B.8.2 Data Dictionary of Private Attributes

The Private Attributes added to created SOP Instances are listed in the Table below. EXAMPLE-INTEGRATED-MODALITY reserves blocks of private attributes in groups 0009, 0019 and 0029. Further details on usage of these private attributes are contained in Section B.8.1.

Table B.8.2-1. Data Dictionary of Private Attributes

Tag

Attribute Name

VR

VM

Attribute Description

(0009,00xx)

Private Creator

LO

1

EXINTMODMAKER

(0009,xx01)

Equipment UID

UI

1

(0009,xx02)

Service UID

UI

1

(0019,00xx)

Private Creator

LO

1

EXINTMODMAKER

(0019,xx10)

Edge Enhancement Percent

IS

1

(0019,xx20)

Landmark

IS

1

(0019,xx30)

Pixel Shift Horizontal

DS

1

(0019,xx40)

Pixel Shift Vertical

DS

1

(0029,00xx)

Private Creator

LO

1

EXINTMODMAKER

(0029,xx40)

Application Header Sequence

SQ

1

(0029,xx41)

Application Header Type

CS

1

(0029,xx42)

Application Header ID

LO

1

(0029,xx43)

Application Header Version

LO

1

(0029,xx44)

Application Header Data

OB

1

(0029,xx50)

Workflow Control Flags

LO

8

(0029,xx51)

Archive Management Flag - Keep Online

CS

1

(0029,xx52)

Archive Management Flag - Do Not Archive

CS

1


DICOM PS3.2 2022a - Conformance