Table A.53-1. X-Ray 3D Angiographic Image IOD Modules
The Overlay Plane Module, VOI LUT Module and Softcopy Presentation LUT Module shall not be used in a Standard Extended SOP Class of the X-Ray 3D Angiographic Image.
The VOI LUT function is provided by a Frame VOI LUT Functional Group.
The Curve Module (Retired) was previously included in the Image IE for this IOD but has been retired. See PS3.3-2004.
For X-Ray 3D Angiographic Images created from SOP Instances of the Enhanced XA SOP Class (1.2.840.10008.5.1.4.1.1.12.1.1) the isocenter coordinate system is used to describe the positioning of the table and positioner (see Section C.8.19.6.13), and shall use only the Defined Term ISOCENTER for Equipment Coordinate System Identification (0028,9537).