DICOM PS3.3 2019a - Information Object Definitions

C.8.8.19 Measured Dose Reference Record Module

Table C.8-55. Measured Dose Reference Record Module Attributes

Attribute Name

Tag

Type

Attribute Description

Measured Dose Reference Sequence

(3008,0010)

1

Sequence of doses measured during treatment delivery, summed over entire session.

One or more Items shall be included in this Sequence.

>Referenced Dose Reference Number

(300C,0051)

1C

Uniquely identifies Dose Reference specified by Dose Reference Number (300A,0012) in Dose Reference Sequence (300A,0010) in RT Prescription Module of referenced RT Plan. Required only if Measured Dose Reference Number (3008,0064) is not present. It shall not be present otherwise.

>Measured Dose Reference Number

(3008,0064)

1C

Unique identifier of measured dose point. Required only if Referenced Dose Reference Number (300C,0051) is not present. It shall not be present otherwise.

>Dose Units

(3004,0002)

1

Units used to describe measured dose.

Enumerated Values:

GY

Gray

RELATIVE

Dose relative to implicit reference value

>Measured Dose Value

(3008,0016)

2

Measured Dose in units specified by Dose Units (3004,0002).

>Measured Dose Type

(3008,0014)

2

Type of dose measurement.

Defined Terms:

DIODE

semiconductor diode

TLD

thermoluminescent dosimeter

ION_CHAMBER

ion chamber

GEL

dose sensitive gel

EPID

electronic portal imaging device

FILM

dose sensitive film

>Measured Dose Description

(3008,0012)

3

User-defined description of Dose Reference (e.g., "Exit dose", "Point A").


DICOM PS3.3 2019a - Information Object Definitions