DICOM PS3.16 2016e - Content Mapping Resource

TID 3504 Arterial Pressure Measurement

The Arterial Pressure Measurement Template consists of a CONTAINER, with an acquisition context of the measurement anatomic site, and containing systolic, diastolic, and mean measurements. This implies that the name of the measurement is effectively post-coordinated from the measurement name, the Hemodynamic Measurement Group container (procedure phase) name, and the acquisition context (finding site) value.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3504. Arterial Pressure Measurement

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (P2-36102, SRT, "Arterial pressure measurements")

1

M

2

>

INCLUDE

DTID 3530 “Hemodynamic Acquisition Context”

1

M

$LocationName = EV (G-C0E3, SRT, "Finding Site")

$LocationValue = DCID 3606 “Arterial Source Locations”

3

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (8480-6, LN, "Intravascular arterial Systolic pressure")

$Units = DCID 3500 “Pressure Units”

4

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (8462-4, LN, "Intravascular arterial Diastolic pressure")

$Units = DCID 3500 “Pressure Units”

5

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (8478-0, LN, "Intravascular arterial mean pressure")

$Units = DCID 3500 “Pressure Units”

6

>

CONTAINS

INCLUDE

DTID 3550 “Pressure Waveform Measurements”

1-n

U


DICOM PS3.16 2016e - Content Mapping Resource