DICOM PS3.2 2016c - Conformance

A.1 Conformance Statement Overview

The Overview consist of typically 5-10 lines describing the network services and media storage capabilities supported by the product in layman's terms (i.e., no DICOM acronyms should be used).

A table of Supported Networking DICOM Service (SOP) Classes is provided with roles (User/Provider), organized in 4 categories:

The first column shall specify the SOP Classes exactly as named in PS3.6., Registry of DICOM Unique Identifiers. The phrase "and specializations" may be added to indicate support of all specializations negotiated through the SOP Class Common Extended Negotiation. If the implementation supports all SOP Classes of a particular Service Class through SOP Class Common Extended Negotiation, the first column shall specify "All services of the <x> Service Class".

Table A.1-1. Network Services

SOP Classes

User of Service (SCU)

Provider of Service (SCP)

Transfer

CT Image Storage

Yes

No

US Image Storage

Yes

Yes

Query/Retrieve

Patient Root Information Model FIND

Option

No

Notes, Reports, Measurements Transfer

Comprehensive SR, and specializations

No

Yes


The services can be specified as a SCU, SCP or as an Option, which means that it is either configurable or that it can be purchased separately.

Note

Verification SCP (C-Echo) is not included in the table above because it is required for any Acceptor of an Association. The Verification SCU details are covered in the details of the conformance statement.

The SOP Classes are categorized as follows:

Table A.1-2. UID Values

UID Value

UID Name

Category

1.2.840.10008.1.20.1

Storage Commitment Push Model SOP Class

Workflow Management

1.2.840.10008.1.40

Procedural Event Logging SOP Class

Workflow Management

1.2.840.10008.1.42

Substance Administration Logging

Workflow Management

1.­2.840.10008.3.1.2.3.3

Modality Performed Procedure Step SOP Class

Workflow Management

1.­2.840.10008.3.1.2.3.4

Modality Performed Procedure Step Retrieve SOP Class

Workflow Management

1.­2.840.10008.3.1.2.3.5

Modality Performed Procedure Step Notification SOP Class

Workflow Management

1.2.840.10008.4.2

Storage Service Class

Transfer

1.2.840.10008.5.1.1.1

Basic Film Session SOP Class

Print Management

1.2.840.10008.5.1.1.2

Basic Film Box SOP Class

Print Management

1.2.840.10008.5.1.1.4

Basic Grayscale Image Box SOP Class

Print Management

1.2.840.10008.5.1.1.4.1

Basic Color Image Box SOP Class

Print Management

1.2.840.10008.5.1.1.9

Basic Grayscale Print Management Meta SOP Class

Print Management

1.2.840.10008.5.1.1.14

Print Job SOP Class

Print Management

1.2.840.10008.5.1.1.15

Basic Annotation Box SOP Class

Print Management

1.2.840.10008.5.1.1.16

Printer SOP Class

Print Management

1.2.840.10008.5.1.1.16.376

Printer Configuration Retrieval SOP Class

Print Management

1.2.840.10008.5.1.1.18

Basic Color Print Management Meta SOP Class

Print Management

1.2.840.10008.5.1.1.23

Presentation LUT SOP Class

Print Management

1.2.840.10008.5.1.1.24.1

Basic Print Image Overlay Box SOP Class

Print Management

1.2.840.10008.5.1.1.33

Media Creation Management SOP Class UID

Print Management

1.2.840.10008.5.1.4.1.1.1

Computed Radiography Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.1.1

Digital X-Ray Image Storage - For Presentation

Transfer

1.2.840.10008.5.1.4.1.1.1.1.1

Digital X-Ray Image Storage - For Processing

Transfer

1.2.840.10008.5.1.4.1.1.1.2

Digital Mammography X-Ray Image Storage - For Presentation

Transfer

1.2.840.10008.5.1.4.1.1.1.2.1

Digital Mammography X-Ray Image Storage - For Processing

Transfer

1.2.840.10008.5.1.4.1.1.1.3

Digital Intra-oral X-Ray Image Storage - For Presentation

Transfer

1.2.840.10008.5.1.4.1.1.1.3.1

Digital Intra-oral X-Ray Image Storage - For Processing

Transfer

1.2.840.10008.5.1.4.1.1.2

CT Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.2.1

Enhanced CT Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.3.1

Ultrasound Multi-frame Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.4

MR Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.4.1

Enhanced MR Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.4.2

MR Spectroscopy Storage

Transfer

1.2.840.10008.5.1.4.1.1.4.3

Enhanced MR Color Image Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.6.1

Ultrasound Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.6.2

Enhanced US Volume Storage

Transfer

1.2.840.10008.5.1.4.1.1.7

Secondary Capture Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.7.1

Multi-frame Single Bit Secondary Capture Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.7.2

