DICOM PS3.4 2016c - Service Class Specifications |
---|
The Implant Template Information Model consists of a single entity. In response to a given C-FIND request, the SCP shall send one C-FIND response per matching Implant Template Instance.
The Implant Assembly Template Information Model consists of a single entity. In response to a given C-FIND request, the SCP shall send one C-FIND response per matching Implant Assembly Template Instance.
The Implant Template Group Information Model consists of a single entity. In response to a given C-FIND request, the SCP shall send one C-FIND response per matching Implant Template Group Instance.
Table BB.6-1 defines the Attributes of the Generic Implant Template Information Model:
Table BB.6-1. Attributes for the Implant Template Information Model
This Attribute is required if expanded or replacement character sets are used. See Section C.2.2.2 and Section C.4.1.1. |
||||
Shall be retrieved with Single Value, Wild Card, or Universal Matching. |
||||
Shall be retrieved with Single Value, Wild Card, or Universal Matching. |
||||
Shall be retrieved with Single Value, Wild Card, or Universal Matching. |
||||
Shall be retrieved with Single Value, Wild Card, or Universal Matching. |
||||
This Attribute shall be retrieved with Sequence or Universal matching. |
||||
This Attribute shall be retrieved with Sequence or Universal matching. |
||||
This Attribute shall be retrieved with Sequence or Universal matching. |
||||
This Attribute shall be retrieved with Sequence or Universal matching. |
||||
This Attribute shall be retrieved with Single Value or Universal matching. |
||||
This Attribute shall be retrieved with Single Value or Universal matching. |
||||
This Attribute shall be retrieved with Single Value or Universal matching. |
||||
This Attribute shall be retrieved with Sequence or Universal matching. |
||||
This Attribute shall be retrieved with Single Value or Universal matching. |
||||
This Attribute shall be retrieved with Single Value or Universal matching. |
||||
This Attribute shall be retrieved with Sequence or Universal matching. |
||||
This Attribute shall be retrieved with Single Value or Universal matching. |
||||
This Attribute shall be retrieved with Single Value or Universal matching. |
||||
This Attribute shall be retrieved with Sequence or Universal matching. |
||||
This Attribute shall be retrieved with Single Value or Universal matching. |
||||
This Attribute shall be retrieved with Single Value or Universal matching. |
||||
Table BB.6-2 defines the Attributes of the Implant Assembly Template Information Model:
Table BB.6-2. Attributes for the Implant Assembly Template Information Model
This Attribute is required if expanded or replacement character sets are used. See Section C.2.2.2 and Section C.4.1.1. |
||||
Shall be retrieved with Single Value, Wild Card, or Universal Matching. |
||||
Shall be retrieved with Single Value, Wild Card, or Universal Matching. |
||||
This Attribute shall be retrieved with Single Value or Universal matching. |
||||
This Attribute shall be retrieved with Single Value or Universal matching. |
||||
This Attribute shall be retrieved with Single Value or Universal matching. |
||||
Shall be retrieved with Single Value, Wild Card, or Universal Matching. |
Table BB.6-3 defines the Attributes of the Implant Template Group Information Model:
Table BB.6-3. Attributes for the Implant Template Group Information Model
This Attribute is required if expanded or replacement character sets are used. See Section C.2.2.2 and Section C.4.1.1. |
||||
Shall be retrieved with Single Value, Wild Card, or Universal Matching. |
||||
Shall be retrieved with Single Value, Wild Card, or Universal Matching. |
||||
An implementation may conform to one of the Implant Template, Implant Assembly Template, or Implant Template Group Information Model SOP Classes as an SCU, SCP, any combination of two of these, or all three. The Conformance Statement shall be in the format defined in PS3.2.
An implementation that conforms to one of the Implant Template, Implant Assembly Template, or Implant Template Group Information Model SOP Classes shall support queries against the appropriate Information Model using the C-FIND SCU Behavior described for the Basic Worklist Management Service Class (see Section K.4.1.2 and Section BB.4.1).
An implementation that conforms to one of the Implant Template, Implant Assembly Template, or Implant Template Group Information Model SOP Classes as an SCU shall state in its Conformance Statement whether it requests Type 3 Return Key Attributes, and shall list these Optional Return Key Attributes.
An implementation that conforms to one of the Implant Template, Implant Assembly Template, or Implant Template Group Information Model SOP Classes as an SCU shall state in its Conformance Statement how it makes use of Specific Character Set (0008,0005) when encoding queries and interpreting responses.
An implementation that conforms to one of the Implant Template, Implant Assembly Template, or Implant Template Group Information Model SOP Classes as an SCU shall support transfers against the appropriate Information Model, using the C-MOVE SCU baseline behavior described for the Query/Retrieve Service Class (see Section C.4.2.2.1 and Section BB.4.2).
An implementation that conforms to one of the Implant Template, Implant Assembly Template, or Implant Template Group Information Model SOP Classes as an SCU shall support transfers against the appropriate Information Model, using the C-GET SCU baseline behavior described for the Query/Retrieve Service Class (see Section C.4.3.2).
An implementation that conforms to one of the Implant Template, Implant Assembly Template, or Implant Template Group Information Model SOP Classes as an SCP shall support queries against the appropriate Template Information Model, using the C-FIND SCP Behavior described for the Basic Worklist Management Service Class (see Section K.4.1.3).
An implementation that conforms to one of the Implant Template, Implant Assembly Template, or Implant Template Group Information Model SOP Classes as an SCP shall state in its Conformance Statement whether it supports Type 3 Return Key Attributes, and shall list these Optional Return Key Attributes.
An implementation that conforms to one of the Implant Template, Implant Assembly Template, or Implant Template Group Information Model SOP Classes as an SCP shall state in its Conformance Statement how it makes use of Specific Character Set (0008,0005) when interpreting queries, performing matching and encoding responses.
An implementation that conforms to one of the Implant Template, Implant Assembly Template, or Implant Template Group Information Model SOP Classes as an SCP shall support transfers against the appropriate Information Model, using the C-MOVE SCP baseline behavior described for the Query/Retrieve Service Class (see Section C.4.2.3.1).
An implementation that conforms to one of the Implant Template, Implant Assembly Template, or Implant Template Group Information Model SOP Classes as an SCP, which generates transfers using the C-MOVE operation, shall state in its Conformance Statement appropriate Storage Service Class, under which it shall support the C-STORE sub-operations generated by the C-MOVE.
An implementation that conforms to one of the SOP Classes of the Implant Template, Implant Assembly Template, or Implant Template Group Information Model SOP Class Group as an SCP shall support retrievals against the Query/Retrieve Information Model described in Section C.6.1.1 using the C-GET SCP Behavior described in Section C.4.3.3.
The SOP Classes of the Implant Template Information Models in the Implant Template Query/Retrieve Service Class identify the Implant Template Information Models, and the DIMSE-C operations supported. The SOP Classes of the Implant Assembly Template Information Models in the Implant Assembly Template Query/Retrieve Service Class identify the Implant Assembly Template Information Models, and the DIMSE-C operations supported. The SOP Classes of the Implant Template Group Information Models in the Implant Template Group Query/Retrieve Service Class identify the Implant Template Group Information Models, and the DIMSE-C operations supported. The following Standard SOP Classes are identified:
DICOM PS3.4 2016c - Service Class Specifications |
---|