DICOM PS3.2 2022a - Conformance

A.1 Conformance Statement Overview

The Overview consist of typically 5-10 lines describing the network services and media storage capabilities supported by the product in layman's terms (i.e., no DICOM acronyms should be used).

A table of Supported Networking DICOM Service (SOP) Classes is provided with roles (User/Provider), organized in 4 categories:

The first column shall specify the SOP Classes exactly as named in PS3.6., Registry of DICOM Unique Identifiers. The phrase "and specializations" may be added to indicate support of all specializations negotiated through the SOP Class Common Extended Negotiation. If the implementation supports all SOP Classes of a particular Service Class through SOP Class Common Extended Negotiation, the first column shall specify "All services of the <x> Service Class".

Table A.1-1. Network Services

SOP Classes

User of Service (SCU)

Provider of Service (SCP)

Transfer

CT Image Storage

Yes

No

US Image Storage

Yes

Yes

Query/Retrieve

Patient Root Information Model FIND

Option

No

Notes, Reports, Measurements Transfer

Comprehensive SR, and specializations

No

Yes


The services can be specified as a SCU, SCP or as an Option, which means that it is either configurable or that it can be purchased separately.

Note

Verification SCP (C-Echo) is not included in the table above because it is required for any Acceptor of an Association. The Verification SCU details are covered in the details of the conformance statement.

The SOP Classes are categorized as follows:

Table A.1-2. UID Values

UID Value

UID Name

Category

1.2.840.10008.1.20.1

Storage Commitment Push Model SOP Class

Workflow Management

1.2.840.10008.1.40

Procedural Event Logging SOP Class

Workflow Management

1.2.840.10008.1.42

Substance Administration Logging SOP Class

Workflow Management

1.­2.840.10008.3.1.2.3.3

Modality Performed Procedure Step SOP Class

Workflow Management

1.­2.840.10008.3.1.2.3.4

Modality Performed Procedure Step Retrieve SOP Class

Workflow Management

1.­2.840.10008.3.1.2.3.5

Modality Performed Procedure Step Notification SOP Class

Workflow Management

1.2.840.10008.4.2

Storage Service Class

Transfer

1.2.840.10008.5.1.1.1

Basic Film Session SOP Class

Print Management

1.2.840.10008.5.1.1.2

Basic Film Box SOP Class

Print Management

1.2.840.10008.5.1.1.4

Basic Grayscale Image Box SOP Class

Print Management

1.2.840.10008.5.1.1.4.1

Basic Color Image Box SOP Class

Print Management

1.2.840.10008.5.1.1.9

Basic Grayscale Print Management Meta SOP Class

Print Management

1.2.840.10008.5.1.1.14

Print Job SOP Class

Print Management

1.2.840.10008.5.1.1.15

Basic Annotation Box SOP Class

Print Management

1.2.840.10008.5.1.1.16

Printer SOP Class

Print Management

1.2.840.10008.5.1.1.16.376

Printer Configuration Retrieval SOP Class

Print Management

1.2.840.10008.5.1.1.18

Basic Color Print Management Meta SOP Class

Print Management

1.2.840.10008.5.1.1.23

Presentation LUT SOP Class

Print Management

1.2.840.10008.5.1.1.24.1

Basic Print Image Overlay Box SOP Class

Print Management

1.2.840.10008.5.1.1.33

Media Creation Management SOP Class

Print Management

1.2.840.10008.5.1.4.1.1.1

Computed Radiography Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.1.1

Digital X-Ray Image Storage - For Presentation SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.1.1.1

Digital X-Ray Image Storage - For Processing SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.1.2

Digital Mammography X-Ray Image Storage - For Presentation SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.1.2.1

Digital Mammography X-Ray Image Storage - For Processing SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.1.3

Digital Intra-oral X-Ray Image Storage - For Presentation SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.1.3.1

Digital Intra-oral X-Ray Image Storage - For Processing SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.2

