DICOM PS3.2 2016c - Conformance

F.4.2 AE Specifications

F.4.2.1 STORAGE-SCU Application Entity Specification

F.4.2.1.1 SOP Classes

The STORAGE-SCU AE provides Standard Conformance to the following DICOM V3.0 SOP Classes:

Table F.4.2-1. SOP Classes for STORAGE-SCU AE

SOP Class Name

SOP Class UID

SCU

SCP

Verification

1.2.840.10008.1.1

Yes

No

US Image Storage (Retired)

1.2.840.10008.5.1.4.1.1.6

Yes

No

US Image Storage

1.2.840.10008.5.1.4.1.1.6.1

Yes

No

US Multi-frame Storage (Retired)

1.2.840.10008.5.1.4.1.1.3

Yes

No

US Multi-frame Storage

1.2.840.10008.5.1.4.1.1.3.1

Yes

No

Computed Radiography Image Storage

1.2.840.10008.5.1.4.1.1.1

Yes

No

CT Image Storage

1.2.840.10008.5.1.4.1.1.2

Yes

No

MR Image Storage

1.2.840.10008.5.1.4.1.1.4

Yes

No

Secondary Capture Image Storage

1.2.840.10008.5.1.4.1.1.7

Yes

No


STORAGE-SCU AE can be configured to use the retired US Image objects (US Image Storage, 1.2.840.10008.5.1.4.1.1.6, and US Multi-frame Storage, 1.2.840.10008.5.1.4.1.1.3) rather than the current US SOP Classes for ultrasound images or vice-versa, making any necessary changes to make the transformed image objects conformant to the corresponding SOP Class. This is only done if the external Storage SCP AE does not support the SOP Instance's original SOP Class.

By altering the configuration it is possible to support additional or fewer SOP Classes.

F.4.2.1.2 Association Establishment Policies
F.4.2.1.2.1 General

The STORAGE-SCU AE can only form Associations when requested to do so by the QUERY-RETRIEVE-SCP AE. The STORAGE-SCU AE can only request the opening of an Association. It cannot accept requests to open Associations from external Application Entities.

The DICOM standard Application Context Name for DICOM is always proposed:

Table F.4.2-2. DICOM Application Context for STORAGE-SCU AE

Application Context Name

1.2.840.10008.3.1.1.1


F.4.2.1.2.2 Number of Associations

The maximum number of simultaneous Associations is configurable, but is usually limited to a maximum of 10. This configuration largely depends on whether relatively quick response to multiple simultaneous C-MOVE Destination AEs is required or maximum throughput performance is required. If the latter is the case, then no simultaneous Associations are permitted, in order to reduce disk thrashing and thus maximize throughput. The STORAGE-SCU AE can initiate simultaneous Associations to a given external C-MOVE Destination AE up to the maximum number configured. There is no separate limit on the maximum number permitted to the same C-MOVE Destination AE.

If the first attempt to open an Association fails then the STORAGE-SCU AE will reschedule the task to attempt it again after a configurable time delay. The number of times to reattempt Association establishment is configurable, with the default being zero.

Table F.4.2-3. Number of Associations as a SCU for STORAGE-SCU AE

Maximum number of simultaneous Associations

10 (Configurable)


F.4.2.1.2.3 Asynchronous Nature

The STORAGE-SCU AE does not support asynchronous communication (multiple outstanding transactions over a single Association). All Association requests must be completed and acknowledged before a new operation can be initiated.

Table F.4.2-4. Asynchronous Nature as a SCU for STORAGE-SCU AE

Maximum number of outstanding asynchronous transactions

1 (Not Configurable)


F.4.2.1.2.4 Implementation Identifying Information

Table F.4.2-5. DICOM Implementation Class and Version for STORAGE-SCU AE

Implementation Class UID

1.840.xxxxxxx.yyy.etc…

Implementation Version Name

EX_VERS_01


Note that the STORAGE-SCU AE and QUERY-RETRIEVE-SCP AE use the same Implementation Class UID. All EXAMPLE-QUERY-RETRIEVE-SERVER AEs use the same Implementation Version Name. This Version Name is updated with each new release of the product software, as the different AE versions are never released independently.

