Quantitative Ventricular Analysis Report SR IOD Templates 

The templates that comprise the Quantitative Ventricular Analysis SR are interconnected as in Figure A-3:

Quantitative Ventricular Analysis Report SR IOD Template Structure

Figure A-3. Quantitative Ventricular Analysis Report SR IOD Template Structure


TID 3202 Ventricular Analysis

The Ventricular Analysis Template provides a CONTAINER with a structure for reporting the result of the ventricular analysis.

Type:

Extensible

Order:

Significant

Table TID 3202. Ventricular Analysis

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (122292, DCM, "Quantitative Ventriculography Report")

1

M

2

>

HAS CONCEPT MOD

INCLUDE

DTID 1204 “Language of Content Item and Descendants”

1

M

3

>

HAS OBS CONTEXT

INCLUDE

DTID 1002 “Observer Context”

1-n

M

4

>

CONTAINS

INCLUDE

DTID 3602 “Cardiovascular Patient Characteristics”

1

U

5

>

CONTAINS

CONTAINER

EV (122144, DCM, "Quantitative Analysis")

1-n

M

6

>>

HAS OBS CONTEXT

INCLUDE

DTID 1002 “Observer Context”

1-n

U

7

>>

HAS OBS CONTEXT

TEXT

EV (111001, DCM, "Algorithm Name")

1

M

8

>>

HAS OBS CONTEXT

TEXT

EV (111003, DCM, "Algorithm Version")

1

M

9

>>

HAS OBS CONTEXT

TEXT

EV (122405, DCM, "Algorithm Manufacturer")

1

M

10

>>

CONTAINS

IMAGE

EV (121112, DCM, "Source of Measurement")

1-n

M

11

>>>

HAS CONCEPT MOD

CODE

EV (G-A60B, SRT, "Cardiac Phase")

1

M

DCID 12233 “Cardiac Phase”

12

>>>

HAS CONCEPT MOD

CODE

EV (111031, DCM, "Image View")

1

MC

If Biplane Analysis

DCID 3466 “Plane Identification”

13

>>

HAS ACQ CONTEXT

INCLUDE

DTID 3205 “Calibration”

1-2

U

VM = 1: Single plane analysis, VM = 2: Biplane analysis

$CalibrationPlane = DCID 3466 “Plane Identification”

14

>>

HAS ACQ CONTEXT

INCLUDE

DTID 3520 “Hemodynamic Clinical Context”

1

U

15

>>

CONTAINS

INCLUDE

DTID 3206 “VA Main Results”

1

M

16

>>

CONTAINS

INCLUDE

DTID 3207 “AA Main Results”

1

U

17

>>

CONTAINS

INCLUDE

DTID 3208 “Frame-to-Frame Results”

1

U

18

>>

CONTAINS

INCLUDE

DTID 3209 “Centerline Wall Motion”

1-2

U

VM = 1: Single plane analysis, VM = 2: Biplane analysis

19

>>

CONTAINS

INCLUDE

DTID 3210 “Radial Based Wall Motion”

1-2

U

VM = 1: Single plane analysis, VM = 2: Biplane analysis

20

>>

CONTAINS

INCLUDE

DTID 3211 “Landmark Based Wall Motion”

1-2

U

VM = 1: Single plane analysis, VM = 2: Biplane analysis


Content Item Descriptions

Row 7

Identifies the Ventricular Analysis program

Row 8

Identifies the Ventricular Analysis program version

Row 9

Identifies the Ventricular Analysis program manufacturer

Row 10

Identifies the ES and ED images on which the analysis is based, for frame by frame analysis the analyzed image are specified in the frame by frame results (3208) template

TID 3205 Calibration

The Calibration Template consists of a CONTAINER, with a structure for reporting of the calibration of images used in the analysis.

Table TID 3205. Parameters

Parameter Name

Parameter Usage

$CalibrationPlane

XA Imaging plane


Type:

Extensible

Order:

Significant

Table TID 3205. Calibration

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (122505, DCM, "Calibration")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (111031, DCM, "Image View")

1

U

$CalibrationPlane

3

>

HAS OBS CONTEXT

TEXT

EV (111001, DCM, "Algorithm Name")

1

MC

IF different from Analysis program specified in the invoking template

4

>

HAS OBS CONTEXT

TEXT

EV (111003, DCM, "Algorithm Version")

1

MC

IF different from Analysis program specified in the invoking template

5

>

HAS OBS CONTEXT

TEXT

EV (122405, DCM, "Algorithm Manufacturer")

1

MC

IF different from Analysis program specified in the invoking template

6

>

CONTAINS

CODE

EV (122422, DCM, "Calibration Method")

1

M

DCID 3452 “Calibration Methods”

7

>

CONTAINS

CODE

EV (122421, DCM, "Calibration Object")

1

MC

If row 6 value specifies Calibration Object Used

DCID 3451 “Calibration Objects”

8

>

CONTAINS

NUM

EV (122423, DCM, "Calibration Object Size")

1

MC

If row 6 value specifies Calibration Object Used

DCID 3510 “Catheter Size Units”

9

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (111026, DCM, "Horizontal Pixel Spacing")

$Unit = DT (mm/{pixel}, UCUM, "mm/pixel")

$ImagePurpose = EV (121112, DCM, "Source of Measurement")

10

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (111066, DCM, "Vertical Pixel Spacing")

$Unit = DT (mm/{pixel}, UCUM, "mm/pixel")

$ImagePurpose = EV (121112, DCM, "Source of Measurement")

11

>

CONTAINS

IMAGE

No purpose of reference

1

U


Content Item Descriptions

Row 3

Identifies the Calibration program

Row 4

Identifies the Calibration program version

Row 5

Identifies the Calibration program manufacturer

Row 7

Besides a Sphere and a Catheter, a Distance can be identified as a Calibration Object. In this case a distance measurement of a known dimension of the object is used to calculate the pixel size.

