DICOM PS3.11 2019a - Media Storage Application Profiles

B.3 STD-XA1K Application Profile Class Requirements

B.3.1 SOP Classes and Transfer Syntaxes

This Application Profile Class is based on the Media Storage Service Class (see PS3.4).

SOP Classes and corresponding Transfer Syntaxes supported by this Application Profile are specified in Table B.3-1.

Table B.3-1. STD-XA1K SOP Classes and Transfer Syntaxes

Information Object Definition

SOP Class UID

Transfer Syntax and UID

FSC Requirement

FSR Requirement

FSU Requirement (see Note 1)

Basic Directory

1.2.840.10008.1.3.10

Explicit VR Little Endian Uncompressed

1.2.840.10008.1.2.1

Mandatory

Mandatory

Mandatory

X-Ray Angiographic Image

1.2.840.10008.5.1.4.1.1.12.1

JPEG Lossless Process 14 (selection value 1)

1.2.840.10008.1.2.4.70

Mandatory

Mandatory

Optional

X-Ray Angiographic Image

1.2.840.10008.5.1.4.1.1.12.1

JPEG Lossy, Baseline Sequential with Huffman Coding (Process 1)

1.2.840.10008.1.2.4.50

Optional for DVD; Disallowed for CD

Mandatory for DVD; Disallowed for CD

Undefined for DVD; Disallowed for CD

X-Ray Angiographic Image

1.2.840.10008.5.1.4.1.1.12.1

JPEG Extended (Process 2 & 4):

Default Transfer Syntax for Lossy JPEG 12 Bit Image Compression (Process 4 only)

1.2.840.10008.1.2.4.51

Optional for DVD; Disallowed for CD

Mandatory for DVD; Disallowed for CD

Undefined for DVD; Disallowed for CD

Secondary Capture Image Storage

1.2.840.10008.5.1.4.1.1.7

Explicit VR Little Endian Uncompressed

1.2.840.10008.1.2.1

Optional

Mandatory

Optional

Grayscale Softcopy Presentation State Storage

1.2.840.10008.5.1.4.1.1.11.1

Explicit VR Little Endian Uncompressed

1.2.840.10008.1.2.1

Optional

Optional

Optional


Note

  1. The FSU requirement is not defined for STD-XA1K-DVD profile.

  2. The Standalone Overlay, Standalone Curve and Detached Patient management SOP Classes were formerly defined in these profiles, but have been retired. The Grayscale Softcopy Presentation State Storage SOP Class has been added as the preferred mechanism for conveying annotations.