F.4.2.1.3 Association Initiation Policy
F.4.2.1.3.1 Activity - Send Images Requested By an External Peer AE
F.4.2.1.3.1.1 Description and Sequencing of Activity

The STORAGE-SCU AE will initiate a new Association when the QUERY-RETRIEVE-SCP AE invokes the STORAGE-SCU AE to transmit images. The QUERY-RETRIEVE-SCP AE will issue such a command whenever it receives a valid C-MOVE Request. An Association Request is sent to the specified C-MOVE Destination AE and upon successful negotiation of the required Presentation Context the image transfer is started. In all cases an attempt will be made to transmit all the indicated images in a single Association, but this may not always be possible. The Association will be released when all the images have been sent. If an error occurs during transmission over an open Association then the image transfer is halted. The STORAGE-SCU AE will not attempt to independently retry the image export.

Note that the STORAGE-SCU AE does not support the unsolicited sending of SOP Instances using the DICOM Storage Service Class. It will only send SOP Instances in response to a C-MOVE Request from a peer AE.

Sequencing of Activity - Send Images Requested By an External Peer AE

Figure F.4.2-1. Sequencing of Activity - Send Images Requested By an External Peer AE


The following sequencing constraints illustrated in Figure F.4.2-1 apply to the STORAGE-SCU AE:

  1. Peer AE requests retrieval of Study, Series, or Images from QUERY-RETRIEVE-SCP AE (C-MOVE-RQ).

  2. QUERY-RETRIEVE-SCP AE signals STORAGE-SCU AE to send the image Composite SOP Instances indicated in the C-MOVE-RQ to the C-MOVE Destination AE.

  3. STORAGE-SCU AE opens a new Association with the indicated C-MOVE Destination AE.

  4. STORAGE-SCU AE sends the indicated Composite SOP Instances.

  5. STORAGE-SCU AE closes the Association.

  6. The Verification Service is only supported as a utility function for Service staff. It is used only as a diagnostic tool.

F.4.2.1.3.1.2 Proposed Presentation Contexts

STORAGE-SCU AE will propose Presentation Contexts as shown in the following table:

Table F.4.2-6. Proposed Presentation Contexts By the STORAGE-SCU AE

Presentation Context Table

Abstract Syntax

Transfer Syntax

Role

Ext. Neg.

Name

UID

Name

UID

Verification

1.2.840.10008.1.1

DICOM Implicit VR Little Endian

1.2.840.10008.1.2

SCU

None

US Image Storage

(Retired)

1.2.840.10008.5.1.4.1.1.6

DICOM Implicit VR Little Endian

1.2.840.10008.1.2

SCU

None

US Image Storage

(Retired)

1.2.840.10008.5.1.4.1.1.6

DICOM Explicit VR Little Endian

1.2.840.10008.1.2.1

SCU

None

US Image Storage

(Retired)

1.2.840.10008.5.1.4.1.1.6

DICOM Explicit JPEG baseline lossy compression

1.2.840.10008.1.2.4.50

SCU

None

US Image Storage

1.2.840.10008.5.1.4.1.1.6.1

DICOM Implicit VR Little Endian

1.2.840.10008.1.2

SCU

None

US Image Storage

1.2.840.10008.5.1.4.1.1.6.1

DICOM Explicit VR Little Endian

1.2.840.10008.1.2.1

SCU

None

US Image Storage

1.2.840.10008.5.1.4.1.1.6.1

DICOM Explicit JPEG baseline lossy compression

1.2.840.10008.1.2.4.50

SCU

None

US Multi-frame Storage (Retired)

1.2.840.10008.5.1.4.1.1.3

DICOM Implicit VR Little Endian

1.2.840.10008.1.2

SCU

None

US Multi-frame Storage (Retired)

1.2.840.10008.5.1.4.1.1.3

DICOM Explicit VR Little Endian

1.2.840.10008.1.2.1

SCU

None

US Multi-frame Storage (Retired)