CT Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.2.1

Enhanced CT Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.3.1

Ultrasound Multi-frame Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.4

MR Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.4.1

Enhanced MR Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.4.2

MR Spectroscopy Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.4.3

Enhanced MR Color Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.6.1

Ultrasound Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.6.2

Enhanced US Volume Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.7

Secondary Capture Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.7.1

Multi-frame Single Bit Secondary Capture Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.7.2

Multi-frame Grayscale Byte Secondary Capture Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.7.3

Multi-frame Grayscale Word Secondary Capture Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.7.4

Multi-frame True Color Secondary Capture Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.9.1.1

12-lead ECG Waveform Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.9.1.2

General ECG Waveform Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.9.1.3

Ambulatory ECG Waveform Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.9.2.1

Hemodynamic Waveform Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.9.3.1

Cardiac Electrophysiology Waveform Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.9.4.1

Basic Voice Audio Waveform Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.9.4.2

General Audio Waveform Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.9.5.1

Arterial Pulse Waveform Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.9.6.1

Respiratory Waveform Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.9.6.2

Multi-channel Respiratory Waveform Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.9.7.1

Routine Scalp Electroencephalogram Waveform Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.9.7.2

Electromyogram Waveform Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.9.7.3

Electrooculogram Waveform Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.9.7.4

Sleep Electroencephalogram Waveform Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.9.8.1

Body Position Waveform Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.11.1

Grayscale Softcopy Presentation State Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.11.2

Color Softcopy Presentation State Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.11.3

Pseudo-Color Softcopy Presentation State Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.11.4

Blending Softcopy Presentation State Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.11.5

XA/XRF Grayscale Softcopy Presentation State Storage SOP Class

Transfer

1.2.840.10008.​5.​1.​4.​1.​1.​11.​6

Grayscale Planar MPR Volumetric Presentation State Storage SOP Class

Transfer

1.2.840.10008.​5.​1.​4.​1.​1.​11.​7

Compositing Planar MPR Volumetric Presentation State Storage SOP Class

Transfer

1.2.840.10008.​5.​1.​4.​1.​1.​11.​8

Advanced Blending Presentation State Storage SOP Class

Transfer

1.2.840.10008.​5.​1.​4.​1.​1.​11.​9

Volume Rendering Volumetric Presentation State Storage SOP Class

Transfer

1.2.840.10008.​5.​1.​4.​1.​1.​11.​10

Segmented Volume Rendering Volumetric Presentation State Storage SOP Class

Transfer

1.2.840.10008.​5.​1.​4.​1.​1.​11.​11

Multiple Volume Rendering Volumetric Presentation State Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.12.1

X-Ray Angiographic Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.12.1.1

Enhanced XA Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.12.2

X-Ray Radiofluoroscopic Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.12.2.1

Enhanced XRF Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.13.1.1

X-Ray 3D Angiographic Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.13.1.2

X-Ray 3D Craniofacial Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.13.1.3

Breast Tomosynthesis Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.13.1.4

Breast Projection X-Ray Image Storage - For Presentation SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.13.1.5

Breast Projection X-Ray Image Storage - For Processing SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.14.1

Intravascular Optical Coherence Tomography Image Storage - For Presentation SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.14.2

Intravascular Optical Coherence Tomography Image Storage - For Processing SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.20

Nuclear Medicine Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.30

Parametric Map Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.66

Raw Data Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.66.1

Spatial Registration Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.66.2

Spatial Fiducials Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.66.3

Deformable Spatial Registration Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.66.4

Segmentation Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.66.5

Surface Segmentation Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.66.6

Tractography Results Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.67

Real World Value Mapping Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.68.1

Surface Scan Mesh Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.68.2

Surface Scan Point Cloud Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.77.1.1

VL Endoscopic Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.77.1.1.1

Video Endoscopic Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.77.1.2

VL Microscopic Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.77.1.2.1

Video Microscopic Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.77.1.3

VL Slide-Coordinates Microscopic Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.77.1.4

VL Photographic Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.77.1.4.1

Video Photographic Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.77.1.5.1

Ophthalmic Photography 8 Bit Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.77.1.5.2

Ophthalmic Photography 16 Bit Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.77.1.5.3

Stereometric Relationship Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.77.1.5.4

Ophthalmic Tomography Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.77.1.5.5

