The content of this section provides recommendations on how to express the concepts from draft ASE guidelines with measurement type concept names and concept name modifiers.
The leftmost column is the name of the ASE concept. The Base Measurement Concept Name is the concept name of the numeric measurement content item. The modifiers column specifies a set of modifiers for the base measurement concept name. Each modifier consists of a modifier concept name (e.g., method or mode) and its value (e.g., Continuity). Where no Concept Modifier appears, the base concept matches the ASE concept.
Name of ASE Concept |
Base Measurement Concept Name |
Concept or Acquisition Context Modifiers |
---|---|---|
Aortic Root Diameter |
(18015-8, LN, "Aortic Root Diameter") |
|
Ascending Aortic Diameter |
(18012-5, LN, "Ascending Aortic Diameter") |
|
Aortic Arch Diameter |
(18011-7, LN, "Aortic Arch Diameter") |
|
Descending Aortic Diameter |
(18013-3, LN, "Descending Aortic Diameter") |
Name of ASE Concept |
Base Measurement Concept Name |
Concept or Acquisition Context Modifiers |
---|---|---|
Aortic Valve Cusp Separation |
(17996-0, LN, "Aortic Valve Cusp Separation") |
|
Aortic Valve Systolic Peak Velocity |
(11726-7, LN, "Peak Velocity") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") |
Aortic Valve Systolic Velocity Time Integral |
(20354-7, LN, "Velocity Time Integral") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") |
Aortic Valve Systolic Area |
(G-038E, SRT, "Cardiovascular Orifice Area") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") |
Aortic Valve Planimetered Systolic Area |
(G-038E, SRT, "Cardiovascular Orifice Area") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") (G-C036, SRT, "Measurement Method") = (125220, DCM, "Planimetry") |
Aortic Valve Systolic Area by Continuity |
(G-038E, SRT, "Cardiovascular Orifice Area") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") (G-C036, SRT, "Measurement Method") = (125212, DCM, "Continuity Equation") |
Aortic Valve Systolic Area by Continuity of Peak Velocity |
(G-038E, SRT, "Cardiovascular Orifice Area") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") (G-C036, SRT, "Measurement Method") = (125214. DCM, "Continuity Equation Peak Velocity") |
Aortic Valve Systolic Area by Continuity of Mean Velocity |
(G-038E, SRT, "Cardiovascular Orifice Area") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") (G-C036, SRT, "Measurement Method") = (125213. DCM, "Continuity Equation by Mean Velocity") |
Aortic Valve Systolic Area by Continuity of VTI |
(G-038E, SRT, "Cardiovascular Orifice Area") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") (G-C036, SRT, "Measurement Method") = (1252125, DCM, "Continuity Equation by Velocity Time Integral") |
Aortic Valve Systolic Peak Instantaneous Gradient |
(20247-3 LN, "Peak Gradient") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") |
Aortic Valve Systolic Mean Gradient |
(20256-4, LN, "Mean Gradient") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") |
Aortic Annulus Systolic Diameter |
(G-038F, SRT, "Cardiovascular Orifice Diameter") |
(G-C0E3, SRT, "Finding Site") = (T-35410, SRT, "Aortic Valve Ring") (G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") |
Aortic Valve Regurgitant Diastolic Deceleration Slope |
(20216-8, LN, "Deceleration Slope") |
(G-C048, SRT, "Direction of Flow") = (R-42E61, SRT, "Regurgitant Flow") |
Aortic Valve Regurgitant Diastolic Deceleration Time |
(20217-6, LN, "Deceleration Time") |
(G-C048, SRT, "Direction of Flow") = (R-42E61, SRT, "Regurgitant Flow") |
Aortic Valve Regurgitant Diastolic Pressure Half-time |
(20280-4, LN, "Pressure Half-Time") |
(G-C048, SRT, "Direction of Flow") = (R-42E61, SRT, "Regurgitant Flow") |
Aortic Insufficiency, End-Diastolic Pressure Gradient |
(20247-3, LN, "Peak Gradient") |
(G-C048, SRT, "Direction of Flow") = (R-42E61, SRT, "Regurgitant Flow") |
Aortic Insufficiency, End Diastolic Velocity |
(11653-3, LN, "End Diastolic Velocity") |
(G-C048, SRT, "Direction of Flow") = (R-42E61, SRT, "Regurgitant Flow") |
Aortic Valve measurements appear in TID 5202 “Echo Section”, which specifies the Finding Site to be Aortic Valve with the concept modifier (G-C0E3, SRT, "Finding Site") = (T-35400, SRT, "Aortic Valve").Therefore, the Finding Site modifier does not appear in the right column.
