DICOM PS3.2 2022a - Conformance

B.8 Annexes

B.8.1 IOD Contents

B.8.1.1 Created SOP Instances

Examples of X-Ray Radiofluoroscopic images and Grayscale Softcopy Presentation States created by EXAMPLE-INTEGRATED-MODALITY can be downloaded from:

http://www.example-imaging-products.nocom/example-integrated-modality/example-images

Table B.8.1-1 specifies the attributes of an X-Ray Radiofluoroscopic Image transmitted by the EXAMPLE-INTEGRATED-MODALITY storage application.

Table B.8.1-2 specifies the attributes of a Grayscale Softcopy Presentation State transmitted by the EXAMPLE­INTEGRATED-MODALITY storage application.

The following tables use a number of abbreviations. The abbreviations used in the "Presence of …" column are:

VNAP

Value Not Always Present (attribute sent zero length if no value is present)

ANAP

Attribute Not Always Present

ALWAYS

Always Present

EMPTY

Attribute is sent without a value

The abbreviations used in the "Source" column:

MWL

the attribute value source Modality Worklist

USER

the attribute value source is from User input

AUTO

the attribute value is generated automatically

MPPS

the attribute value is the same as that use for Modality Performed Procedure Step

CONFIG

the attribute value source is a configurable parameter

Note

All dates and times are encoded in the local configured calendar and time. Date, Time and Time zone are configured using the Service/Installation Tool.

B.8.1.1.1 X-Ray Radiofluoroscopic Image IOD

Table B.8.1-1. IOD of Created Rf SOP Instances

IE

Module

Reference

Presence of Module

Patient

Patient

Table B.8.1-3

ALWAYS

Study

General Study

Table B.8.1-4

ALWAYS

Patient Study

Table B.8.1-5

ALWAYS

Series

General Series

Table B.8.1-6

ALWAYS

Equipment

General Equipment

Table B.8.1-7

ALWAYS

Image

General Image

Table B.8.1-8

ALWAYS

Image Pixel

Table B.8.1-10

ALWAYS

Cine

Table B.8.1-11

Only if Multi-frame

Multi-Frame

Table B.8.1-12

Only if Multi-frame

Frame Pointers

Table B.8.1-13

Only if Multi-frame

Mask

Table B.8.1-14

ALWAYS

X-Ray Image

Table B.8.1-15

ALWAYS

X-Ray Acquisition

Table B.8.1-16

ALWAYS

Modality LUT

Table B.8.1-17

Only if Pixel Intensity Relationship (0028,1040) is LOG

VOI LUT

Table B.8.1-18

ALWAYS

SOP Common

Table B.8.1-19

ALWAYS

Private Application

Table B.8.1-8

ALWAYS


B.8.1.1.2 Grayscale Softcopy Presentation State IOD

Table B.8.1-2. IOD of Created Grayscale Softcopy Presentation State SOP Instances

IE

Module

Reference

Presence of Module

Patient

Patient

Table B.8.1-3

ALWAYS

Study

General Study

Table B.8.1-4

ALWAYS

Patient Study

Table B.8.1-5

ALWAYS

Series

General Series

Table B.8.1-6

ALWAYS

Presentation Series

Table B.8.1-20

ALWAYS

Equipment

General Equipment

Table B.8.1-7

ALWAYS

Presentation State

Presentation State

Table B.8.1-21

ALWAYS

Display Shutter

Table B.8.1-22

Only if Shutter applied

Displayed Area

Table B.8.1-23

ALWAYS

Graphic Annotation

Table B.8.1-24

Only if Graphic Annotations are present

Spatial Transformation

Table B.8.1-25

Only if Spacial Transformation applied

Graphic Layer

Table B.8.1-26

Only if Graphic Annotations are present

Modality LUT

Table B.8.1-27

ALWAYS

Softcopy VOI LUT

Table B.8.1-28

ALWAYS

Softcopy Presentation LUT

Table B.8.1-29

ALWAYS

SOP Common

Table B.8.1-19

ALWAYS

Private Application

Table B.8.1-8

ALWAYS


B.8.1.1.3 Common Modules

Table B.8.1-3. Patient Module of Created SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

Patient's Name

(0010,0010)

PN

From Modality Worklist or user input. Values supplied via Modality Worklist will be entered as received. Values supplied via user input will contain all 5 components (some possibly empty). . Maximum 64 characters.

