DICOM PS3.16 2019a - Content Mapping Resource

TID 4010 Mammography CAD Calcification Cluster

This Template provides the detail specific to a calcification cluster.

Type:

Non-Extensible

Order:

Significant

Root:

No

Table TID 4010. Mammography CAD Calcification Cluster

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

EV (111009, DCM, "Calcification Type")

1-n

MC

At least one of rows 1, 2, 3, 4, 5 shall be present

DCID 6010 “Mammography Calcification Types”

2

CODE

EV (111008, DCM, "Calcification Distribution")

1

MC

At least one of rows 1, 2, 3, 4, 5 shall be present

DCID 6012 “Calcification Distribution Modifier”

3

NUM

EV (111038, DCM, "Number of calcifications")

1

MC

At least one of rows 1, 2, 3, 4, 5 shall be present

UNITS = EV (1, UCUM, "no units")

Value = Integer > = 1

4

INCLUDE

DTID 1400 “Linear Measurement”

1-n

MC

At least one of rows 1, 2, 3, 4, 5 shall be present

If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID 1400 “Linear Measurement” shall be used.

5

INCLUDE

DTID 1401 “Area Measurement”

1-n

MC

At least one of rows 1, 2, 3, 4, 5 shall be present

If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID 1401 “Area Measurement” shall be used.

6

INCLUDE

DTID 1402 “Volume Measurement”

1-n

U

If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID 1402 “Volume Measurement” shall be used.


DICOM PS3.16 2019a - Content Mapping Resource