DICOM PS3.16 2016e - Content Mapping Resource |
---|
This Template describes a single image finding for a lesion or other object. The details of the finding are expressed in this Template and/or more specific Templates.
Table TID 4104. Chest CAD Single Image Finding
Shall be present IFF value of row 1 is (112005, DCM, "Radiographic anatomy") |
||||||||
May be present IFF value of row 1 is (111102, DCM, "Non-lesion") |
||||||||
IFF value of row 6 is (111151, DCM, "Presentation Optional") and row 1 of TID 4023 “CAD Operating Points” is present for the finding identified in row 1 |
UNITS = DT ({1:n}, UCUM, "range: 1:n"), where n is the maximum specified in Row 1 of TID 4023 “CAD Operating Points” for the finding identified in row 1. Value is restricted to being an integer |
|||||||
Shall be present IFF this finding is duplicated from a different report than its parent. |
||||||||
Shall be present IFF value of row 1 is (111099, DCM, "Selected region") |
||||||||
Shall be present unless value of row 1 is (111101, DCM, "Image quality") |
||||||||
Shall be present IFF value of row 1 is (111101, DCM, "Image quality") and rows 20 and 21 are not present |
||||||||
Shall be present IFF value of row 1 is (111101, DCM, "Image quality") and rows 19 and 21 are not present |
Shall reference an IMAGE Content Item in the (111028, DCM, "Image Library") |
|||||||
Shall be present IFF value of row 1 is (111101, DCM, "Image quality") and rows 19 and 20 are not present |
||||||||
All the row 21 Content Items in a single invocation of this Template shall reference the same IMAGE |
||||||||
All the row 21 Content Items in a single invocation of this Template shall reference the same IMAGE Content Item in the (111028, DCM, "Image Library") |
||||||||
Shall be present IFF value of row 1 is (111101, DCM, "Image quality") |
$QualityFinding = DCID 6135 “Image Quality Finding” $QualityStandard = DCID 6136 “Chest Types of Quality Control Standard” |
An identifier of an anatomic feature when a multiplicity of features of that type may be present, such as "Rib 1", "Rib 2" or thoracic vertebrae "T1" or "T2". |
|
This Content Item constrains the SCP receiving the Chest CAD SR IOD in its use of the contents of this Template and its Target Content Items. Chest CAD devices may opt to use data marked "Not for Presentation" or "Presentation Optional" as input to subsequent chest CAD processing steps. Refer to Section O.2 “Structured Reporting Storage SOP Class SCU and SCP Behavior” in PS3.4 . |
|
Additional information to use when Rendering Intent is "Presentation Optional". A CAD Operating Point of zero is not sent, and is encoded as a Rendering Intent of "Presentation Required". See Section O.2 “Structured Reporting Storage SOP Class SCU and SCP Behavior” in PS3.4 and Section E.4 “CAD Operating Point” in PS3.17 . |
|
The certainty of the CAD device that the finding detected and classified by the Single Image Finding CODE specified is in fact that type of finding. |
DICOM PS3.16 2016e - Content Mapping Resource |
---|