VNAP

MWL/USER

Patient ID

(0010,0020)

LO

From Modality Worklist or user input. Maximum 64 characters.

VNAP

MWL/USER

Patient's Birth Date

(0010,0030)

DA

From Modality Worklist or user input

VNAP

MWL/USER

Patient's Sex

(0010,0040)

CS

From Modality Worklist or user input

VNAP

MWL/USER

Patient Comments

(0010,4000)

LT

From User Input. Maximum 1024 characters.

VNAP

USER


Table B.8.1-4. General Study Module of Created SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

Study Instance UID

(0020,000D)

UI

From Modality Worklist or generated by device

ALWAYS

MWL/AUTO

Study Date

(0008,0020)

DA

<yyyymmdd>

ALWAYS

AUTO

Study Time

(0008,0030)

TM

<hhmmss>

ALWAYS

AUTO

Referring Physician's Name

(0008,0090)

PN

From Modality Worklist

VNAP

MWL

Study ID

(0020,0010)

SH

Requested Procedure ID from Worklist or User Input

VNAP

MWL/USER

Accession Number

(0008,0050)

SH

From Modality Worklist or user input

VNAP

MWL/USER

Study Description

(0008,1030)

LO

Comment text box in study list. Maximum 1024 characters.

VNAP

USER

Referenced Study Sequence

(0008,1110)

SQ

From Modality Worklist

VNAP

MWL

>Referenced SOP Class UID

(0008,1150)

UI

From Modality Worklist

VNAP

MWL

>Referenced SOP Instance UID

(0008,1155)

UI

From Modality Worklist

VNAP

MWL


Table B.8.1-5. Patient Study Module of Created SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

Admitting Diagnosis Description

(0008,1080)

LO

From Modality Worklist

VNAP

MWL

Patient's Age

(0010,1010)

AS

Calculated from DoB input on base of actual Date

ALWAYS

AUTO

Patient's Weight

(0010,1030)

DS

From Modality Worklist or user input

VNAP

MWL/USER


Table B.8.1-6. General Series Module of Created SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

Modality

(0008,0060)

CS

RF

ALWAYS

AUTO

Series Instance UID

(0020,000E)

UI

Generated by device

ALWAYS

AUTO

Series Number

(0020,0011)

IS

Generated by device

ALWAYS

AUTO

Series Date

(0008,0021)

DA

<yyyymmdd>

ALWAYS

AUTO

Series Time

(0008,0031)

TM

<hhmmss>

ALWAYS

AUTO

Performing Physician's Name

(0008,1050)

PN

Physician field in Study list. Maximum 64 characters.

VNAP

USER

Protocol Name

(0018,1030)

LO

Organ program

ALWAYS

AUTO

Series Description

(0008,103E)

LO

Organ from Study list. Maximum 512 characters.

VNAP

USER

Operator's Name

(0008,1070)

PN

Operator field in Study list. Maximum 64 characters.

VNAP

USER

Referenced Performed Procedure Step Sequence

(0008,1111)

SQ

Identifies the MPPS SOP Instance to which this image is related

ALWAYS

MPPS

>Referenced SOP Class UID

(0008,1150)

UI

MPPS SOP Class UID

ALWAYS

MPPS

>Referenced SOP Instance UID

(0008,1155)

UI

MPPS SOP Instance UID

ALWAYS

MPPS

Request Attributes Sequence

(0040,0275)

SQ

Zero or 1 item will be present

ALWAYS

AUTO

>Requested Procedure ID

(0040,1001)

SH

From Modality Worklist

VNAP

MWL

>Scheduled Procedure Step ID

(0040,0009)

SH

From Modality Worklist

VNAP

MWL

>Scheduled Procedure Step Description

(0040,0007)

LO

From Modality Worklist

VNAP

MWL

>Scheduled Protocol Code Sequence

(0040,0008)

SQ

From Modality Worklist

VNAP

MWL

Performed Procedure Step ID

(0040,0253)

SH

Same as MPPS.

ALWAYS

MPPS

Performed Procedure Step Start Date

(0040,0244)

DA

Same as MPPS

ALWAYS

MPPS

Performed Procedure Step Start Time

(0040,0245)

TM

Same as MPPS

ALWAYS

MPPS

Performed Procedure Step Description

(0040,0254)

LO

Same as MPPS. From user input. Maximum 64 characters.