Name of ASE Concept |
Base Measurement Concept Name |
Concept or Acquisition Context Modifiers |
---|---|---|
Left Ventricle Internal End Diastolic Dimension |
(29436-3, LN "Left Ventricle Internal End Diastolic Dimension") |
|
Left Ventricle Internal Systolic Dimension |
(29438-9, LN, "Left Ventricle Internal Systolic Dimension") |
|
Left Ventricle Diastolic Major Axis |
(18077-8, LN, "Left Ventricle Diastolic Major Axis") |
|
Left Ventricle Systolic Major Axis |
(18076-0, LN, "Left Ventricle Systolic Major Axis") |
|
Left Ventricular Fractional Shortening |
(18051-3, LN, "Left Ventricular Fractional Shortening") |
|
Interventricular Septum Diastolic Thickness |
(18154-5, LN, "Interventricular Septum Diastolic Thickness") |
|
Interventricular Septum Systolic Thickness |
(18158-6, LN, "Interventricular Septum Systolic Thickness") |
|
Interventricular Septum % Thickening |
(18054-7, LN, "Interventricular Septum % Thickening") |
|
Left Ventricle Posterior Wall Diastolic Thickness |
(18152-9, LN, "Left Ventricle Posterior Wall Diastolic Thickness") |
|
Left Ventricle Posterior Wall Systolic Thickness |
(18156-0, LN, "Left Ventricle Posterior Wall Systolic Thickness") |
|
Left Ventricle Posterior Wall % Thickening |
(18053-9, LN, "Left Ventricle Posterior Wall % Thickening") |
|
Interventricular Septum to Posterior Wall Thickness ratio |
(18155-2, LN, "Interventricular Septum to Posterior Wall Thickness Ratio") |
|
Left Ventricular Internal End Diastolic Dimension by 2-D |
(29436-3, LN, "Left Ventricle Internal End Diastolic Dimension") |
(G-0373, SRT, "Image Mode") = (G-03A2, SRT, "2D mode") |
Left Ventricular Internal Systolic Dimension by 2-D |
(29438-9, LN, "Left Ventricle Internal Systolic Dimension") |
(G-0373, SRT, "Image Mode") = (G-03A2, SRT, "2D mode") |
Left Ventricular Fractional Shortening by 2-D |
(18051-3, LN, "Left Ventricular Fractional Shortening") |
(G-0373, SRT, "Image Mode") = (G-03A2, SRT, "2D mode") |
Interventricular Septum Diastolic Thickness by 2-D |
(18154-5, LN, "Interventricular Septum Diastolic Thickness") |
(G-0373, SRT, "Image Mode") = (G-03A2, SRT, "2D mode") |
Interventricular Septum Systolic Thickness by 2-D |
(18158-6, LN, "Interventricular Septum Systolic Thickness") |
(G-0373, SRT, "Image Mode") = (G-03A2, SRT, "2D mode") |
Interventricular Septum % Thickening by 2-D |
(18054-7, LN, "Interventricular Septum % Thickening") |
(G-0373, SRT, "Image Mode") = (G-03A2, SRT, "2D mode") |
Left Ventricular Posterior Wall Diastolic Thickness by 2-D |
(18152-9, LN, "Left Ventricle Posterior Wall Diastolic Thickness") |
(G-0373, SRT, "Image Mode") = (G-03A2, SRT, "2D mode") |
Left Ventricle Posterior Wall Systolic Thickness by 2-D |
(18156-0, LN, "Left Ventricle Posterior Wall Systolic Thickness") |
(G-0373, SRT, "Image Mode") = (G-03A2, SRT, "2D mode") |
Left Ventricle Posterior Wall % Thickening by 2-D |
(18053-9, LN, "Left Ventricle Posterior Wall % Thickening") |
G-0373, SRT, "Image Mode") = (G-03A2, SRT, "2D mode") |
Interventricular Septum/ Left Ventricular Posterior Wall Diastolic Thickness Ratio by 2-D |
(18155-2, LN, "Interventricular Septum to Posterior Wall Thickness Ratio") |
(G-0373, SRT, "Image Mode") = (G-03A2, SRT, "2D mode") |
Left Ventricular Internal End Diastolic Dimension by M-Mode |
(29436-3, LN, "Left Ventricle Internal End Diastolic Dimension") |
(G-0373, SRT, "Image Mode") = (G-0394, SRT, "M mode") |
Left Ventricular Internal Systolic Dimension by M-Mode |
(29438-9, LN, "Left Ventricle Internal Systolic Dimension") |
(G-0373, SRT, "Image Mode") = (G-0394, SRT, "M mode") |
Left Ventricular Systolic Fractional Shortening by M-Mode |
(18051-3, LN, "Left Ventricular Fractional Shortening") |
(G-0373, SRT, "Image Mode") = (G-0394, SRT, "M mode") |
Interventricular Septum Diastolic Thickness by M-Mode |
(18154-5, LN, "Interventricular Septum Diastolic Thickness") |
(G-0373, SRT, "Image Mode") = (G-0394, SRT, "M mode") |
Interventricular Septum Systolic Thickness by M-Mode |
(18158-6, LN, "Interventricular Septum Systolic Thickness") |
(G-0373, SRT, "Image Mode") = (G-0394, SRT, "M mode") |
Interventricular Septum % Thickening by M-Mode |
(18054-7, LN, "Interventricular Septum % Thickening") |
G-0373, SRT, "Image Mode") = (G-0394, SRT, "M mode") |
Left Ventricular Posterior Wall Diastolic Thickness by M-Mode |
(18152-9, LN, "Left Ventricle Posterior Wall Diastolic Thickness") |
(G-0373, SRT, "Image Mode") = (G-0394, SRT, "M mode") |
Left Ventricle Posterior Wall Systolic Thickness by M-Mode |
(18156-0, LN, "Left Ventricle Posterior Wall Systolic Thickness") |
(G-0373, SRT, "Image Mode") = (G-0394, SRT, "M mode") |
Left Ventricle Posterior Wall % Thickening by M-Mode |
(18053-9, LN, "Left Ventricle Posterior Wall % Thickening") |
(G-0373, SRT, "Image Mode") = (G-0394, SRT, "M mode") |
Interventricular Septum to Left Ventricular Posterior Wall Ratio by M-Mode |
(18155-2, LN, "Interventricular Septum to Posterior Wall Thickness Ratio") |
(G-0373, SRT, "Image Mode") = (G-0394, SRT, "M mode") |
Name of ASE Concept |
Base Measurement Concept Name |
Concept or Acquisition Context Modifiers |
---|---|---|
Left Ventricular End Diastolic Volume |
(18026-5, LN, "Left Ventricular End Diastolic Volume") |
|
