DICOM PS3.3 2019a - Information Object Definitions

C.13.4 Basic Film Box Relationship Module

Table C.13-4. Basic Film Box Relationship Module Attributes

Attribute Name

Tag

Attribute Description

Referenced Film Session Sequence

(2010,0500)

A Film Session.

Only a single Item shall be included in this Sequence.

>Include Table 10-11 “SOP Instance Reference Macro Attributes”

Referenced Image Box Sequence

(2010,0510)

A set of Image Boxes.

One or more Items shall be included in this Sequence.

>Include Table 10-11 “SOP Instance Reference Macro Attributes”

Referenced Basic Annotation Box Sequence

(2010,0520)

A set of Basic Annotation Boxes.

Zero or more Items shall be included in this Sequence.

>Include Table 10-11 “SOP Instance Reference Macro Attributes”

Referenced Presentation LUT Sequence

(2050,0500)

A Presentation LUT.

Only a single Item shall be included in this Sequence.

>Include Table 10-11 “SOP Instance Reference Macro Attributes”


DICOM PS3.3 2019a - Information Object Definitions