VNAP

MPPS

Performed Protocol Code Sequence

(0040,0260)

SQ

Same as MPPS

ALWAYS

MPPS

Comments on the Performed Procedure Step

(0040,0280)

LO

Same as MPPS. From user input. Maximum 64 characters.

VNAP

MPPS


Table B.8.1-7. General Equipment Module of Created SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

Manufacturer

(0008,0070)

LO

EXAMPLE-IMAGING-PRODUCTS

ALWAYS

AUTO

Institution Name

(0008,0080)

LO

From Configuration

VNAP

CONFIG

Station Name

(0008,1010)

SH

From Configuration

ALWAYS

CONFIG

Manufacturer's Model Name

(0008,1090)

LO

EXAMPLE-INTEGRATED-MODALITY

ALWAYS

AUTO

Device Serial Number

(0018,1000)

LO

From Configuration

ALWAYS

CONFIG

Software Version

(0018,1020)

LO

From Configuration

ALWAYS

CONFIG

Private Creator

(0009,00xx)

LO

EXINTMOD_EQ_01

ALWAYS

AUTO

Equipment UID

(0009,xx01)

UI

From Configuration

ALWAYS

CONFIG

Service UID

(0009,xx02)

UI

From Configuration

ALWAYS

CONFIG


Table B.8.1-8. Private Application Module of Created SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

Private Creator

(0029,00xx)

LO

EXINTMOD_IM_01

ALWAYS

AUTO

Application Header Sequence

(0029,xx40)

SQ

Zero or more items. Each item contains private application data from a different application.

VNAP

AUTO

>Private Creator

(0029,00xx)

LO

EXINTMOD_IM_01

ALWAYS

AUTO

> Application Header Type

(0029,xx41)

CS

One of PLATFORM or PLUGIN

ALWAYS

AUTO

> Application Header ID

(0029,xx42)

LO

One of ACQUISITION, IMAGE PROCESSING, VIEWER, AUDIT, ACCESS, ROUTING or STATUS

ALWAYS

AUTO

> Application Header Version

(0029,xx43)

LO

From Application

ALWAYS

AUTO

> Application Header Data

(0029,xx44)

OB

From Application

ALWAYS

AUTO

Workflow Control Flags

(0029,xx50)

LO

One or more of:P: printedcom: completedrea: readver: verifiedRI: receivedAC: archived and committedE: exportedm: marked

VNAP

AUTO

Archive Management Flag - Keep Online

(0029,xx51)

CS

00 = remote control not required (default)01 = keep instance online.

ALWAYS

AUTO

Archive Management Flag - Do Not Archive

(0029,xx52)

CS

00 = remote control not required (default)01 = do not archive instance.

ALWAYS

AUTO


B.8.1.1.4 X-Ray Radiofluoroscopic Image Modules

Table B.8.1-9. General Image Module of Created Rf SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

Instance Number

(0020,0013)

IS

Generated by device

ALWAYS

AUTO

Patient Orientation

(0020,0020)

CS

Zero length

EMPTY

AUTO

Content Date

(0008,0023)

DA

<yyyymmdd>

ALWAYS

AUTO

Content Time

(0008,0033)

TM

<hhmmss>

ALWAYS

AUTO

Acquisition Number

(0020,0012)

IS

Generated by device

ALWAYS

AUTO

Image Comments

(0020,4000)

LT

From user input. Maximum 1024 characters.

VNAP

USER

Anatomic Region Sequence

(0008,2218)

SQ

From user input.

ALWAYS

USER

> Include 'Code Sequence Macro'

Baseline Context ID is 4009 (see also Section B.8.6)


Table B.8.1-10. Image Pixel Module of Created Rf SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

Pixel Data

(7FE0,0010)

OW

The Pixel Data itself does not contain any burned-in annotation.

ALWAYS

AUTO


Table B.8.1-11. Cine Module of Created Rf SOP

Attribute Name

Tag

VR

Value

Presence of Value

Source

Frame Time

(0018,1063)

DS

Only if multi-frame.

