The creator of an implementation shall document in its Conformance Statement:
the origin server and/or user agent role(s) played,
the resources supported for this transaction,
the media types supported for this transaction,
the optional Query Parameters supported,
the optional Header Fields supported.
The creator of an implementation shall also document:
The Composite SOP classes supported, including:
the Image Storage SOP classes supported,
the Image Storage SOP classes supported by Rendered Presentation States.
If Thumbnails are supported:
A high-level description of the method used to render thumbnails for the study, series, or
instance.
Note
The description could indicate, for example, whether a representative instance is chosen from
a series, and how that instance is selected, or that per-modality fixed content is used.
The minimum and maximum sizes for thumbnails.
Character sets supported for Thumbnail resources (if other than UTF-8).