Left Ventricular End Diastolic Volume by Teichholz Method |
(18026-5, LN, "Left Ventricular End Diastolic Volume") |
(G-C036, SRT, "Measurement Method") = (125209, DCM, "Teichholz") |
Left Ventricular End Diastolic Volume by 2-D Single Plane by Method of Disks (4-Chamber) |
(18026-5, LN, "Left Ventricular End Diastolic Volume") |
(111031, DCM, "Image View") = (G-A19C, SRT, "Apical Four Chamber") (G-C036, SRT, "Measurement Method") = (125208, DCM, "Method of Disks, Single Plane") |
Left Ventricular End Diastolic Volume by 2-D Biplane by Method of Disks |
(18026-5, LN, "Left Ventricular End Diastolic Volume") |
(G-C036, SRT, "Measurement Method") = (125207, DCM, "Method of Disks, Biplane") |
Left Ventricular End Systolic Volume |
(18148-7, LN, "Left Ventricular End Systolic Volume") |
|
Left Ventricular End Systolic Volume by Teichholz Method |
(18148-7, LN, "Left Ventricular End Systolic Volume") |
(G-C036, SRT, "Measurement Method") = (125209, DCM, "Teichholz") |
Left Ventricular End Systolic Volume by 2D Single Plane by Method of Disks (4-Chamber) |
(18148-7, LN, "Left Ventricular End Systolic Volume") |
(111031, DCM, "Image View") = (G-A19C, SRT, "Apical Four Chamber") (G-C036, SRT, "Measurement Method") = (125208, DCM, "Method of Disks, Single Plane") |
Left Ventricular End Systolic Volume by 2-D Biplane by Method of Disks |
(18148-7, LN, "Left Ventricular End Systolic Volume") |
(G-C036, SRT, "Measurement Method") = (125207, DCM, "Method of Disks, Biplane") |
Left Ventricular EF |
(18043-0, LN, "Left Ventricular Ejection Fraction") |
|
Left Ventricular EF by Teichholz Method |
(18043-0, LN, "Left Ventricular Ejection Fraction") |
(G-C036, SRT, "Measurement Method") = (125209, DCM, "Teichholz") |
Left Ventricular EF by 2D Single Plane by Method of Disks (4-Chamber) |
(18043-0, LN, "Left Ventricular Ejection Fraction") |
(111031, DCM, "Image View") = (G-A19C, SRT, "Apical Four Chamber ") (G-C036, SRT, "Measurement Method") = (125208, DCM, "Method Of Disks, Single Plane") |
Left Ventricular EF by 2-D Biplane by Method of Disks |
(18043-0, LN, "Left Ventricular Ejection Fraction") |
(G-C036, SRT, "Measurement Method") = (1252087, DCM, "Method of Disks, Biplane") |
Name of ASE Concept |
Base Measurement Concept Name |
Concept or Acquisition Context Modifiers |
---|---|---|
Left Ventricular Stroke Volume |
(F-32120, SRT, "Stroke Volume") |
|
Left Ventricular Stroke Volume by Doppler Volume Flow |
(F-32120, SRT, "Stroke Volume") |
(G-C036, SRT, "Measurement Method") = (125219, DCM, "Doppler Volume Flow") (G-C0E3, SRT, "Finding Site") = (T-32650, SRT, "Left Ventricle Outflow Tract") |
Left Ventricular Stroke Volume by Teichholz Method |
(F-32120, SRT, "Stroke Volume") |
(G-C036, SRT, "Measurement Method") = (125209, DCM, "Teichholz") |
Left Ventricular Stroke Volume by 2-D Single Plane by Method of Disks (4-Chamber) |
(F-32120, SRT, "Stroke Volume") |
(1110321 DCM, "Image View")= (G-A19C, SRT, "Apical Four Chamber") (G-C036, SRT, "Measurement Method") = (125208, DCM, "Method of Disks, Single Plane") |
Left Ventricular Stroke Volume by 2-D Biplane by Method of Disks |
(F-32120, SRT, "Stroke Volume") |
(G-C036, SRT, "Measurement Method") = (125207, DCM, "Method of Disks, Biplane") |
Left Ventricular Cardiac Output |
(F-32100, SRT, "Cardiac Output") |
|
Left Ventricular Cardiac Output by Doppler Volume Outflow |
(F-32100, SRT, "Cardiac Output") |
(G-C036, SRT, "Measurement Method") = (125219, DCM, "Doppler Volume Flow") (G-C0E3, SRT, "Finding Site") = (T-32650, SRT, "Left Ventricle Outflow Tract") |
Left Ventricular Cardiac Output by Teichholz Method |
(F-32100, SRT, "Cardiac Output") |
(G-C036, SRT, "Measurement Method") = (125209, DCM, "Teichholz") |
Left Ventricular Cardiac Output by 2-D Single Plane by Method of Disks (4-Chamber) |
(F-32100, SRT, "Cardiac Output") |
(111031, DCM, "Image View") = (G-A19C, SRT, "Apical Four Chamber") (G-C036, SRT, "Measurement Method") = (125208, DCM, "Method of Disks, Single Plane") |
Left Ventricular Cardiac Output by 2-D Biplane by Method of Disks |
(F-32100, SRT, "Cardiac Output") |
(G-C036, SRT, "Measurement Method") = (125207, DCM, "Method of Disks, Biplane") |
Left Ventricular Cardiac Index |
(F-32110, SRT, "Cardiac Index") |
|
Left Ventricular Cardiac Index by Doppler Volume Flow |
(F-32110, SRT, "Cardiac Index") |
(G-C036, SRT, "Measurement Method") = (125219, DCM, "Doppler Volume Flow") |
Left Ventricular Cardiac Index by Teichholz Method |
(F-32110, SRT, "Cardiac Index") |
(G-C036, SRT, "Measurement Method") = (125209, DCM, "Teichholz") |
Left Ventricular Cardiac Index by 2-D Single Plane by Method of Disks (4-Chamber) |
(F-32110, SRT, "Cardiac Index") |
(111031, DCM, "Image View") = (G-A19C, SRT, "Apical Four Chamber") (G-C036, SRT, "Measurement Method") = (125208, DCM, "Method Of Disks, Single Plane") |
Left Ventricular Cardiac Index by 2-D Biplane by Method of Disks |
(F-32110, SRT, "Cardiac Index") |
(G-C036, SRT, "Measurement Method") = (125207, DCM, "Method of Disks, Biplane") |
Measurements in the Left Ventricle section have context of Left Ventricle and do not require a Finding Site modifier (G-C0E3, SRT, "Finding Site") = (T-35400, SRT, "Left Ventricle") to specify the site. The Finding Site modifier appears for more specificity.
