DICOM PS3.11 2019a - Media Storage Application Profiles

E.3 STD-CTMR Profiles

E.3.1 SOP Classes and Transfer Syntaxes

These Application Profiles are based on the Media Storage Service Class (see PS3.4).

SOP Classes and corresponding Transfer Syntaxes supported by these Application Profiles are specified in the Table E.3-1.

Table E.3-1. STD-CTMR SOP Classes and Transfer Syntaxes

Information Object Definition

SOP Class UID

Transfer Syntax and UID

FSC Requirement

FSR Requirement

FSU Requirement (see Note 1)

Basic Directory

1.2.840.10008.1.3.10

Explicit VR Little Endian Uncompressed

1.2.840.10008.1.2.1

Mandatory

Mandatory

Mandatory

CT Image

1.2.840.10008.5.1.4.1.1.2

JPEG Lossless Process 14 (selection value 1)

1.2.840.10008.1.2.4.70

Optional

Mandatory

Optional

CT Image

1.2.840.10008.5.1.4.1.1.2

Explicit VR Little Endian Uncompressed

1.2.840.10008.1.2.1

Optional

Mandatory

Optional

MR Image

1.2.840.10008.5.1.4.1.1.4

JPEG Lossless Process 14 (selection value 1)

1.2.840.10008.1.2.4.70

Optional

Mandatory

Optional

MR Image

1.2.840.10008.5.1.4.1.1.4

Explicit VR Little Endian Uncompressed

1.2.840.10008.1.2.1

Optional

Mandatory

Optional

SC Image (grayscale)

1.2.840.10008.5.1.4.1.1.7

JPEG Lossless Process 14 (selection value 1)

1.2.840.10008.1.2.4.70

Optional

Mandatory

Optional

SC Image (grayscale)

1.2.840.10008.5.1.4.1.1.7

Explicit VR Little Endian Uncompressed

1.2.840.10008.1.2.1

Optional

Mandatory

Optional

SC Image(palette color)

1.2.840.10008.5.1.4.1.1.7

JPEG Lossless Process 14 (selection value 1)

1.2.840.10008.1.2.4.70

Optional

Optional

Optional

SC Image(palette color)

1.2.840.10008.5.1.4.1.1.7

Explicit VR Little Endian Uncompressed

1.2.840.10008.1.2.1

Optional

Optional

Optional

Grayscale Softcopy Presentation State

1.2.840.10008.5.1.4.1.1.11.1

Explicit VR Little Endian Uncompressed

1.2.840.10008.1.2.1

Optional

Optional

Optional

X-Ray Radiation Dose SR

1.2.840.10008.5.1.4.1.1.88.67

Explicit VR Little Endian Uncompressed

1.2.840.10008.1.2.1

Optional

Optional

Optional


Note

  1. The FSU requirement is not defined for STD-CTMR-DVD profile.

  2. The Detached Patient management SOP Class was formerly defined in these profiles, but has been retired.

DICOM PS3.11 2019a - Media Storage Application Profiles