1.2.840.10008.5.1.4.1.1.3

DICOM Explicit JPEG baseline lossy compression

1.2.840.10008.1.2.4.50

SCU

None

US Multi-frame Storage

1.2.840.10008.5.1.4.1.1.3.1

DICOM Implicit VR Little Endian

1.2.840.10008.1.2

SCU

None

US Multi-frame Storage

1.2.840.10008.5.1.4.1.1.3.1

DICOM Explicit VR Little Endian

1.2.840.10008.1.2.1

SCU

None

US Multi-frame Storage

1.2.840.10008.5.1.4.1.1.3.1

DICOM Explicit JPEG baseline lossy compression

1.2.840.10008.1.2.4.50

SCU

None

Computer Radiography Image Storage

1.2.840.10008.5.1.4.1.1.1

DICOM Implicit VR Little Endian

1.2.840.10008.1.2

SCU

None

Computer Radiography Image Storage

1.2.840.10008.5.1.4.1.1.1

DICOM Explicit VR Little Endian

1.2.840.10008.1.2.1

SCU

None

CT Image Storage

1.2.840.10008.5.1.4.1.1.2

DICOM Implicit VR Little Endian

1.2.840.10008.1.2

SCU

None

CT Image Storage

1.2.840.10008.5.1.4.1.1.2

DICOM Explicit VR Little Endian

1.2.840.10008.1.2.1

SCU

None

MR Image Storage

1.2.840.10008.5.1.4.1.1.4

DICOM Implicit VR Little Endian

1.2.840.10008.1.2

SCU

None

MR Image Storage

1.2.840.10008.5.1.4.1.1.4

DICOM Explicit VR Little Endian

1.2.840.10008.1.2.1

SCU

None

Secondary Capture Image Storage

1.2.840.10008.5.1.4.1.1.7

DICOM Implicit VR Little Endian

1.2.840.10008.1.2

SCU

None

Secondary Capture Image Storage

1.2.840.10008.5.1.4.1.1.7

DICOM Explicit VR Little Endian

1.2.840.10008.1.2.1

SCU

None

Secondary Capture Image Storage

1.2.840.10008.5.1.4.1.1.7

DICOM Explicit JPEG baseline lossy compression

1.2.840.10008.1.2.4.50

SCU

None


Note

The SOP Classes and Transfer Syntaxes that the STORAGE-SCU AE proposes, as listed above, represent the default behavior. The STORAGE-SCU AE can be configured to propose a subset of these contexts or additional Presentation Contexts. Also, the default Behavior is to propose just a single Transfer Syntax per Presentation Context. However, this can be altered so that every proposed Presentation Context has a unique SOP Class and one or more Transfer Syntaxes. That is, the default behavior is to determine the Transfer Syntaxes the SCP can accept as opposed to which it prefers.

F.4.2.1.3.1.3 SOP Specific Conformance for Verification SOP Class

Standard conformance is provided to the DICOM Verification Service Class as an SCU. The Verification Service as an SCU is actually only supported as a diagnostic service tool for network communication issues.

F.4.2.1.3.1.4 SOP Specific Conformance for Image SOP Classes

Composite DICOM SOP Instances are maintained as DICOM Part 10 compliant files in the EXAMPLE-QUERY-RETRIEVE-SERVER database. The entire set of tags received with the image will be saved in EXAMPLE-QUERY-RETRIEVE-SERVER; this includes all Private and SOP Extended Elements. When a SOP Instance is selected for export from EXAMPLE-QUERY-RETRIEVE-SERVER, its content will be exported as it was originally received except for a few possible exceptions. Some of the Patient demographic and Study information Elements whose values can have been altered due to changes administered on EXAMPLE-QUERY-RETRIEVE-SERVER or changes to the state of the image data due to compression can be altered when the SOP Instance is exported.

The Patient demographic and Study information can be entered or altered by several means: manually, or from HL7 messaging,. The replacement behavior depends on which specific DICOM and HL7 services are supported. Also, this behavior is configurable. Values can be altered without changing the SOP Instance UID unless otherwise noted. Refer to the Annex for the specific details of which Elements can have their values altered at time of export.

The EXAMPLE-QUERY-RETRIEVE-SERVER creates files called Service Logs that can be used to monitor their status and diagnose any problems that may arise. If any error occurs during DICOM communication then appropriate messages are always output to these Service Logs. In addition, error messages may be output as alerts to the User Interface in certain cases.

The STORAGE-SCU AE will exhibit the following Behavior according to the Status Code value returned in a C-STORE Response from a destination C-STORE SCP:

Table F.4.2-7. STORAGE-SCU AE C-STORE Response Status Handling Behavior

Service Status

Further Meaning

Error Code

Behavior

Success

Success

0000

The SCP has successfully stored the exported SOP Instance. A message is sent to the QUERY-RETRIEVE-SCP AE indicating successful export. The QUERY-RETRIEVE-SCP AE will send the appropriate PENDING or SUCCESS Status in the C-MOVE Response.

Success indication message is output to the Service Logs.

No message is posted to the User Interface.

Refused

Out of Resources

A700 -A7FF

This is treated as a permanent Failure. A message is sent to the QUERY-RETRIEVE-SCP AE indicating an export failure and the Association is released. The QUERY-RETRIEVE-SCP AE will send an appropriate Status in the C-MOVE Response.

Error indication message is output to the Service Logs.

No message is posted to the User Interface.

