DICOM PS3.3 2022a - Information Object Definitions

A.66.3 Intravascular Optical Coherence Tomography Image IOD Module Table

Table A.66.3-1 specifies the Modules of the Intravascular Optical Coherence Tomography Image IOD.

Table A.66.3-1. Intravascular Optical Coherence Tomography Image IOD Modules

IE

Module

Reference

Usage

Patient

Patient

C.7.1.1

M

Clinical Trial Subject

C.7.1.3

U

Study

General Study

C.7.2.1

M

Patient Study

C.7.2.2

U

Clinical Trial Study

C.7.2.3

U

Series

General Series

C.7.3.1

M

Clinical Trial Series

C.7.3.2

U

Intravascular OCT Series

C.8.27.1

M

Frame of Reference

Frame of Reference

C.7.4.1

M

Synchronization

C.7.4.2

M

Equipment

General Equipment

C.7.5.1

M

Enhanced General Equipment

C.7.5.2

M

Image

Image Pixel

C.7.6.3

M

Supplemental Palette Color Lookup Table

C.7.6.19

C - Required if Pixel Presentation (0008,9205) in the Intravascular OCT Image Module equals COLOR.

Enhanced Contrast/Bolus

C.7.6.4

M

Multi-frame Functional Groups

C.7.6.16

M

Multi-frame Dimension

C.7.6.17

M

Device

C.7.6.12

U

Acquisition Context

C.7.6.14

M

Cardiac Synchronization

C.7.6.18.1

C - Required if cardiac synchronization was used

Intravascular OCT Image

C.8.27.2

M

Intravascular OCT Acquisition Parameters

C.8.27.3

M

Intravascular OCT Processing Parameters

C.8.27.4

C - Required if Presentation Intent Type (0008,0068) is FOR PROCESSING.

Intravascular Image Acquisition Parameters

C.8.27.5

M

SOP Common

C.12.1

M

Common Instance Reference

C.12.2

M

Frame Extraction

C.12.3

C - Required if the SOP Instance was created in response to a Frame-Level retrieve request


A.66.3.1 Intravascular Optical Coherence Tomography Image IOD Content Constraints

The following constraints on Image Attributes take precedence over the descriptions given in the Module Attribute Tables.

A.66.3.1.1 Contrast/Bolus Agent Sequence

For Contrast/Bolus Agent Sequence (0018,0012) DCID 3850 “Contrast Bolus Substance” shall be used.

A.66.3.1.2 Prohibited Modules

The Overlay Plane Module and VOI LUT Module shall not be used in a Standard Extended SOP Class of the Intravascular Optical Coherence Tomography Image.

Note

In order to annotate images, whether during acquisition or subsequently, SOP Instances of the Grayscale Softcopy Presentation State Storage or the Structured Report Storage SOP Classes that reference the image SOP Instance may be used.

Pseudo-color presentation information may be applied through the use of separate Pseudo-color Softcopy Presentation State SOP Instances.

No standard mechanism is provided for inclusion of annotations within the image SOP Instance itself and implementers are discouraged from using private extensions to circumvent this restriction.

DICOM PS3.3 2022a - Information Object Definitions