Table CC.2.8-2 defines the status code values that might be returned in a C-FIND response. Fields related to status code values are defined in PS3.7.
Table CC.2.8-2. C-FIND Response Status Values
Service Status
|
Further Meaning
|
Status Codes
|
Related Fields
|
Failure
|
Refused: Out of Resources
|
A700
|
(0000,0902)
|
Identifier Does Not Match SOP Class
|
A900
|
(0000,0901)
(0000,0902)
|
SOP Class not Supported
|
0122
|
|
Unable to process
|
(any value C000 through CFFF as assigned by the implementation)
|
(0000,0901)
(0000,0902)
|
Cancel
|
Matching terminated due to Cancel request
|
FE00
|
None
|
Success
|
Matching is complete - No final Identifier is supplied.
|
0000
|
None
|
Pending
|
Matches are continuing - Current Match is supplied and any Optional Keys were supported in the same manner as Required Keys.
|
FF00
|
Identifier
|
Matches are continuing - Warning that one or more Optional Keys were not supported for existence for this Identifier.
|
FF01
|
Identifier
|
Note
Status Codes are returned in DIMSE response messages (see PS3.7). The code values stated in column "Status Codes" are returned in Status Command Element (0000,0900).