Multi-frame Grayscale Byte Secondary Capture Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.7.3

Multi-frame Grayscale Word Secondary Capture Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.7.4

Multi-frame True Color Secondary Capture Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.9.1.1

12-lead ECG Waveform Storage

Transfer

1.2.840.10008.5.1.4.1.1.9.1.2

General ECG Waveform Storage

Transfer

1.2.840.10008.5.1.4.1.1.9.1.3

Ambulatory ECG Waveform Storage

Transfer

1.2.840.10008.5.1.4.1.1.9.2.1

Hemodynamic Waveform Storage

Transfer

1.2.840.10008.5.1.4.1.1.9.3.1

Cardiac Electrophysiology Waveform Storage

Transfer

1.2.840.10008.5.1.4.1.1.9.4.1

Basic Voice Audio Waveform Storage

Transfer

1.2.840.10008.5.1.4.1.1.9.4.2

General Audio Waveform Storage

Transfer

1.2.840.10008.5.1.4.1.1.9.5.1

Arterial Pulse Waveform Storage

Transfer

1.2.840.10008.5.1.4.1.1.9.6.1

Respiratory Waveform Storage

Transfer

1.2.840.10008.5.1.4.1.1.11.1

Grayscale Softcopy Presentation State Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.11.2

Color Softcopy Presentation State Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.11.3

Pseudo-Color Softcopy Presentation State Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.11.4

Blending Softcopy Presentation State Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.11.5

XA/XRF Grayscale Softcopy Presentation State Storage

Transfer

1.2.840.10008.​5.​1.​4.​1.​1.​11.​6

Grayscale Planar MPR Volumetric Presentation State Storage SOP Class

Transfer

1.2.840.10008.​5.​1.​4.​1.​1.​11.​7

Compositing Planar MPR Volumetric Presentation State Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.12.1

X-Ray Angiographic Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.12.1.1

Enhanced XA Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.12.2

X-Ray Radiofluoroscopic Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.12.2.1

Enhanced XRF Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.13.1.1

X-Ray 3D Angiographic Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.13.1.2

X-Ray 3D Craniofacial Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.13.1.3

Breast Tomosynthesis Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.13.1.4

Breast Projection X-Ray Image Storage - For Presentation

Transfer

1.2.840.10008.5.1.4.1.1.13.1.5

Breast Projection X-Ray Image Storage - For Processing

Transfer

1.2.840.10008.5.1.4.1.1.14.1

Intravascular Optical Coherence Tomography Image Storage - For Presentation

Transfer

1.2.840.10008.5.1.4.1.1.14.2

Intravascular Optical Coherence Tomography Image Storage - For Processing

Transfer

1.2.840.10008.5.1.4.1.1.20

Nuclear Medicine Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.30

Parametric Map Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.66

Raw Data Storage

Transfer

1.2.840.10008.5.1.4.1.1.66.1

Spatial Registration Storage

Transfer

1.2.840.10008.5.1.4.1.1.66.2

Spatial Fiducials Storage

Transfer

1.2.840.10008.5.1.4.1.1.66.3

Deformable Spatial Registration SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.66.4

Segmentation SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.66.5

Surface Segmentation Storage

Transfer

1.2.840.10008.5.1.4.1.1.66.6

Tractography Results Storage

Transfer

1.2.840.10008.5.1.4.1.1.67

Real World Value Mapping Storage

Transfer

1.2.840.10008.5.1.4.1.1.68.1

Surface Scan Mesh Storage

Transfer

1.2.840.10008.5.1.4.1.1.68.2

Surface Scan Point Cloud Storage

Transfer

1.2.840.10008.5.1.4.1.1.77.1.1

VL Endoscopic Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.77.1.1.1

Video Endoscopic Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.77.1.2

VL Microscopic Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.77.1.2.1

Video Microscopic Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.77.1.3

VL Slide-Coordinates Microscopic Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.77.1.4

VL Photographic Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.77.1.4.1

Video Photographic Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.77.1.5.1

Ophthalmic Photography 8 Bit Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.77.1.5.2

Ophthalmic Photography 16 Bit Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.77.1.5.3

Stereometric Relationship Storage

Transfer

1.2.840.10008.5.1.4.1.1.77.1.5.4

Ophthalmic Tomography Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.77.1.5.5

Wide Field Ophthalmic Photography Stereographic Projection Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.77.1.5.6

Wide Field Ophthalmic Photography 3D Coordinates Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.77.1.6

VL Whole Slide Microscopy Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.78.1