Name of ASE Concept |
Base Measurement Concept Name |
Concept or Acquisition Context Modifiers |
---|---|---|
Left Ventricular Outflow Tract Systolic Diameter |
(G-038F, SRT, "Cardiovascular Orifice Diameter") |
(G-C0E3, SRT, "Finding Site") = (T-32650, SRT, "Left Ventricular Outflow Tract") |
Left Ventricular Outflow Tract Systolic Cross Sectional Area |
(G-038E, SRT, "Cardiovascular Orifice Area") |
(G-C0E3, SRT, "Finding Site") = (T-32650, SRT, "Left Ventricular Outflow Tract") |
Left Ventricular Outflow Tract Systolic Peak Velocity |
(11726-7, LN, "Peak Velocity") |
(G-C0E3, SRT, "Finding Site") = (T-32650, SRT, "Left Ventricular Outflow Tract") |
Left Ventricular Outflow Tract Systolic Peak Instantaneous Gradient |
(20247-3, LN, "Peak Gradient") |
(G-C0E3, SRT, "Finding Site") = (T-32650, SRT, "Left Ventricular Outflow Tract") |
Left Ventricular Outflow Tract Systolic Mean Velocity |
(20352-1, LN, "Mean Velocity") |
(G-C0E3, SRT, "Finding Site") = (T-32650, SRT, "Left Ventricular Outflow Tract") |
Left Ventricular Outflow Tract Systolic Mean Gradient |
(20256-4, LN, "Mean Gradient") |
(G-C0E3, SRT, "Finding Site") = (T-32650, SRT, "Left Ventricular Outflow Tract") |
Left Ventricular Outflow Tract Systolic Velocity Time Integral |
(20354-7, LN, "Velocity Time Integral") |
(G-C0E3, SRT, "Finding Site") = (T-32650, SRT, "Left Ventricular Outflow Tract") |
Name of ASE Concept |
Base Measurement Concept Name |
Concept or Acquisition Context Modifiers |
---|---|---|
Left Ventricle Mass |
(18087-7, LN, "Left Ventricle Mass") |
|
Left Ventricular Mass by 2-D Method of Disks, Single Plane (4-Chamber) |
(18087-7, LN, "Left Ventricle Mass") |
(G-0373, SRT, "Image Mode") = (G-03A2, SRT, "2D mode") (G-C036, SRT, "Measurement Method") = (125208, DCM, "Method Of Disks, single plane") |
Left Ventricular Mass by 2-D Biplane by Method of Disks |
(18087-7, LN, "Left Ventricle Mass") |
(G-0373, SRT, "Image Mode") = (G-03A2, SRT, "2D mode") (G-C036, SRT, "Measurement Method") = (1252087, DCM, "Method of disks, biplane") |
Left Ventricular Mass by M-Mode |
(18087-7, LN, "Left Ventricle Mass") |
(G-0373, SRT, "Image Mode") = (G-0394, SRT, "M mode") |
Name of ASE Concept |
Base Measurement Concept Name |
Concept or Acquisition Context Modifiers |
---|---|---|
Left Ventricular Isovolumic Relaxation Time |
(18071-1, LN, "Left Ventricular Isovolumic Relaxation Time") |
|
Left Ventricular Isovolumic Contraction Time |
(G-037E, SRT, "Left Ventricular Isovolumic Contraction Time") |
|
Left Ventricular Peak Early Diastolic Tissue Velocity at the Medial Mitral Annulus |
(G-037A, SRT, "Left Ventricular Peak Early Diastolic Tissue Velocity") |
(G-C0E3, SRT, "Finding Site") = (G-0391, SRT, "Medial Mitral Annulus") |
Left Ventricular Peak Early Diastolic Tissue Velocity at the Lateral Mitral Annulus |
(G-037A, SRT, "Left Ventricular Peak Early Diastolic Tissue Velocity") |
(G-C0E3, SRT, "Finding Site") = (G-0392, SRT, "Lateral Mitral Annulus") |
Ratio of Mitral Valve E-Wave Peak Velocity to Left Ventricular Peak Early Diastolic Tissue Velocity at the Medial Mitral Annulus |
(G-037B, SRT, "Ratio of MV Peak Velocity to LV Peak Tissue Velocity E-Wave") |
(G-C0E3, SRT, "Finding Site") = (G-0391, SRT, "Medial Mitral Annulus") |
Ratio of Mitral Valve E-Wave Peak Velocity to Left Ventricular Peak Early Diastolic Tissue Velocity at the Lateral Mitral Annulus |
(G-037B, SRT, "Ratio of MV Peak Velocity to LV Peak Tissue Velocity E-Wave") |
(G-C0E3, SRT, "Finding Site") = (G-0392, SRT, "Lateral Mitral Annulus") |
Left Ventricular Peak Diastolic Tissue Velocity at the Medial Mitral Annulus During Atrial Systole |
(G-037C, SRT, "LV Peak Diastolic Tissue Velocity During Atrial Systole") |
(G-C0E3, SRT, "Finding Site") = (G-0391, SRT, "Medial Mitral Annulus") |
Left Ventricular Peak Diastolic Tissue Velocity at the Lateral Mitral Annulus During Atrial Systole |
(G-037C, SRT, "LV Peak Diastolic Tissue Velocity During Atrial Systole") |
(G-C0E3, SRT, "Finding Site") = (G-0392, SRT, "Lateral Mitral Annulus") |
Left Ventricular Peak Systolic Tissue Velocity at the Medial Mitral Annulus |
(G-037D, SRT, "Left Ventricular Peak Systolic Tissue Velocity") |
(G-C0E3, SRT, "Finding Site") = (G-0391, SRT, "Medial Mitral Annulus") |
Left Ventricular Peak Systolic Tissue Velocity at the Lateral Mitral Annulus |
(G-037D, SRT, "Left Ventricular Peak Systolic Tissue Velocity") |
