DICOM PS3.3 2019a - Information Object Definitions

10.9.2 Extended Content Identification Macro

The Extended Content Identification Macro identifies content using a label supporting lower case characters and specified character sets. If a Code String is required, see Section 10.9 “Content Identification Macro”.

Table 10.9.2-1. Extended Content Identification Macro Attributes

Attribute Name

Tag

Type

Attribute Description

User Content Long Label

(3010,0034)

1

User-defined label for the content of this SOP Instance.

See Section 10.9.2.1.1.

Content Description

(0070,0081)

2

User-defined description for the content of this SOP Instance.

See Section 10.9.2.1.1.

Content Creator's Name

(0070,0084)

2

Name of operator (such as a technologist or physician) creating the content of the SOP Instance.

Content Creator's Identification Code Sequence

(0070,0086)

3

Identification of the person who created the content.

Only a single Item is permitted in this Sequence.

>Include Table 10-1 “Person Identification Macro Attributes Description”


10.9.2.1 Extended Content Identification Macro Attribute Descriptions

10.9.2.1.1 User Content Long Label and Content Description

User Content Long Label (3010,0034) shall represent a user-defined free text providing the primary identification of this entity to other users. Content Description (0070,0081) allows a longer string containing additional descriptive identifying text.

This information is intended for display to human readers. Shall not be used for structured processing.

DICOM PS3.3 2019a - Information Object Definitions