Wide Field Ophthalmic Photography Stereographic Projection Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.77.1.5.6

Wide Field Ophthalmic Photography 3D Coordinates Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.77.1.5.7

Ophthalmic Optical Coherence Tomography En Face Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.77.1.5.8

Ophthalmic Optical Coherence Tomography B-scan Volume Analysis Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.77.1.6

VL Whole Slide Microscopy Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.77.1.7

Dermoscopic Photography Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.78.1

Lensometry Measurements Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.78.2

Autorefraction Measurements Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.78.3

Keratometry Measurements Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.78.4

Subjective Refraction Measurements Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.78.5

Visual Acuity Measurements Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.78.6

Spectacle Prescription Report Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.78.7

Ophthalmic Axial Measurements Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.78.8

Intraocular Lens Calculations Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.79.1

Macular Grid Thickness and Volume Report Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.80.1

Ophthalmic Visual Field Static Perimetry Measurements Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.81.1

Ophthalmic Thickness Map Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.82.1

Corneal Topography Map Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.88.11

Basic Text SR Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.88.22

Enhanced SR Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.88.33

Comprehensive SR Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.88.34

Comprehensive 3D SR Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.88.40

Procedure Log Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.88.50

Mammography CAD SR Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.88.59

Key Object Selection Document Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.88.65

Chest CAD SR Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.88.67

X-Ray Radiation Dose SR Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.88.68

Radiopharmaceutical Radiation Dose SR Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.88.69

Colon CAD SR Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.88.70

Implantation Plan SR Document Storage SOP Class

Transfer

1.2.840.10008.5.​1.​4.​1.​1.​88.​71

Acquisition Context SR Storage SOP Class

Transfer

1.2.840.10008.5.​1.​4.​1.​1.​88.​72

Simplified Adult Echo SR Storage SOP Class

Transfer

1.2.840.10008.5.​1.​4.​1.​1.​88.​73

Patient Radiation Dose SR Storage SOP Class

Transfer

1.2.840.10008.5.​1.​4.​1.​1.​88.​74

Planned Imaging Agent Administration SR Storage SOP Class

Transfer

1.2.840.10008.5.​1.​4.​1.​1.​88.​75

Performed Imaging Agent Administration SR Storage SOP Class

Transfer

1.2.840.10008.5.​1.​4.​1.​1.​88.​76

Enhanced X-Ray Radiation Dose SR Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.90.1

Content Assessment Results Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.91.1

Microscopy Bulk Simple Annotations Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.104.1

Encapsulated PDF Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.104.2

Encapsulated CDA Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.104.3

Encapsulated STL Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.104.4

Encapsulated OBJ Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.104.5

Encapsulated MTL Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.128

Positron Emission Tomography Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.130

Enhanced PET Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.131

Basic Structured Display Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.200.1

CT Defined Procedure Protocol Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.200.2

CT Performed Procedure Protocol Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.200.3

Protocol Approval Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.200.7

XA Defined Procedure Protocol Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.200.8

XA Performed Procedure Protocol Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.481.1

RT Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.481.2

RT Dose Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.481.3

RT Structure Set Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.481.4

RT Beams Treatment Record Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.481.5

RT Plan Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.481.6

RT Brachy Treatment Record Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.481.7

RT Treatment Summary Record Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.481.8

RT Ion Plan Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.481.9

RT Ion Beams Treatment Record Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.481.10

RT Physician Intent Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.481.11

RT Segment Annotation Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.481.12