(G-C0E3, SRT, "Finding Site") = (G-0392, SRT, "Lateral Mitral Annulus") |
Name of ASE Concept |
Base Measurement Concept Name |
Concept or Acquisition Context Modifiers |
---|---|---|
Mitral Valve Area |
(G-038E, SRT, "Cardiovascular Orifice Area") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") |
Mitral Valve Area by Continuity |
(G-038E, SRT, "Cardiovascular Orifice Area") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") (G-C036, SRT, "Measurement Method") = (125212. DCM, "Continuity Equation") |
Mitral Valve Area by Planimetry |
(G-038E, SRT, "Cardiovascular Orifice Area") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") (G-C036, SRT, "Measurement Method") = (125220, DCM, "Planimetry") |
Mitral Valve Area by Pressure Half-time |
(G-038E, SRT, "Cardiovascular Orifice Area") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") (G-C036, SRT, "Measurement Method") = (125210, DCM, "Area by PHT") |
Mitral Valve Area by Proximal Isovelocity Surface Area |
(G-038E, SRT, "Cardiovascular Orifice Area") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") (G-C036, SRT, "Measurement Method") = (125216, DCM, "Proximal Isovelocity Surface Area") |
Mitral Valve Pressure Half-time |
(20280-4, LN, "Pressure Half-Time") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") |
Mitral Valve A-Wave Peak Velocity |
(17978-8, LN, "Mitral Valve A-Wave Peak Velocity") |
|
Mitral Valve E-Wave Peak Velocity |
(18037-2, LN, "Mitral Valve E-Wave Peak Velocity") |
|
Mitral Valve E to A Ratio |
(18038-0, LN, "Mitral Valve E to A Ratio") |
|
Mitral Valve E-Wave Deceleration Time |
(G-0384, SRT, "Mitral Valve E-Wave Deceleration Time") |
|
Mitral Valve E-F Slope by M-Mode |
(18040-6, LN, "Mitral Valve E-F Slope by M-Mode") |
|
Mitral Valve Velocity Time Integral |
(20354-7, LN, "Velocity Time Integral") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") |
Mitral Valve Diastolic Peak Instantaneous Gradient |
(20247-3, LN, "Peak Gradient") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") |
Mitral Valve Diastolic Mean Gradient |
(20256-4, LN, "Mean Gradient") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") |
Mitral Valve Annulus Diastolic Velocity Time Integral |
(20354-7, LN, "Velocity Time Integral") |
(G-C0E3, SRT, "Finding Site") = (T-35313, SRT, "Mitral Annulus") (G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") |
Mitral Valve Annulus Diastolic Diameter |
(G-038F, SRT, "Cardiovascular Orifice Diameter") |
(G-C0E3, SRT, "Finding Site") = (T-35313, SRT, "Mitral Annulus") (G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") |
Mitral Regurgitant Peak Velocity |
(11726-7, LN, "Peak Velocity") |
(G-C048, SRT, "Direction of Flow") = (R-42E61, SRT, "Regurgitant Flow") |
Mitral Valve Effective Regurgitant Orifice by Proximal Isovelocity Surface Area Method |
(G-038E, SRT, "Cardiovascular Orifice Area") |
(G-C048, SRT, "Direction of Flow") = (R-42E61, SRT, "Regurgitant Flow")G-C036, SRT, "Measurement Method") = (125216, DCM, "Proximal Isovelocity Surface Area") |
Mitral Valve Regurgitant Volume by Proximal Isovelocity Surface Area Method |
(33878-0, LN, "Volume Flow") |
(G-C0E3, SRT, "Finding Site") = (T-35313, SRT, "Mitral Annulus") (G-C048, SRT, "Direction of Flow") = (R-42E61, SRT, "Regurgitant Flow") (G-C036, SRT, "Measurement Method") = (125216, DCM, "Proximal Isovelocity Surface Area") |
Mitral Valve Regurgitant Fraction |
(G-0390, SRT, "Regurgitant Fraction") |
|
Mitral Valve Regurgitant Fraction by PISA |
(G-0390, SRT, "Regurgitant Fraction") |
(G-C036, SRT, "Measurement Method") = (125216, DCM, "Proximal Isovelocity Surface Area") |
Mitral Valve Regurgitant Fraction by Mitral Annular Flow |
(G-0390, SRT, "Regurgitant Fraction") |
(G-C0E3, SRT, "Finding Site") = (T-35313, SRT, "Mitral Annulus") (G-C036, SRT, "Measurement Method") = (125219, DCM, "Doppler Volume Flow") |
Mitral Regurgitation Peak Gradient |
(20247-3, LN, "Peak Gradient") |
(G-C048, SRT, "Direction of Flow") = (R-42E61, SRT, "Regurgitant Flow") |
Left Ventricular dP/dt derived from Mitral Regurgitation velocity |
(18035-6, LN, "Mitral Regurgitation dP/dt derived from Mitral Regurgitation velocity") |
Mitral Valve measurements appear in TID 5202 “Echo Section”, which specifies the Finding Site to be Mitral Valve with the concept modifier (G-C0E3, SRT, "Finding Site") = (T-35300, SRT, "Mitral Valve").Therefore, the Finding Site modifier does not appear in the right column.