ANAP

AUTO

Recommended Display Frame Rate

(0008,2144)

IS

Only if multi-frame Same as Cine Rate

ANAP

AUTO

Cine Rate

(0018,0040)

IS

Only if multi-frame

ANAP

AUTO


Table B.8.1-12. Multi-Frame Module of Created Rf SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

Number of Frames

(0028,0008)

IS

Only if multi-frame

ANAP

AUTO


Table B.8.1-13. Frame Pointers Module of Created Rf SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

Representative Frame Number

(0028,6010)

US

Only if multi-frame

ANAP

AUTO


Table B.8.1-14. Mask Module of Created Rf SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

Mask Subtraction Sequence

(0028,6100)

SQ

Only if multi-frame and (0028,1040) = LOG

ANAP

AUTO

> Mask Operation

(0028,6101)

CS

AVG_SUB

ANAP

AUTO

> Mask Frame Numbers

(0028,6110)

US

Mask Frame Number

ANAP

AUTO

Recommended Viewing Mode

(0028,1090)

CS

NAT or SUB

ALWAYS

AUTO


Table B.8.1-15. X-Ray Image Module of Created Rf SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

Frame Increment Pointer

(0028,0009)

AT

<0018,1063> only if multi-frame

ANAP

AUTO

Image Type

(0008,0008)

CS

ORIGINAL\PRIMARY\SINGLE PLANE\DAS

(acquired images)

ORIGINAL\DERIVED\SINGLE PLANE

(post-processed images)

ALWAYS

AUTO

Pixel Intensity Relationship

(0028,1040)

CS

LIN or LOG

ALWAYS

AUTO

Samples per Pixel

(0028,0002)

US

1

ALWAYS

AUTO

Photometric Interpretation

(0028,0004)

CS

MONOCHROME2

ALWAYS

AUTO

Rows

(0028,0010)

US

1024

ALWAYS

AUTO

Columns

(0028,0011)

US

1024

ALWAYS

AUTO

Bits Allocated

(0028,0100)

US

16

ALWAYS

AUTO

Bits Stored

(0028,0101)

US

10

ALWAYS

AUTO

High Bit

(0028,0102)

US

9

ALWAYS

AUTO

Pixel Representation

(0028,0103)

US

0000H

ALWAYS

AUTO


Table B.8.1-16. X-Ray Acquisition Module of Created Rf SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

KVP

(0018,0060)

DS

From Acquisition parameters

ALWAYS

AUTO

Radiation Setting

(0018,1155)

CS

GR

ALWAYS

AUTO

X-Ray Tube Current

(0018,1151)

IS

From Acquisition parameters

ALWAYS

AUTO

Exposure Time

(0018,1150)

IS

From Acquisition parameters

ALWAYS

AUTO

Radiation Mode

(0018,115A)

CS

CONTINUOUS

ALWAYS

AUTO

Intensifier Size

(0018,1162)

DS

From Acquisition parameters

ALWAYS

AUTO

Private Creator

(0019,00xx)

LO

EXINTMOD_AQ_01

ALWAYS

AUTO

Edge Enhancement Percent

(0019,xx10)

IS

0 .. 100

VNAP

AUTO

Landmark

(0019,xx20)

IS

0 .. 100

VNAP

AUTO

Pixel Shift Horizontal

(0019,xx30)

DS

-20 .. +20

VNAP

AUTO

Pixel Shift Vertical

(0019,xx40)

DS

-20 .. +20

VNAP

AUTO


Table B.8.1-17. Modality LUT Module of Created Rf SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

Modality LUT Sequence

(0028,3000)

SQ

present if (0028,1040) = LOG

ANAP

AUTO

> LUT Descriptor

(0028,3002)

US

<1024,0,16>

ANAP

AUTO

> Modality LUT Type

(0028,3004)

LO

US

ANAP

AUTO

> LUT Data

(0028,3006)

US

LUT

ANAP

AUTO


Table B.8.1-18. VOI LUT Module of Created RF SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

Window Center

(0028,1050)

DS

0…1023

ALWAYS

AUTO

Window Width

(0028,1051)

DS

1…1024

ALWAYS

AUTO


Table B.8.1-19. SOP Common Module of Created Rf SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

Specific Character Set

(0008,0005)

CS

"ISO_IR 100" or "ISO_IR 144"

ALWAYS

CONFIG

SOP Class UID

(0008,0016)

UI

1.2.840.10008.5.1.4.1.1.12.2

ALWAYS

AUTO

SOP Instance UID

(0008,0018)

UI

Generated by device

ALWAYS

AUTO


B.8.1.1.5 Grayscale Softcopy Presentation State Modules

Table B.8.1-20. Presentation Series Module of Created GSPS SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

Modality

(0008,0060)

CS

PR

ALWAYS

AUTO


Table B.8.1-21. Presentation State Module of Created GSPS SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

Instance Number

(0020,0013)

IS

Generated by device

ALWAYS

AUTO

Presentation Label

(0070,0080)

CS

From user input.