RT Radiation Set Storage

Transfer

1.2.840.10008.5.1.4.1.1.481.13

C-Arm Photon-Electron Radiation Storage

Transfer

1.2.840.10008.5.1.4.1.1.481.14

Tomotherapeutic Radiation Storage

Transfer

1.2.840.10008.5.1.4.1.1.481.15

Robotic-Arm Radiation Storage

Transfer

1.2.840.10008.5.1.4.1.1.481.16

RT Radiation Record Set Storage

Transfer

1.2.840.10008.5.1.4.1.1.481.17

RT Radiation Salvage Record Storage

Transfer

1.2.840.10008.5.1.4.1.1.481.18

Tomotherapeutic Radiation Record Storage

Transfer

1.2.840.10008.5.1.4.1.1.481.19

C-Arm Photon Electron Radiation Record Storage

Transfer

1.2.840.10008.5.1.4.1.1.481.20

Robotic-Arm Radiation Record Storage

Transfer

1.2.840.10008.5.1.4.1.1.481.21

RT Radiation Set Delivery Instruction Storage

Transfer

1.2.840.10008.5.1.4.1.1.481.22

RT Treatment Preparation Storage

Transfer

1.2.840.10008.5.1.4.1.2.1.1

Patient Root Query/Retrieve Information Model - FIND SOP Class

Query/Retrieve

1.2.840.10008.5.1.4.1.2.1.2

Patient Root Query/Retrieve Information Model - MOVE SOP Class

Query/Retrieve

1.2.840.10008.5.1.4.1.2.1.3

Patient Root Query/Retrieve Information Model - GET SOP Class

Query/Retrieve

1.2.840.10008.5.1.4.1.2.2.1

Study Root Query/Retrieve Information Model - FIND SOP Class

Query/Retrieve

1.2.840.10008.5.1.4.1.2.2.2

Study Root Query/Retrieve Information Model - MOVE SOP Class

Query/Retrieve

1.2.840.10008.5.1.4.1.2.2.3

Study Root Query/Retrieve Information Model - GET SOP Class

Query/Retrieve

1.2.840.10008.5.1.4.1.2.4.2

Composite Instance Root Retrieve - MOVE SOP Class

Query/Retrieve

1.2.840.10008.5.1.4.1.2.4.3

Composite Instance Root Retrieve - GET SOP Class

Query/Retrieve

1.2.840.10008.5.1.4.1.2.5.3

Composite Instance Retrieve Without Bulk Data - GET SOP Class

Query/Retrieve

1.2.840.10008.5.1.4.20.1

Defined Procedure Protocol Information Model - FIND SOP Class

Query/Retrieve

1.2.840.10008.5.1.4.20.2

Defined Procedure Protocol Information Model - MOVE SOP Class

Query/Retrieve

1.2.840.10008.5.1.4.20.3

Defined Procedure Protocol Information Model - GET SOP Class

Query/Retrieve

1.2.840.10008.5.1.4.1.1.200.4

Protocol Approval Information Model - FIND SOP Class

Query/Retrieve

1.2.840.10008.5.1.4.1.1.200.5

Protocol Approval Information Model - MOVE SOP Class

Query/Retrieve

1.2.840.10008.5.1.4.1.1.200.6

Protocol Approval Information Model - GET SOP Class

Query/Retrieve

1.2.840.10008.5.1.4.31

Modality Worklist Information Model - FIND SOP Class

Workflow Management

1.2.840.10008.5.1.4.32.1

General Purpose Worklist Information Model - FIND SOP Class (Retired)

Workflow Management

1.2.840.10008.5.1.4.32.2

General Purpose Scheduled Procedure Step SOP Class (Retired)

Workflow Management

1.2.840.10008.5.1.4.32.3

General Purpose Performed Procedure Step SOP Class (Retired)

Workflow Management

1.2.840.10008.5.1.4.32

General Purpose Worklist Management Meta SOP Class (Retired)

