DICOM PS3.11 2019a - Media Storage Application Profiles

M.3 STD-GEN-BD and STD-GEN-SEC-BD Profile Classes

M.3.1 SOP Classes and Transfer Syntaxes

This Application Profile is based on the Media Storage Service Class with the Interchange Option (see PS3.4).

Table M.3-1. STD-GEN-BD and STD-GEN-SEC-BD SOP Classes and Transfer Syntaxes

Information Object Definition

SOP Class UID

Transfer Syntax and UID

FSC Requirement

FSR Requirement

FSU Requirement

Basic Directory

1.2.840.10008.1.3.10

Explicit VR Little Endian Uncompressed

1.2.840.10008.1.2.1

Mandatory

Mandatory

Mandatory

Composite IODs for which a Media Storage SOP Class is defined in PS3.4

See PS3.4

Explicit VR Little Endian Uncompressed

1.2.840.10008.1.2.1

Defined in Conformance Statement

Mandatory for all SOP Classes defined in Conformance Statement

Defined in Conformance Statement

Composite IODs for which a Media Storage SOP Class is defined in PS3.4

See PS3.4

JPEG Lossless Process 14 (selection value 1)

1.2.840.10008.1.2.4.70

Defined in Conformance Statement

Mandatory for JPEG profiles for all SOP Classes defined in Conformance Statement

Defined in Conformance Statement

Composite IODs for which a Media Storage SOP Class is defined in PS3.4

See PS3.4

JPEG Lossy, Baseline Sequential with Huffman Coding (Process 1)

1.2.840.10008.1.2.4.50

Defined in Conformance Statement

Mandatory for JPEG profiles for all SOP Classes defined in Conformance Statement

Defined in Conformance Statement

Composite IODs for which a Media Storage SOP Class is defined in PS3.4

See PS3.4

JPEG Extended (Process 2 & 4):

Default Transfer Syntax for Lossy JPEG 12 Bit Image Compression (Process 4 only)

1.2.840.10008.1.2.4.51

Defined in Conformance Statement

Mandatory for JPEG profiles for all SOP Classes defined in Conformance Statement

Defined in Conformance Statement

Composite IODs for which a Media Storage SOP Class is defined in PS3.4

See PS3.4

JPEG 2000 Image Compression (Lossless Only)

1.2.840.10008.1.2.4.90

Defined in Conformance Statement

Mandatory for J2K profiles for all SOP Classes defined in Conformance Statement

Defined in Conformance Statement

Composite IODs for which a Media Storage SOP Class is defined in PS3.4

See PS3.4

JPEG 2000 Image Compression

1.2.840.10008.1.2.4.91

Defined in Conformance Statement

Mandatory for J2K profiles for all SOP Classes defined in Conformance Statement

Defined in Conformance Statement

Multi-frame Composite IODs for which a Media Storage SOP Class is defined in PS3.4

See PS3.4

MPEG2 Main Profile @ Main Level

1.2.840.10008.1.2.4.100

Defined in Conformance Statement

Mandatory for all SOP Classes defined in Conformance Statement

Defined in Conformance Statement

Multi-frame Composite IODs for which a Media Storage SOP Class is defined in PS3.4

See PS3.4

MPEG2 Main Profile @ High Level

1.2.840.10008.1.2.4.101

Defined in Conformance Statement

Mandatory for all SOP Classes defined in Conformance Statement

Defined in Conformance Statement

Multi-frame Composite IODs for which a Media Storage SOP Class is defined in PS3.4

See PS3.4

MPEG-4 AVC/H.264 High Profile / Level 4.1

1.2.840.10008.1.2.4.102

Defined in Conformance Statement

Mandatory for all SOP Classes defined in Conformance Statement

Defined in Conformance Statement

Multi-frame Composite IODs for which a Media Storage SOP Class is defined in PS3.4

See PS3.4

MPEG-4 AVC/H.264 BD-compatible High Profile / Level 4.1

1.2.840.10008.1.2.4.103

Defined in Conformance Statement

Mandatory for all SOP Classes defined in Conformance Statement

Defined in Conformance Statement


The SOP Classes and corresponding Transfer Syntax supported by this Application Profile are specified in the Table M.3-1. The supported Storage SOP Class(es) shall be listed in the Conformance Statement using a table of the same form.

DICOM PS3.11 2019a - Media Storage Application Profiles