ALWAYS

USER

Presentation Description

(0070,0081)

LO

From user input.

VNAP

USER

Presentation Creation Date

(0070,0082)

DA

Generated by device

ALWAYS

AUTO

Presentation Creation Time

(0070,0083)

TM

Generated by device

ALWAYS

AUTO

Presentation Creator's Name

(0070,0084)

PN

Generated by device according to currently active user.

ALWAYS

AUTO

Referenced Series Sequence

(0008,1115)

SQ

One or more items.

ALWAYS

AUTO

>Series Instance UID

(0020,000E)

UI

From referenced image

ALWAYS

AUTO

>Referenced Image Sequence

(0008,1140)

SQ

From referenced image

ALWAYS

AUTO

>>Referenced SOP Class UID

(0008,1150)

UI

From referenced image

ALWAYS

AUTO

>>Referenced SOP Instance UID

(0008,1155)

UI

From referenced image

ALWAYS

AUTO

>>Referenced Frame Number

(0008,1160)

IS

If referenced image is a multi-frame image

ANAP

AUTO

Shutter Presentation Value

(0018,1622)

US

Generated by device if shutter present

ANAP

AUTO


Table B.8.1-22. Display Shutter Module of Created GSPS SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

Shutter Shape

(0018,1600)

CS

If shutter applied: RECTANGULAR\CIRCULAR

ANAP

AUTO

Shutter Left Vertical Edge

(0018,1602)

IS

If RECTANGULAR shutter applied

ANAP

AUTO

Shutter Right Vertical Edge

(0018,1604)

IS

If RECTANGULAR shutter applied

ANAP

AUTO

Shutter Upper Horizontal Edge

(0018,1606)

IS

If RECTANGULAR shutter applied

ANAP

AUTO

Shutter Lower Horizontal Edge

(0018,1608)

IS

If RECTANGULAR shutter applied

ANAP

AUTO

Center of Circular Shutter

(0018,1610)

IS

If CIRCULAR shutter applied

ANAP

AUTO

Radius of Circular Shutter

(0018,1612)

IS

If CIRCULAR shutter applied

ANAP

AUTO


Table B.8.1-23. Displayed Area Module of Created GSPS SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

Displayed Area Selection Sequence

(0070,005A)

SQ

One or more items

ALWAYS

AUTO

>Referenced Image Sequence

(0008,1140)

SQ

One or more items

ALWAYS

AUTO

>>Referenced SOP Class UID

(0008,1150)

UI

From referenced image

ALWAYS

AUTO

>>Referenced SOP Instance UID

(0008,1155)

UI

From referenced image

ALWAYS

AUTO

>>Referenced Frame Number

(0008,1160)

IS

If referenced image is a multi-frame image

ANAP

AUTO

>Displayed Area Top Left Hand Corner

(0070,0052)

SL

From current display setting

ALWAYS

AUTO

>Displayed Area Bottom Right Hand Corner

(0070,0053)

SL

From current display setting

ALWAYS

AUTO

>Presentation Size Mode

(0070,0100)

CS

From current display setting

ALWAYS

AUTO

>Presentation Pixel Spacing

(0070,0101)

DS

From current display setting

ANAP

AUTO

>Presentation Pixel Aspect Ratio

(0070,0102)

IS

From current display setting

ANAP

AUTO

>Presentation Pixel Magnification Ratio

(0070,0103)

FL

From current display setting

ANAP

AUTO


Table B.8.1-24. Graphic Annotation Module of Created GSPS SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

Graphic Annotation Sequence

(0070,0001)

SQ

One or more items

ANAP

AUTO

>Referenced Image Sequence

(0008,1140)

SQ

One or more items

ALWAYS

AUTO

>>Referenced SOP Class UID

(0008,1150)

UI

From referenced image

ALWAYS

AUTO

>>Referenced SOP Instance UID

(0008,1155)

UI

From referenced image

ALWAYS

AUTO

>>Referenced Frame Number

(0008,1160)

IS

If referenced image is a multi-frame image

ANAP

AUTO

>Graphic Layer

(0070,0002)

CS

Layer in Graphic Layer Module

ALWAYS

AUTO

>Text Object Sequence

(0070,0008)

SQ

One or more items if text annotation present

ANAP

AUTO

>>Anchor Point Annotation Units

(0070,0004)