Name of ASE Concept |
Base Measurement Concept Name |
Concept or Acquisition Context Modifiers |
---|---|---|
Pulmonary Vein Systolic Peak Velocity |
(29450-4, LN, "Pulmonary Vein Systolic Peak Velocity") |
|
Pulmonary Vein Diastolic Peak Velocity |
(29451-2, LN, "Pulmonary Vein Diastolic Peak Velocity") |
|
Pulmonary Vein Systolic to Diastolic Ratio |
(29452-0, LN, "Pulmonary Vein Systolic to Diastolic Ratio") |
|
Pulmonary Vein Atrial Contraction Reversal Peak Velocity |
(29453-8, LN, "Pulmonary Vein Atrial Contraction Reversal Peak Velocity") |
|
Right Upper Pulmonary Vein Peak Systolic Velocity |
(29450-4, LN, "Pulmonary Vein Systolic Peak Velocity") |
(G-A1F8G-A1F8, SRT, "Topographical Modifier") = (R-404A0, SRT, "Right Upper Segment") |
Right Upper Pulmonary Vein Diastolic Peak Velocity |
(29451-2, LN, "Pulmonary Vein Diastolic Peak Velocity") |
(G-A1F8G-A1F8, SRT, "Topographical Modifier") = (R-404A0, SRT, "Right Upper Segment") |
Right Upper Pulmonary Vein Systolic to Diastolic Velocity Ratio |
(29452-0, LN, "Pulmonary Vein Systolic to Diastolic Ratio") |
(G-A1F8, SRT, "Anatomic Site Modifier") = (R-404A0, SRT, "Right Upper Segment") |
Right Lower Pulmonary Vein Peak Systolic Velocity |
(29450-4, LN, "Pulmonary Vein Systolic Peak Velocity") |
(G-A1F8, SRT, "Topographical Modifier") = (R-4049E, SRT, "Right Lower Segment") |
Right Lower Pulmonary Vein Diastolic Peak Velocity |
(29451-2, LN, "Pulmonary Vein Diastolic Peak Velocity") |
(G-A1F8, SRT, "Topographical Modifier") = (R-4049E, SRT, "Right Lower Segment") |
Right Lower Pulmonary Vein Systolic to Diastolic Velocity Ratio |
(29452-0, LN, "Pulmonary Vein Systolic to Diastolic Ratio") |
(G-A1F8, SRT, "Topographical Modifier") = (R-4049E, SRT, "Right Lower Segment") |
Left Upper Pulmonary Vein Peak Systolic Velocity |
(29450-4, LN, "Pulmonary Vein Systolic Peak Velocity") |
(G-A1F8, SRT, "Topographical Modifier") = (R-40491, SRT, "Left Upper Segment") |
Left Upper Pulmonary Vein Velocity Peak Diastolic |
(29451-2, LN, "Pulmonary Vein Diastolic Peak Velocity") |
(G-A1F8, SRT, "Topographical Modifier") = (R-40491, SRT, "Left Upper Segment") |
Left Upper Pulmonary Vein Systolic to Diastolic Velocity Ratio |
(29452-0, LN, "Pulmonary Vein Systolic to Diastolic Ratio") |
(G-A1F8, SRT, "Topographical Modifier") = (R-40491, SRT, "Left Upper Segment") |
Left Lower Pulmonary Vein Peak Systolic Velocity |
(29450-4, LN, "Pulmonary Vein Systolic Peak Velocity") |
(G-A1F8, SRT, "Topographical Modifier") = (R-4214B, SRT, "Left Lower Segment") |
Left Lower Pulmonary Vein Diastolic Peak Velocity |
(29451-2, LN, "Pulmonary Vein Diastolic Peak Velocity") |
(G-A1F8, SRT, "Topographical Modifier") = (R-4214B, SRT, "Left Lower Segment") |
Left Lower Pulmonary Vein Systolic to Diastolic Velocity Ratio |
(29452-0, LN, "Pulmonary Vein Systolic to Diastolic Ratio") |
(G-A1F8, SRT, "Topographical Modifier") = (R-4214B, SRT, "Left Lower Segment") |
Name of ASE Concept |
Base Measurement Concept Name |
Concept or Acquisition Context Modifiers |
---|---|---|
Left Atrium Antero-posterior Systolic Dimension |
(29469-4, LN, "Left Atrium Antero-posterior Systolic Dimension") |
|
Left Atrial Antero-posterior Systolic Dimension by M-Mode |
(29469-4, LN, "Left Atrium Antero-posterior Systolic Dimension") |
(G-0373, SRT, "Image Mode") = (G-0394, SRT, "M mode") |
Left Atrial Antero-posterior Systolic Dimension by 2-D |
(29469-4, LN, "Left Atrium Antero-posterior Systolic Dimension") |
(G-0373, SRT, "Image Mode") = (G-03A2, SRT, "2D mode") |
Left Atrium to Aortic Root Ratio |
(17985-3, LN, "Left Atrium to Aortic Root Ratio") |
|
Left Atrial Appendage Peak Velocity |
(29486-8, LN, "Left Atrial Appendage Peak Velocity") |
|
Left Atrium Systolic Area |
(17977-0, LN, "Left Atrium Area A4C view") |
(R-4089A, SRT, "Cardiac Cycle Point") = (F-32020, SRT, "Systole") |
Left Atrium Systolic Volume |
(G-0383, SRT, "Left Atrium Systolic Volume") |
Name of ASE Concept |
Base Measurement Concept Name |
Concept or Acquisition Context Modifiers |
---|---|---|
Right Ventricular Internal Diastolic Dimension by M-Mode |
(20304-2, LN, "Right Ventricular Internal Diastolic Dimension") |
(G-0373, SRT, "Image Mode") = (G-0394, SRT, "M mode") |
Right Ventricular Internal Diastolic Dimension by 2-D |
(20304-2, LN, "Right Ventricular Internal Diastolic Dimension") |
(G-0373, SRT, "Image Mode") = (G-03A2, SRT, "2D mode") |
Right Ventricular Outflow Tract Systolic Peak Velocity |
(11726-7, LN, "Peak Velocity") |
(G-C0E3, SRT, "Finding Site") = (T-32550, SRT, "Right Ventricular Outflow Tract") |
Right Ventricular Outflow Tract Systolic Velocity Time Integral |
(20354-7, LN, "Velocity Time Integral") |
(G-C0E3, SRT, "Finding Site") = (T-32550, SRT, "Right Ventricular Outflow Tract") |
Right Ventricular Outflow Systolic Diameter by 2-D |
(G-038F, SRT, "Cardiovascular Orifice Diameter") |
(G-C0E3, SRT, "Finding Site") = (T-32550, SRT, "Right Ventricular Outflow Tract") (G-0373, SRT, "Image Mode") = (G-03A2, SRT, "2D mode") |
Right Ventricular Outflow Tract Systolic Peak Instantaneous Gradient |