Lensometry Measurements Storage

Transfer

1.2.840.10008.5.1.4.1.1.78.2

Autorefraction Measurements Storage

Transfer

1.2.840.10008.5.1.4.1.1.78.3

Keratometry Measurements Storage

Transfer

1.2.840.10008.5.1.4.1.1.78.4

Subjective Refraction Measurements Storage

Transfer

1.2.840.10008.5.1.4.1.1.78.5

Visual Acuity Measurements Storage

Transfer

1.2.840.10008.5.1.4.1.1.78.6

Spectacle Prescription Report Storage

Transfer

1.2.840.10008.5.1.4.1.1.78.7

Ophthalmic Axial Measurements Storage

Transfer

1.2.840.10008.5.1.4.1.1.78.8

Intraocular Lens Calculations Storage

Transfer

1.2.840.10008.5.1.4.1.1.79.1

Macular Grid Thickness and Volume Report Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.80.1

Ophthalmic Visual Field Static Perimetry Measurements Storage

Transfer

1.2.840.10008.5.1.4.1.1.81.1

Ophthalmic Thickness Map Storage

Transfer

1.2.840.10008.5.1.4.1.1.82.1

Corneal Topography Map Storage

Transfer

1.2.840.10008.5.1.4.1.1.88.11

Basic Text SR

Transfer

1.2.840.10008.5.1.4.1.1.88.22

Enhanced SR

Transfer

1.2.840.10008.5.1.4.1.1.88.33

Comprehensive SR

Transfer

1.2.840.10008.5.1.4.1.1.88.34

Comprehensive 3D SR

Transfer

1.2.840.10008.5.1.4.1.1.88.40

Procedure Log Storage

Transfer

1.2.840.10008.5.1.4.1.1.88.50

Mammography CAD SR

Transfer

1.2.840.10008.5.1.4.1.1.88.59

Key Object Selection Document

Transfer

1.2.840.10008.5.1.4.1.1.88.65

Chest CAD SR

Transfer

1.2.840.10008.5.1.4.1.1.88.67

X-Ray Radiation Dose SR

Transfer

1.2.840.10008.5.1.4.1.1.88.68

Radiopharmaceutical Radiation Dose SR

Transfer

1.2.840.10008.5.1.4.1.1.88.69

Colon CAD SR

Transfer

1.2.840.10008.5.1.4.1.1.88.70

Implantation Plan SR Document Storage

Transfer

1.2.840.10008.5.​1.​4.​1.​1.​88.​71

Acquisition Context SR Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.90.1

Content Assessment Results Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.104.1

Encapsulated PDF Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.104.2

Encapsulated CDA Storage SOP Class

Transfer

1.2.840.10008.5.1.4.1.1.128

Positron Emission Tomography Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.130

Enhanced PET Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.131

Basic Structured Display Storage

Transfer

1.2.840.10008.5.1.4.1.1.481.1

RT Image Storage

Transfer

1.2.840.10008.5.1.4.1.1.481.2

RT Dose Storage

Transfer

1.2.840.10008.5.1.4.1.1.481.3

RT Structure Set Storage

Transfer

1.2.840.10008.5.1.4.1.1.481.4

RT Beams Treatment Record Storage

Transfer

1.2.840.10008.5.1.4.1.1.481.5

RT Plan Storage

Transfer

1.2.840.10008.5.1.4.1.1.481.6

RT Brachy Treatment Record Storage

Transfer

1.2.840.10008.5.1.4.1.1.481.7

RT Treatment Summary Record Storage

Transfer

1.2.840.10008.5.1.4.1.2.1.1

Patient Root Query/Retrieve Information Model - FIND

Query/Retrieve

1.2.840.10008.5.1.4.1.2.1.2

Patient Root Query/Retrieve Information Model - MOVE

Query/Retrieve

1.2.840.10008.5.1.4.1.2.1.3

Patient Root Query/Retrieve Information Model - GET

Query/Retrieve

1.2.840.10008.5.1.4.1.2.2.1

Study Root Query/Retrieve Information Model - FIND

Query/Retrieve

1.2.840.10008.5.1.4.1.2.2.2

Study Root Query/Retrieve Information Model - MOVE

Query/Retrieve

1.2.840.10008.5.1.4.1.2.2.3

Study Root Query/Retrieve Information Model - GET

Query/Retrieve

1.2.840.10008.5.1.4.1.2.4.2

Composite Instance Root Retrieve - MOVE

Query/Retrieve

1.2.840.10008.5.1.4.1.2.4.3

Composite Instance Root Retrieve - GET

Query/Retrieve

1.2.840.10008.5.1.4.1.2.5.3

Composite Instance Retrieve Without Bulk Data - GET

Query/Retrieve

1.2.840.10008.5.1.4.31

Modality Worklist Information Model - FIND

Workflow Management

1.2.840.10008.5.1.4.32.1

General Purpose Worklist Information Model - FIND (Retired)

