DICOM PS3.18 2019e - Web Services

I.2.2 Failure Reason

Table I.2-2 defines the semantics for which the associated value shall be used for the Failure Reason (0008,1197). Implementation specific warning and error codes shall be defined in the conformance statement:

Table I.2-2. Store Instances Response Failure Reason Values

Status Code (hexadecimal)

Status Code (decimal)

Meaning

Explanation

A7xx

42752 - 43007

Refused out of Resources

The Studies Store Transaction (Section 10.5) did not store the instance because it was out of resources.

A9xx

43264 - 43519

Error: Data Set does not match SOP Class

The Studies Store Transaction (Section 10.5) did not store the instance because the instance does not conform to its specified SOP Class.

Cxxx

49152 - 53247

Error: Cannot understand

The Studies Store Transaction (Section 10.5) did not store the instance because it cannot understand certain Data Elements.

C122

49442

Referenced Transfer Syntax not supported

The Studies Store Transaction (Section 10.5) did not store the instance because it does not support the requested Transfer Syntax for the instance.

0110

272

Processing failure

The Studies Store Transaction (Section 10.5) did not store the instance because of a general failure in processing the operation.

0122

290

Referenced SOP Class not supported

The Studies Store Transaction (Section 10.5) did not store the instance because it does not support the requested SOP Class.


Additional codes may be used for the Failure Reason (0008,1197) to address the semantics of other issues.

In the event that multiple codes may apply, the single most appropriate code shall be used.

DICOM PS3.18 2019e - Web Services