CS

PIXEL

ALWAYS

AUTO

>>Unformatted Text Value

(0070,0006)

ST

From user input

ALWAYS

USER

>>Anchor Point

(0070,0014)

FL

From user input

ALWAYS

USER

>>Anchor Point Visibility

(0070,0015)

CS

From user input

ALWAYS

USER

>Graphic Object Sequence

(0070,0009)

SQ

One or more items if graphic annotation present

ANAP

AUTO

>>Graphic Annotation Units

(0070,0005)

CS

PIXEL

ALWAYS

AUTO

>>Graphic Dimensions

(0070,0020)

US

2

ALWAYS

AUTO

>>Number of Graphic Points

(0070,0021)

US

From user input

ALWAYS

USER

>>Graphic Data

(0070,0022)

FL

From user input

ALWAYS

USER

>>Graphic Type

(0070,0023)

CS

One of POINT, POLYLINE, INTERPOLATED, CIRCLE or ELLIPSE

ALWAYS

USER

>>Graphic Filled

(0070,0024)

CS

From user input

ANAP

USER

Attribute Name

Tag

VR

Value

Presence of Value

Source

Graphic Annotation Sequence

(0070,0001)

SQ

One or more items

ANAP

AUTO

>Referenced Image Sequence

(0008,1140)

SQ

One or more items

ALWAYS

AUTO

>>Referenced SOP Class UID

(0008,1150)

UI

From referenced image

ALWAYS

AUTO

>>Referenced SOP Instance UID

(0008,1155)

UI

From referenced image

ALWAYS

AUTO

>>Referenced Frame Number

(0008,1160)

IS

If referenced image is a multi-frame image

ANAP

AUTO

>Graphic Layer

(0070,0002)

CS

Layer in Graphic Layer Module

ALWAYS

AUTO

>Text Object Sequence

(0070,0008)

SQ

One or more items if text annotation present

ANAP

AUTO

>>Anchor Point Annotation Units

(0070,0004)

CS

PIXEL

ALWAYS

AUTO

>>Unformatted Text Value

(0070,0006)

ST

From user input

ALWAYS

USER

>>Anchor Point

(0070,0014)

FL

From user input

ALWAYS

USER

>>Anchor Point Visibility

(0070,0015)

CS

From user input

ALWAYS

USER

>Graphic Object Sequence

(0070,0009)

SQ

One or more items if graphic annotation present

ANAP

AUTO

>>Graphic Annotation Units

(0070,0005)

CS

PIXEL

ALWAYS

AUTO

>>Graphic Dimensions

(0070,0020)

US

2

ALWAYS

AUTO

>>Number of Graphic Points

(0070,0021)

US

From user input

ALWAYS

USER

>>Graphic Data

(0070,0022)

FL

From user input

ALWAYS

USER

>>Graphic Type

(0070,0023)

CS

One of POINT, POLYLINE, INTERPOLATED, CIRCLE or ELLIPSE

ALWAYS

USER

>>Graphic Filled

(0070,0024)

CS

From user input

ANAP

USER


Table B.8.1-25. Spatial Transformation Module of Created GSPS SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

Image Rotation

(0070,0042)

US

From current display setting

ANAP

AUTO

Image Horizontal Flip

(0070,0041)

CS

From current display setting

ANAP

AUTO


Table B.8.1-26. Graphic Layer Module of Created GSPS SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

Graphic Layer Sequence

(0070,0060)

SQ

One or more items

ANAP

AUTO

>Graphic Layer

(0070,0002)

CS

LAYER1. LAYER2, LAYER3, …

ALWAYS

AUTO

>Graphic Layer Order

(0070,0062)

IS

From current display setting

ALWAYS

AUTO


Table B.8.1-27. Modality LUT Module of Created GSPS SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

Modality LUT Sequence

(0028,3000)

SQ

One item

ANAP

AUTO

>LUT Descriptor

(0028,3002)

US

<1024,0,16>

ALWAYS

AUTO

>Modality LUT Type

(0028,3004)

LO

US

ALWAYS

AUTO

>LUT Data

(0028,3006)

US

LUT

ALWAYS

AUTO

Rescale Intercept

(0028,1052)

DS

1

ANAP

AUTO

Rescale Slope

(0028,1053)

DS

0

ANAP

AUTO

Rescale Type

(0028,1054)

LO

US

ANAP

AUTO


Table B.8.1-28. Softcopy VOI LUT Module of Created GSPS SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