Workflow Management

1.2.840.10008.5.1.4.32.2

General Purpose Scheduled Procedure Step SOP Class (Retired)

Workflow Management

1.2.840.10008.5.1.4.32.3

General Purpose Performed Procedure Step SOP Class (Retired)

Workflow Management

1.2.840.10008.5.1.4.32

General Purpose Worklist Management Meta SOP Class (Retired)

Workflow Management

1.2.840.10008.5.1.4.33

Instance Availability Notification SOP Class

Workflow Management

1.2.840.10008.5.1.4.34.6.1

Unified Procedure Step - Push SOP Class

Workflow Management

1.2.840.10008.5.1.4.34.6.2

Unified Procedure Step - Watch SOP Class

Workflow Management

1.2.840.10008.5.1.4.34.6.3

Unified Procedure Step - Pull SOP Class

Workflow Management

1.2.840.10008.5.1.4.34.6.4

Unified Procedure Step - Event SOP Class

Workflow Management

1.2.840.10008.5.1.4.34.7

RT Beams Delivery Instruction Storage

Transfer

1.2.840.10008.5.1.4.34.8

RT Conventional Machine Verification

Workflow Management

1.2.840.10008.5.1.4.34.9

RT Ion Machine Verification

Workflow Management

1.2.840.10008.5.1.4.34.10

RT Brachy Application Setup Delivery Instruction Storage

Transfer

1.2.840.10008.5.1.4.37.1

General Relevant Patient Information Query

Query/Retrieve

1.2.840.10008.5.1.4.37.2

Breast Imaging Relevant Patient Information Query

Query/Retrieve

1.2.840.10008.5.1.4.37.3

Cardiac Relevant Patient Information Query

Query/Retrieve

1.2.840.10008.5.1.4.38.1

Hanging Protocol Storage

Transfer

1.2.840.10008.5.1.4.38.2

Hanging Protocol Information Model - FIND

Query/Retrieve

1.2.840.10008.5.1.4.38.3

Hanging Protocol Information Model - MOVE

Query/Retrieve

1.2.840.10008.5.1.4.39.1

Color Palette Storage

Transfer

1.2.840.10008.5.1.4.39.2

Color Palette Information Model - FIND

Query/Retrieve

1.2.840.10008.5.1.4.39.3

Color Palette Information Model - MOVE

Query/Retrieve

1.2.840.10008.5.1.4.39.4

Color Palette Information Model - GET

Query/Retrieve

1.2.840.10008.5.1.4.41

Product Characteristics Query

Query/Retrieve

1.2.840.10008.5.1.4.42

Substance Approval Query

Query/Retrieve

1.2.840.10008.5.1.4.43.1

Generic Implant Template Storage

Transfer

1.2.840.10008.5.1.4.43.2

Generic Implant Template Information Model - FIND

Query / Retrieve

1.2.840.10008.5.1.4.43.3

Generic Implant Template Information Model - MOVE

Query / Retrieve

1.2.840.10008.5.1.4.43.4

Generic Implant Template Information Model - GET

Query / Retrieve

1.2.840.10008.5.1.4.44.1

Implant Assembly Template Storage

Transfer

1.2.840.10008.5.1.4.44.2

Implant Assembly Template Information Model - FIND

Query / Retrieve

1.2.840.10008.5.1.4.44.3

Implant Assembly Template Information Model - MOVE

Query / Retrieve

1.2.840.10008.5.1.4.44.4

Implant Assembly Template Information Model - GET

Query / Retrieve

1.2.840.10008.5.1.4.45.1

Implant Template Group Storage

Transfer

1.2.840.10008.5.1.4.45.2

Implant Template Group Information Model - FIND

Query / Retrieve

1.2.840.10008.5.1.4.45.3

Implant Template Group Information Model - MOVE

Query / Retrieve

1.2.840.10008.5.1.4.45.4

Implant Template Group Information Model - GET

Query / Retrieve


A table of Supported Media Storage Application Profiles (with roles) is provided, organized in categories:

Table A.1-3. Media Services

Media Storage Application Profile

Write Files (FSC or FSU)

Read Files (FSR)

Compact Disk - Recordable

General Purpose CD-R

Option

Yes

Magneto-Optical Disk

CT/MR 2.3 GB MOD

Yes

Yes

DVD

General Purpose DVD-RAM

Yes

Yes

BD

General Purpose BD Interchange with MPEG-4 AVC/H.264 BD-Compatible HiP@Level4.1

Yes

Yes

USB and Flash Memory

General Purpose USB Media Interchange with JPEG

Yes

Yes

Email

General Purpose MIME Interchange

Yes

No

General Purpose ZIP Email

Yes

No


DICOM PS3.2 2016c - Conformance