Error

Data Set does not match SOP Class

A900 -A9FF

This is treated as a permanent Failure. A message is sent to the QUERY-RETRIEVE-SCP AE indicating an export failure and the Association is released. The QUERY-RETRIEVE-SCP AE will send an appropriate Status in the C-MOVE Response.

Error indication message is output to the Service Logs.

No message is posted to the User Interface.

Error

Cannot Understand

C000 -CFFF

This is treated as a permanent Failure. A message is sent to the QUERY-RETRIEVE-SCP AE indicating an export failure and the Association is released. The QUERY-RETRIEVE-SCP AE will send an appropriate Status in the C-MOVE Response.

Error indication message is output to the Service Logs.

No message is posted to the User Interface.

Warning

Coercion of Data Elements

B000

Image transmission is considered successful. A message is sent to the QUERY-RETRIEVE-SCP AE indicating successful export. The QUERY-RETRIEVE-SCP AE will send the appropriate PENDING or SUCCESS Status in the C-MOVE Response.

Warning indication message is output to the Service Logs.

No message is posted to the User Interface.

Warning

Data Set does not match SOP Class

B007

Image transmission is considered successful. A message is sent to the QUERY-RETRIEVE-SCP AE indicating successful export. The QUERY-RETRIEVE-SCP AE will send the appropriate PENDING or SUCCESS Status in the C-MOVE Response.

Warning indication message is output to the Service Logs.

No message is posted to the User Interface.

Warning

Elements Discarded

B006

Image transmission is considered successful. A message is sent to the QUERY-RETRIEVE-SCP AE indicating successful export. The QUERY-RETRIEVE-SCP AE will send the appropriate PENDING or SUCCESS Status in the C-MOVE Response.

Warning indication message is output to the Service Logs.

No message is posted to the User Interface.

Warning

Attribute List Error

0107

Image transmission is considered successful. A message is sent to the QUERY-RETRIEVE-SCP AE indicating successful export. The QUERY-RETRIEVE-SCP AE will send the appropriate PENDING or SUCCESS Status in the C-MOVE Response.

Warning indication message is output to the Service Logs.

No message is posted to the User Interface.

Warning

Attribute Value Out of Range

0116

Image transmission is considered successful. A message is sent to the QUERY-RETRIEVE-SCP AE indicating successful export. The QUERY-RETRIEVE-SCP AE will send the appropriate PENDING or SUCCESS Status in the C-MOVE Response.

Warning indication message is output to the Service Logs.

No message is posted to the User Interface.

*

*

Any other status code.

This is treated as a permanent Failure. A message is sent to the QUERY-RETRIEVE-SCP AE indicating an export failure and the Association is released. The QUERY-RETRIEVE-SCP AE will send an appropriate Status in the C-MOVE Response.

Error indication message is output to the Service Logs.

No message is posted to the User Interface.


All Status Codes indicating an error or refusal are treated as a permanent failure. The STORAGE-SCU AE never automatically resends images when an error Status Code is returned in a C-STORE Response. For specific behavior regarding Status Code values returned in C-MOVE Responses, refer to the Services Supported as an SCP by the QUERY-RETRIEVE-SCP AE.

Table F.4.2-8. STORAGE-SCU AE Communication Failure Behavior

Exception

Behavior

Timeout expiry for an expected DICOM Message Response (DIMSE level timeout).

The Association is aborted using a DICOM A-ABORT and a message is sent to the QUERY-RETRIEVE-SCP AE indicating an export failure. The QUERY-RETRIEVE-SCP AE will send an appropriate Status in the C-MOVE Response.

Error indication message is output to the Service Logs.

No message is posted to the User Interface.

Timeout expiry for an expected DICOM PDU or TCP/IP packet (Low-level timeout).

The Association is aborted using a DICOM A-ABORT and a message is sent to the QUERY-RETRIEVE-SCP AE indicating an export failure. The QUERY-RETRIEVE-SCP AE will send an appropriate Status in the C-MOVE Response.

Error indication message is output to the Service Logs.

No message is posted to the User Interface.

Association A-ABORTed by the SCP or the network layers indicate communication loss (i.e., low-level TCP/IP socket closure)

A message is sent to the QUERY-RETRIEVE-SCP AE indicating an export failure. The QUERY-RETRIEVE-SCP AE will send an appropriate Status in the C-MOVE Response.

Error indication message is output to the Service Logs.

No message is posted to the User Interface.


F.4.2.1.4 Association Acceptance Policy

The STORAGE-SCU AE does not accept Associations.

DICOM PS3.2 2016c - Conformance