Softcopy VOI LUT Sequence

(0028,3110)

SQ

One or more items

ALWAYS

AUTO

>Referenced Image Sequence

(0008,1140)

SQ

One or more items

ALWAYS

AUTO

>>Referenced SOP Class UID

(0008,1150)

UI

From referenced image

ALWAYS

AUTO

>>Referenced SOP Instance UID

(0008,1155)

UI

From referenced image

ALWAYS

AUTO

>>Referenced Frame Number

(0008,1160)

IS

If referenced image is a multi-frame image

ANAP

AUTO

>Window Center

(0028,1050)

DS

From current display setting: 0…1023

ALWAYS

AUTO

>Window Width

(0028,1051)

DS

From current display setting: 1…1024

ALWAYS

AUTO

>Window Center & Width Explanation

(0028,1055)

LO

Name of Window Preset

ANAP

AUTO


Table B.8.1-29. Softcopy Presentation LUT Module of Created GSPS SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

Presentation LUT Shape

(2050,0020

CS

IDENTITY

ALWAYS

AUTO


Table B.8.1-30. SOP Common Module of Created GSPS SOP Instances

Attribute Name

Tag

VR

Value

Presence of Value

Source

Specific Character Set

(0008,0005)

CS

"ISO_IR 100" or "ISO_IR 144"

ALWAYS

CONFIG

SOP Class UID

(0008,0016)

UI

1.2.840.10008.5.1.4.1.1.11.1

ALWAYS

AUTO

SOP Instance UID

(0008,0018)

UI

Generated by device

ALWAYS

AUTO


B.8.1.2 Used Fields in Received IOD By Application

The EXAMPLE-INTEGRATED-MODALITY storage application does not receive SOP Instances. The usage of attributes received via Modality Worklist is described in Section B.4.2.2.3.1.3.

B.8.1.3 Attribute Mapping

The relationships between attributes received via Modality Worklist, stored in acquired images and communicated via MPPS are summarized in Table B.8.1-31. The format and conventions used in DICOM Table B.8.1-31 are the same as the corresponding table in Section J.6 in PS3.17 .

Table B.8.1-31. Attribute Mapping Between Modality Worklist, Image and MPPS

Modality Worklist

Image IOD

MPPS IOD

Patient Name

Patient Name

Patient Name

Patient ID

Patient ID

Patient ID

Patient's Birth Date

Patient's Birth Date

Patient's Birth Date

Patient's Sex

Patient's Sex

Patient's Sex

Patient's Weight

Patient's Weight

Referring Physician's Name

Referring Physician's Name

----

----

Scheduled Step Attributes Sequence

Study Instance UID

Study Instance UID

>Study Instance UID

Referenced Study Sequence

Referenced Study Sequence

>Referenced Study Sequence

Accession Number

Accession Number

>Accession Number

----

Request Attributes Sequence

----

Requested Procedure ID

>Requested Procedure ID

>Requested Procedure ID

Requested Procedure Description

>Requested Procedure Description

Scheduled Procedure Step ID

>Scheduled Procedure Step ID

>Scheduled Procedure Step ID

Scheduled Procedure Step Description

>Scheduled Procedure Step Description

>Scheduled Procedure Step Description

Scheduled Protocol Code Sequence

>Scheduled Protocol Code Sequence

----

----

Performed Protocol Code Sequence

Performed Protocol Code Sequence

----

Study ID

Study ID

----

Performed Procedure Step ID

Performed Procedure Step ID

----

Performed Procedure Step Start Date

Performed Procedure Step Start Date

----

Performed Procedure Step Start Time

Performed Procedure Step Start Time

----

Performed Procedure Step Description

Performed Procedure Step Description

----

Comments on the Performed Procedure Step

Comments on the Performed Procedure Step

----

----

Performed Series Sequence

Scheduled Performing Physician's Name

Performing Physician's Name

>Performing Physician's Name

Requested Procedure Code Sequence

----

Procedure Code Sequence

----

Referenced Study Component Sequence

----

----

>Referenced SOP Class UID

SOP Class UID

----

>Referenced SOP Instance UID

SOP Instance UID

----

Protocol Name

Protocol Name


B.8.1.4 Coerced/Modified Fields

The Modality Worklist AE will truncate attribute values received in the response to a Modality Worklist Query if the value length is longer than the maximum length permitted by the attribute's VR.