DICOM PS3.3 2020b - Information Object Definitions |
---|
This section describes the specific Modules for the Enhanced MR Image IOD.
This section describes the Enhanced MR Image Module.
Table C.8-79 specifies the Attributes of the Enhanced MR Image Module.
Table C.8-79. Enhanced MR Image Module Attributes
Include Table C.8-83 “MR Image and Spectroscopy Instance Macro Attributes” |
|||
Image characteristics. See Section C.8.16.1 and Section C.8.13.1.1.1. |
|||
Include Table C.8-131 “Common CT/MR Image Description Macro Attributes” |
|||
Include Table C.8-84 “MR Image Description Macro Attributes” |
|||
Number of samples (planes) in this image. For Enumerated Values See Section C.8.13.1.1.2. |
|||
Specifies the intended interpretation of the pixel data. Enumerated Values are specified in the IOD that invokes this Module. See Section C.7.6.3.1.2 for definition of this term. |
|||
Number of bits allocated for each pixel sample. Each sample shall have the same number of bits allocated. For Enumerated Values See Section C.8.13.1.1.2. |
|||
Number of bits stored for each pixel sample. Each sample shall have the same number of bits stored. For Enumerated Values See Section C.8.13.1.1.2. |
|||
Most significant bit for pixel sample data. Each sample shall have the same high bit. Shall be one less than the value in Bits Stored (0028,0101). |
|||
Data representation of the pixel samples. Each sample shall have the same pixel representation. For Enumerated Values See Section C.8.13.1.1.2 |
|||
Indicates whether the pixel data are encoded color-by-plane or color-by-pixel. Required if Samples per Pixel (0028,0002) has a value greater than 1. See Section C.7.6.3.1.3 and Section C.8.13.1.1.2 for further explanation. |
|||
Indicates whether or not the image contains sufficient burned in annotation to identify the patient and date the image was acquired. This means that images that contain this Module shall not contain such burned in annotations. Required if SOP Class UID is not "1.2.840.10008.5.1.4.1.1.4.4" (Legacy Converted), may be present otherwise. |
|||
Indicates whether or not the image contains sufficiently recognizable visual features to allow the image or a reconstruction from a set of images to identify the patient. If this Attribute is absent, then the image may or may not contain recognizable visual features. |
|||
Specifies whether an Image has undergone lossy compression (at a point in its lifetime). Enumerated Values: Once this value has been set to 01 it shall not be reset. See Section C.7.6.1.1.5. Required if SOP Class UID is not "1.2.840.10008.5.1.4.1.1.4.4" (Legacy Converted). May be present otherwise. |
|||
Describes the approximate lossy compression ratio(s) that have been applied to this image. |
|||
A label for the lossy compression method(s) that have been applied to this image. |
|||
Specifies an identity transformation for the Presentation LUT, such that the output of all grayscale transformations defined in the IOD containing this Module are defined to be P-Values. Required if Photometric Interpretation (0028,0004) is MONOCHROME2. |
|||
See Section C.7.6.1.1.6. |
|||
Include Table 10-25 “Optional View and Slice Progression Direction Macro Attributes” |
Image Type (0008,0008) and Frame Type (0008,9007) are not included in this Macro but one or the other is always included in the Module or Macro that invokes this Macro, and they are therefore described here.
In addition to the requirements specified in Section C.8.16.1 Image Type and Frame Type, the following additional requirements and Defined Terms are specified.
Value 1 of Image Type (0008,0008) and Frame Type (0008,9007) is discussed in Section C.8.16.1.1. No additional requirements or Defined Terms.
Value 2 of Image Type (0008,0008) and Frame Type (0008,9007) is discussed in Section C.8.16.1.2. No additional requirements or Defined Terms.
Table C.8-80 specifies the Defined Terms for MR additional to those defined in Section C.8.16.1.3 for Value 3 for Image Type (0008,0008) and Frame Type (0008,9007).
Table C.8-80. MR-Specific Image Type and Frame Type Value 3
Table C.8-81 specifies the Defined Terms for MR additional to those defined in Section C.8.16.1.4 for Value 4 for Image Type (0008,0008) and Frame Type (0008,9007).
Table C.8-82 specifies the Enumerated Values and allowed combinations of Samples per Pixel (0028,0002), Planar Configuration (0028,0006), Pixel Representation (0028,0103), Bits Allocated (0028,0100) and Bits Stored (0028,0101) for each allowable Photometric Interpretation allowed by the IOD that invokes this Module.
Table C.8-82. Allowed Combinations of Attribute Values for Photometric Interpretation, Samples Per Pixel, Planar Configuration, Pixel Representation, Bits Allocated and Bits Stored
See Section C.8.16.2.1.1. No additional requirements or Defined Terms.
See Section C.8.16.2.1.2. No additional requirements or Defined Terms.
See Section C.8.16.2.1.3. No additional requirements or Defined Terms.
DICOM PS3.3 2020b - Information Object Definitions |
---|