(20247-3, LN, "Peak Gradient") |
(G-C0E3, SRT, "Finding Site") = (T-32550, SRT, "Right Ventricular Outflow Tract") |
Right Ventricular Outflow Tract Systolic Mean Gradient |
(20256-4, LN, "Mean Gradient") |
(G-C0E3, SRT, "Finding Site") = (T-32550, SRT, "Right Ventricular Outflow Tract") |
Right Ventricular Stroke Volume by Doppler Volume Outflow |
(F-32120, SRT, "Stroke Volume") |
(G-C036, SRT, "Measurement Method") = (125219, DCM, "Doppler Volume Flow") (G-C0E3, SRT, "Finding Site") = (T-32550, SRT, "Right Ventricular Outflow Tract") |
Right Ventricular Outflow Tract Area |
(G-038E, SRT, "Cardiovascular Orifice Area") |
(G-C0E3, SRT, "Finding Site") = (T-32550, SRT, "Right Ventricular Outflow Tract") |
Right Ventricular Outflow Tract Mean Velocity |
(20352-1, LN, "Mean Velocity") |
(G-C0E3, SRT, "Finding Site") = (T-32550, SRT, "Right Ventricular Outflow Tract") |
Right Ventricle Anterior Wall Diastolic Thickness |
(18153-7, LN, "Right Ventricle Anterior Wall Diastolic Thickness") |
|
Right Ventricular Anterior Wall Systolic Thickness |
(18157-8, LN, "Right Ventricular Anterior Wall Systolic Thickness") |
|
Right Ventricular Peak Systolic Pressure |
(G-0380, SRT, "Right Ventricular Peak Systolic Pressure") |
Name of ASE Concept |
Base Measurement Concept Name |
Concept or Acquisition Context Modifiers |
---|---|---|
Main Pulmonary Artery Diameter |
(18020-8, LN, "Main Pulmonary Artery Diameter") |
|
Main Pulmonary Artery Velocity |
(G-038A, SRT, "Main Pulmonary Artery Velocity") |
|
Right Pulmonary Artery Diameter |
(18021-6, LN, "Right Pulmonary Artery Diameter") |
|
Left Pulmonary Artery Diameter |
(18019-0, LN, "Left Pulmonary Artery Diameter") |
|
Pulmonic Valve Systolic Peak Instantaneous Gradient |
(20247-3, LN, "Peak Gradient") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") |
Pulmonic Valve Systolic Mean Gradient |
(20256-4, LN, "Mean Gradient") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") |
Pulmonic Valve Systolic Peak Velocity |
(20354-7, LN, 11726-7, LN, "Peak Velocity") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") |
Pulmonic Valve Systolic Velocity Time Integral |
(20354-7, LN, "Velocity Time Integral") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") |
Pulmonic Valve Area by Continuity |
(18096-8, LN, "Pulmonic valve Area by Continuity") |
|
Pulmonic Valve Acceleration Time |
(20168-1, LN, "Acceleration Time") |
(G-C048, SRT, "Direction of Flow") = (R-42E61, SRT, "Regurgitant Flow") |
Pulmonic Valve Regurgitant End Diastolic Velocity |
(11653-3, LN, "End Diastolic Velocity") |
(G-C048, SRT, "Direction of Flow") = (R-42E61, SRT, "Regurgitant Flow") |
Pulmonic Valve Regurgitant Diastolic Peak Velocity |
(11726-7, LN, "Peak Velocity") |
(G-C048, SRT, "Direction of Flow") = (R-42E61, SRT, "Regurgitant Flow") |
Pulmonic Valve measurements appear in TID 5202 “Echo Section”, which specifies the Finding Site to be Pulmonic Valve with the concept modifier (G-C0E3, SRT, "Finding Site") = (T-35100, SRT, "Pulmonic Valve"). Therefore, this Finding Site concept modifier does not appear in the right column.
Name of ASE Concept |
Base Measurement Concept Name |
Concept or Acquisition Context Modifiers |
---|---|---|
Tricuspid Valve Mean Diastolic Velocity |
(20352-1, LN, "Mean Velocity") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") |
Tricuspid Valve E Wave Peak Velocity |
(18031-5, LN, "Tricuspid Valve E Wave Peak Velocity") |
|
Tricuspid Valve A Wave Peak Velocity |
(18030-7, LN, "Tricuspid Valve A Wave Peak Velocity") |
|
Tricuspid Valve Diastolic Velocity Time Integral |
(20354-7, LN, "Velocity Time Integral") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") |
Tricuspid Valve Peak Diastolic Gradient |
(20247-3, LN, Peak Gradient") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") |
Tricuspid Valve Mean Diastolic Gradient |
(20256-4, LN, Mean Gradient") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") |
Tricuspid Valve Annulus Diastolic Diameter |
(G-038F, SRT, Cardiovascular Orifice Diameter") |
(G-C0E3, SRT, "Finding Site") = (T-35111, SRT, "Tricuspid Annulus") (G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") |
Tricuspid Valve Regurgitant Peak Velocity |
(11726-7, LN, "Peak Velocity") |
(G-C048, SRT, "Direction of Flow") = (R-42E61, SRT, "Regurgitant Flow") |
Tricuspid Regurgitation Peak Pressure Gradient |
(20247-3, LN, "Peak Gradient") |
(G-C048, SRT, "Direction of Flow") = (R-42E61, SRT, "Regurgitant Flow") |
Tricuspid Regurgitation Velocity Time Integral |
(20354-7, LN, "Velocity Time Integral") |
(G-C048, SRT, "Direction of Flow") = (R-42E61, SRT, "Regurgitant Flow") |
Tricuspid Valve Deceleration Time |
(20217-6, LN, "Deceleration Time") |
(G-C048, SRT, "Direction of Flow") = (R-42047, SRT, "Antegrade Flow") |
TRICUSPID Valve measurements appear in TID 5202 “Echo Section”, which specifies the Finding Site to be Tricuspid Valve with the concept modifier (G-C0E3, SRT, "Finding Site") = (T-35100, SRT, "Tricuspid Valve"). Therefore, the Finding Site modifier does not appear in the right column.
