DICOM PS3.11 2019a - Media Storage Application Profiles

C.3 General Class Profile

C.3.1 Abstract and Transfer Syntaxes

Application Profiles in this class, STD-US, shall support the appropriate Information Object Definitions (IOD) and Transfer Syntaxes for the Media Storage SOP Class in the following table. In the role of FS-Updater or FS-Creator the application can choose one of the three possible transfer Syntaxes to create an IOD. In the role of FS-Reader an application shall support all transfer Syntaxes defined for the STD-US application profile.

Table C.3-1. Ultrasound SOP Classes and Transfer Syntaxes

Information Object Definition

SOP Class UID

Transfer Syntax

Transfer Syntax UID

DICOM Media Storage Directory

1.2.840.10008.1.3.10

Explicit VR Little Endian Uncompressed

1.2.840.10008.1.2.1 (see Section 8.6 in PS3.10 )

Ultrasound Image Storage

1.2.840.10008.5.1.4.1.1.6.1

Explicit VR Little Endian Uncompressed

1.2.840.10008.1.2.1

Ultrasound Image Storage

1.2.840.10008.5.1.4.1.1.6.1

RLE Lossless Image Compression

1.2.840.10008.1.2.5

Ultrasound Image Storage

1.2.840.10008.5.1.4.1.1.6.1

JPEG Lossy, Baseline Sequential with Huffman Coding (Process 1)

1.2.840.10008.1.2.4.50

Ultrasound Multi-frame Image Storage

1.2.840.10008.5.1.4.1.1.3.1

Explicit VR Little Endian Uncompressed

1.2.840.10008.1.2.1

Ultrasound Multi-frame Image Storage

1.2.840.10008.5.1.4.1.1.3.1

RLE Lossless Image Compression

1.2.840.10008.1.2.5

Ultrasound Multi-frame Image Storage

1.2.840.10008.5.1.4.1.1.3.1

JPEG Lossy, Baseline Sequential with Huffman Coding (Process 1)

1.2.840.10008.1.2.4.50


C.3.1.1 Ultrasound Single and Multi-frame Pixel Formats Supported

The STD-US application profile requires that all ultrasound image objects only be stored using the values described in PS3.3 US Image Module and the specializations used for the Ultrasound Single and Multi-Frame IODs.

In the role of FS-Updater or FS-Creator the application can choose any of the supported Photometric Interpretations described in PS3.3 US Image Module to create an IOD. In the role of FS-Reader, an application shall support all Photometric Interpretations described in PS3.3 US Image Module.

Table C.3-2 describes restrictions on the use of various Transfer Syntaxes with the supported Photometric Interpretations for both single and multi-frame images.

Table C.3-2. Defined Photometric Interpretation and Transfer Syntax Pairs

Photometric Interpretation Value

Transfer Syntax

Transfer Syntax UID

MONOCHROME2

Uncompressed

RLE Lossless Image Compression

1.2.840.10008.1.2.1

1.2.840.10008.1.2.5

RGB

Uncompressed

RLE Lossless Image Compression

1.2.840.10008.1.2.1

1.2.840.10008.1.2.5

PALETTE COLOR

Uncompressed

RLE Lossless Image Compression

1.2.840.10008.1.2.1

1.2.840.10008.1.2.5

YBR_FULL

RLE Lossless Image Compression

1.2.840.10008.1.2.5

YBR_FULL_422

Uncompressed

JPEG Lossy

1.2.840.10008.1.2.1

1.2.840.10008.1.2.4.50


DICOM PS3.11 2019a - Media Storage Application Profiles