DICOM PS3.3 2019a - Information Object Definitions

C.8.27.6 Intravascular OCT Functional Group Macros

The following sections contain Functional Group Macros specific to the Intravascular OCT IOD.

C.8.27.6.1 Intravascular OCT Frame Type Macro

Table C.8.27.6.1-1 specifies the Attributes of the Intravascular OCT Frame Type Functional Group.

Table C.8.27.6.1-1. Intravascular OCT Frame Type Macro Attributes

Attribute Name

Tag

Type

Attribute Description

Intravascular OCT Frame Type Sequence

(0052,0025)

1

Identifies characteristics of this frame.

Only a single Item shall be included in this Sequence.

>Frame Type

(0008,9007)

1

Type of Frame. A multi-valued Attribute analogous to Image Type (0008,0008).

Enumerated Values and Defined Terms are the same as those for the four values of Image Type (0008,0008), except that the value MIXED is not allowed.

See Section C.8.27.2.1.


DICOM PS3.3 2019a - Information Object Definitions