.5

M

Multi-frame

C.7.6.6

M

Frame Pointers

C.7.6.9

U

Palette Color Lookup Table

C.7.9

C - Required if Photometric Interpretation (0028,0004) has a value of PALETTE COLOR

Device

C.7.6.12

U

Specimen

C.7.6.22

U

US Region Calibration

C.8.5.5

U

US Image

C.8.5.6

M

Overlay Plane

C.9.2

U

Multi-frame Overlay

C.9.3

U

VOI LUT

C.11.2

U

ICC Profile

C.11.15

U

SOP Common

C.12.1

M

Common Instance Reference

C.12.2

U

Frame Extraction

C.12.3

C - Required if the SOP Instance was created in response to a Frame-Level retrieve request


Note

The US Frame of Reference Module was previously included in this IOD, but has been retired. See PS3.3-2003.

A.7.4.1 Mutually Exclusive IEs

Note

A Curve IE was previously included in this IOD that was mutually exclusive with the Image IE, but has been retired. See PS3.3-2004.

DICOM PS3.3 2022a - Information Object Definitions