Workflow Management

1.2.840.10008.5.1.4.33

Instance Availability Notification SOP Class

Workflow Management

1.2.840.10008.5.1.4.34.6.1

Unified Procedure Step - Push SOP Class

Workflow Management

1.2.840.10008.5.1.4.34.6.2

Unified Procedure Step - Watch SOP Class

Workflow Management

1.2.840.10008.5.1.4.34.6.3

Unified Procedure Step - Pull SOP Class

Workflow Management

1.2.840.10008.5.1.4.34.6.4

Unified Procedure Step - Event SOP Class

Workflow Management

1.2.840.10008.5.1.4.34.6.5

Unified Procedure Step - Query SOP Class

Workflow Management

1.2.840.10008.5.1.4.34.7

RT Beams Delivery Instruction Storage SOP Class

Transfer

1.2.840.10008.5.1.4.34.8

RT Conventional Machine Verification SOP Class

Workflow Management

1.2.840.10008.5.1.4.34.9

RT Ion Machine Verification SOP Class

Workflow Management

1.2.840.10008.5.1.4.34.10

RT Brachy Application Setup Delivery Instruction Storage SOP Class

Transfer

1.2.840.10008.5.1.4.37.1

General Relevant Patient Information Query SOP Class

Query/Retrieve

1.2.840.10008.5.1.4.37.2

Breast Imaging Relevant Patient Information Query SOP Class

Query/Retrieve

1.2.840.10008.5.1.4.37.3

Cardiac Relevant Patient Information Query SOP Class

Query/Retrieve

1.2.840.10008.5.1.4.38.1

Hanging Protocol Storage SOP Class

Transfer

1.2.840.10008.5.1.4.38.2

Hanging Protocol Information Model - FIND SOP Class

Query/Retrieve

1.2.840.10008.5.1.4.38.3

Hanging Protocol Information Model - MOVE SOP Class

Query/Retrieve

1.2.840.10008.5.1.4.39.1

Color Palette Storage SOP Class

Transfer

1.2.840.10008.5.1.4.39.2

Color Palette Information Model - FIND SOP Class

Query/Retrieve

1.2.840.10008.5.1.4.39.3

Color Palette Information Model - MOVE SOP Class

Query/Retrieve

1.2.840.10008.5.1.4.39.4

Color Palette Information Model - GET SOP Class

Query/Retrieve

1.2.840.10008.5.1.4.41

Product Characteristics Query SOP Class

Query/Retrieve

1.2.840.10008.5.1.4.42

Substance Approval Query SOP Class

Query/Retrieve

1.2.840.10008.5.1.4.43.1

Generic Implant Template Storage SOP Class

Transfer

1.2.840.10008.5.1.4.43.2

Generic Implant Template Information Model - FIND SOP Class

Query / Retrieve

1.2.840.10008.5.1.4.43.3

Generic Implant Template Information Model - MOVE SOP Class

Query / Retrieve

1.2.840.10008.5.1.4.43.4

Generic Implant Template Information Model - GET SOP Class

Query / Retrieve

1.2.840.10008.5.1.4.44.1

Implant Assembly Template Storage SOP Class

Transfer

1.2.840.10008.5.1.4.44.2

Implant Assembly Template Information Model - FIND SOP Class

Query / Retrieve

1.2.840.10008.5.1.4.44.3

Implant Assembly Template Information Model - MOVE SOP Class

Query / Retrieve

1.2.840.10008.5.1.4.44.4

Implant Assembly Template Information Model - GET SOP Class

Query / Retrieve

1.2.840.10008.5.1.4.45.1

Implant Template Group Storage SOP Class

Transfer

1.2.840.10008.5.1.4.45.2

Implant Template Group Information Model - FIND SOP Class

Query / Retrieve

1.2.840.10008.5.1.4.45.3

Implant Template Group Information Model - MOVE SOP Class

Query / Retrieve

1.2.840.10008.5.1.4.45.4

Implant Template Group Information Model - GET SOP Class

Query / Retrieve


A table of Supported Media Storage Application Profiles (with roles) is provided, organized in categories:

Table A.1-3. Media Services

Media Storage Application Profile

Write Files (FSC or FSU)

Read Files (FSR)

Compact Disk - Recordable

General Purpose CD-R

Option

Yes

Magneto-Optical Disk

CT/MR 2.3 GB MOD

Yes

Yes

DVD

General Purpose DVD-RAM

Yes

Yes

BD

General Purpose BD Interchange with MPEG-4 AVC/H.264 BD-Compatible HiP@Level4.1

Yes

Yes

USB and Flash Memory

General Purpose USB Media Interchange with JPEG

Yes

Yes

Email

General Purpose MIME Interchange

Yes

No

General Purpose ZIP Email

Yes

No


DICOM PS3.2 2022a - Conformance