Name of ASE Concept |
Base Measurement Concept Name |
Concept or Acquisition Context Modifiers |
---|---|---|
Right Atrium Systolic Pressure |
(18070-3, LN, "Right Atrium Systolic Pressure") |
|
Right Atrium Systolic Area |
(17988-7, LN, "Right Atrium Area A4C view") |
(R-4089A, SRT, "Cardiac Cycle Point") = (F-32020, SRT, "Systole") |
Inferior Vena Cava Diameter |
(18006-7, LN, "Inferior Vena Cava Diameter") |
|
Inferior Vena Cava Diameter at Inspiration |
(18006-7, LN, "Inferior Vena Cava Diameter") |
(R-40899, SRT, "Respiratory Cycle Point") = (F-20010, SRT, "During Inspiration") |
Inferior Vena Cava Diameter at Expiration |
(18006-7, LN, "Inferior Vena Cava Diameter") |
(R-40899, SRT, "Respiratory Cycle Point") = (F-20020, SRT, "During Expiration") |
Inferior Vena Cava % Collapse |
(18050-5, LN, "Inferior Vena Cava % Collapse") |
|
Hepatic Vein Systolic Peak Velocity |
(29471-0, LN, "Hepatic Vein Systolic Peak Velocity") |
|
Hepatic Vein Diastolic Peak Velocity |
(29472-8, LN, "Hepatic Vein Diastolic Peak Velocity") |
|
Hepatic Vein Systolic to Diastolic Ratio |
(29473-6, LN, "Hepatic Vein Systolic to Diastolic Ratio") |
|
Hepatic Vein Atrial Contraction Reversal Peak Velocity |
(29474-4, LN, "Hepatic Vein Atrial Contraction Reversal Peak Velocity") |
|
Hepatic Vein Peak Systolic Velocity at Inspiration |
(29471-0, LN, "Hepatic Vein Systolic Peak Velocity") |
(R-40899, SRT, "Respiratory Cycle Point") = (F-20010, SRT, "During Inspiration") |
Hepatic Vein Peak Diastolic Velocity at Inspiration |
(29472-8, LN, "Hepatic Vein Diastolic Peak Velocity") |
(R-40899, SRT, "Respiratory Cycle Point") = (F-20010, SRT, "During Inspiration") |
Hepatic Vein Systolic to Diastolic Ratio at Inspiration |
(29473-6, LN, "Hepatic Vein Systolic to Diastolic Ratio") |
(R-40899, SRT, "Respiratory Cycle Point") = (F-20010, SRT, "During Inspiration") |
Hepatic Vein Peak Atrial Contraction Reversal Velocity at Inspiration |
(29474-4, LN, "Hepatic Vein Atrial Contraction Reversal Peak Velocity") |
(R-40899, SRT, "Respiratory Cycle Point") = (F-20010, SRT, "During Inspiration") |
Hepatic Vein Peak Systolic Velocity at Expiration |
(29471-0, LN, "Hepatic Vein Systolic Peak Velocity") |
(R-40899, SRT, "Respiratory Cycle Point") = (F-20020, SRT, "During Expiration") |
Hepatic Vein Peak Diastolic Velocity at Expiration |
(29472-8, LN, "Hepatic Vein Diastolic Peak Velocity") |
(R-40899, SRT, "Respiratory Cycle Point") = (F-20020, SRT, "During Expiration") |
Hepatic Vein Systolic to Diastolic Ratio at Expiration |
(29473-6, LN, "Hepatic Vein Systolic to Diastolic Ratio") |
(R-40899, SRT, "Respiratory Cycle Point") = (F-20020, SRT, "During Expiration") |
Hepatic Vein Peak Atrial Contraction Reversal Velocity at Expiration |
(29474-4, LN, "Hepatic Vein Atrial Contraction Reversal Peak Velocity") |
(R-40899, SRT, "Respiratory Cycle Point") = (F-20020, SRT, "During Expiration") |
Name of ASE Concept |
Base Measurement Concept Name |
Concept or Acquisition Context Modifiers |
---|---|---|
Thoracic Aorta Coarctation Systolic Peak Velocity |
(29460-3, LN, "Thoracic Aorta Coarctation Systolic Peak Velocity") |
|
Thoracic Aorta Coarctation Systolic Peak Instantaneous Gradient |
(20256-4, LN, "Mean Gradient") |
(G-C0E3, SRT, "Finding Site") = (D4-32030, SRT, "Thoracic Aortic Coarctation") |
Thoracic Aorta Coarctation Systolic Mean Gradient |
(17995-2, LN, "Thoracic Aorta Coarctation Systolic Peak Instantaneous Gradient") |
|
Ventricular Septal Defect Diameter |
(G-038F, SRT, "Cardiovascular Orifice Diameter") |
(G-C0E3, SRT, "Finding Site") = (D4-31150, SRT, "Ventricular Septal Defect") |
Ventricular Septal Defect Systolic Peak Instantaneous Gradient |
(20247-3, LN, "Peak Gradient") |
(G-C0E3, SRT, "Finding Site") = (D4-31150, SRT, "Ventricular Septal Defect") |
Ventricular Septal Defect Systolic Mean Gradient |
(20256-4, LN, "Mean Gradient") |
(G-C0E3, SRT, "Finding Site") = (D4-31150, SRT, "Ventricular Septal Defect") |
Ventricular Septum Defect Systolic Peak Velocity |
(11726-7, LN, "Peak Velocity") |
(G-C0E3, SRT, "Finding Site") = (D4-31150, SRT, "Ventricular Septal Defect") |
Atrial Septal Defect Diameter |
(G-038F, SRT, "Cardiovascular Orifice Diameter") |
(G-C0E3, SRT, "Finding Site") = (D4-31220, SRT, "Atrial Septal Defect") |
Pulmonary-to-Systemic Shunt Flow Ratio |
(29462-9, LN, "Pulmonary-to-Systemic Shunt Flow Ratio") |
|
Pulmonary-to-Systemic Shunt Flow Ratio by Doppler Volume Flow |
(29462-9, LN, "Pulmonary-to-Systemic Shunt Flow Ratio") |
(G-C036, SRT, "Measurement Method") = (125219, DCM, "Doppler Volume Flow") |