DICOM PS3.16 2019a - Content Mapping Resource |
---|
This Template records details related to the use of the NEMA Dose Check Standard (NEMA XR-25-2010).
Table TID 10015. CT Dose Check Details
IFF value of Row 2 is (R-0038D, SRT, "Yes") |
||||||||
IFF value of Row 3 is (R-0038D, SRT, "Yes") |
||||||||
IF Accumulated DLP Forward Estimate (Row 6) exceeds DLP Alert Value (Row 4) |
||||||||
IF Accumulated CTDIvol Forward Estimate (Row 7) exceeds CTDIvol Alert Value (Row 5) |
||||||||
IFF Accumulated DLP Forward Estimate (Row 6) exceeds DLP Alert Value (Row 4) or Accumulated CTDIvol Forward Estimate (Row 7) exceeds CTDIvol Alert Value (Row 5) |
||||||||
IF Accumulated DLP Forward Estimate (Row 6) exceeds DLP Alert Value (Row 4) or Accumulated CTDIvol Forward Estimate (Row 7) exceeds CTDIvol Alert Value (Row 5) |
$PersonProcedureRole = EV (113850, DCM, "Irradiation Authorizing") |
|||||||
IFF value of Row 11 is (R-0038D, SRT, "Yes") |
||||||||
IFF value of Row 12 is (R-0038D, SRT, "Yes") |
||||||||
IF DLP Forward Estimate (Row 15) exceeds DLP Notification Value (Row 13) |
||||||||
IF CTDIvol Forward Estimate (Row 16) exceeds CTDIvol Notification Value (Row 14) |
||||||||
IFF DLP Forward Estimate (Row 15) exceeds DLP Notification Value (Row 13) or CTDIvol Forward Estimate (Row 16) exceeds CTDIvol Notification Value (Row 14) |
||||||||
IFF DLP Forward Estimate (Row 15) exceeds DLP Notification Value (Row 13) or CTDIvol Forward Estimate (Row 16) exceeds CTDIvol Notification Value (Row 14) |
$PersonProcedureRole = EV (113850, DCM, "Irradiation Authorizing") |
DICOM PS3.16 2019a - Content Mapping Resource |
---|