Row 8

The catheter size units is also used to specify the size of other calibration objects (e.g., sphere)

Row 9, 10

Spacing in the patient body. Point to a single frame containing the image used for calibration if applicable, the actual measurements may be indicated by a SCOORD (see TID 320 “Image or Spatial Coordinates”, row 3)

Row 11

Secondary Capture image with calibration position

TID 3206 VA Main Results

The VA Main Results Template consists of a CONTAINER with a structure for reporting the main ventricular analysis measurements.

Type:

Extensible

Order:

Significant

Table TID 3206. VA Main Results

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (121070, DCM, "Findings")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (G-C0E3, SRT, "Finding Site")

1

M

DCID 3462 “Chamber Identification”

3

>

CONTAINS

CODE

EV (122429, DCM, "Volume Method")

1

M

DCID 3453 “Cardiac Volume Methods”

4

>

CONTAINS

NUM

EV (122435, DCM, "Regression Volume Exponent")

1

U

Unit = DT (1, UCUM, "no units")

5

>

CONTAINS

NUM

EV (122431, DCM, "Regression Slope ED")

1

U

Unit = DT ({ratio}, UCUM, "ratio")

6

>

CONTAINS

NUM

EV (122432, DCM, "Regression Offset ED")

1

U

Unit = DT (ml, UCUM, "ml")

7

>

CONTAINS

NUM

EV (122433, DCM, "Regression Slope ES")

1

U

Unit = DT ({ratio}, UCUM, "ratio")

8

>

CONTAINS

NUM

EV (122434, DCM, "Regression Offset ES")

1

U

Unit = DT (ml, UCUM, "ml")

9

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = DCID 3467 “Ejection Fraction”

$Unit = DT (%, UCUM, "%")

10

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = DCID 3468 “ED Volume”

$Unit = DT (ml, UCUM, "ml")

11

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = DCID 3469 “ES Volume”

$Unit = DT (ml, UCUM, "ml")

12

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (F-32120, SRT, "Stroke Volume")

$Unit = DT (ml, UCUM, "ml")

13

>

CONTAINS

NUM

EV (8867-4, LN, "Heart Rate")

1

U

Unit = DT ({H.B.}/min, UCUM, "BPM")

14

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = DCID 3468 “ED Volume”

$ModType = EV (121425, DCM, "Index")

$ModValue = DCID 3455 “Index Methods”

$Unit = DT (ml/m2, UCUM, "ml/m2")

15

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = DCID 3468 “ED Volume”

$ModType = EV (121425, DCM, "Index")

$ModValue = EV (29463-7, LN, "Patient Weight")

$Unit = DT (ml/kg, UCUM, "ml/kg")

16

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = DCID 3469 “ES Volume”

$ModType = EV (121425, DCM, "Index")

$ModValue = DCID 3455 “Index Methods”

$Unit = DT (ml/m2, UCUM, "ml/m2")

17

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = DCID 3469 “ES Volume”

$ModType = EV (121425, DCM, "Index")

$ModValue = EV (29463-7, LN, "Patient Weight")

$Unit = DT (ml/kg, UCUM, "ml/kg")

18

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (F-32120, SRT, "Stroke Volume")

$ModType = EV (121425, DCM, "Index")

$ModValue = DCID 3455 “Index Methods”

$Unit = DT (ml/m2, UCUM, "ml/m2")

19

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (F-32120, SRT, "Stroke Volume")

$ModType = EV (121425, DCM, "Index")

$ModValue = EV (29463-7, LN, "Patient Weight")

$Unit = DT (ml/kg, UCUM, "ml/kg")

20

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (F-32100, SRT, "Cardiac Output")

$Unit = DT (l/min, UCUM, "l/min")

21

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (F-32110, SRT, "Cardiac Index")

$ModType = EV (121425, DCM, "Index")

$ModValue = DCID 3455 “Index Methods”

$Unit = DT (l/min/m2, UCUM, "l/min/m2")

22

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122445, DCM, "Wall Thickness")

$Unit = DT (mm, UCUM, "mm")

23

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122446, DCM, "Wall Volume")

$Unit = DT (ml, UCUM, "ml")

24

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122447, DCM, "Wall Mass")

$Unit = DT (g, UCUM, "gram")

25

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122447, DCM, "Wall Mass")

$ModType = EV (121425, DCM, "Index")

$ModValue = DCID 3455 “Index Methods”

$Unit = DT (g/m2, UCUM, "gram/m2")

26

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122447, DCM, "Wall Mass")

$ModType = EV (121425, DCM, "Index")

$ModValue = EV (29463-7, LN, "Patient Weight")

$Unit = DT (g/kg, UCUM, "gram/kg")

27

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122448, DCM, "Wall Stress")

$Unit = DT (dyn/cm2, UCUM, "dynes/cm2")

28

>

CONTAINS

IMAGE

No purpose of reference

1-n

U


Content Item Descriptions

Row 28

Secondary Capture image with ED and/or ES contours

TID 3207 AA Main Results

The AA Main Results Template consists of a CONTAINER with a structure for reporting the main atrial analysis measurements.

Type:

Extensible

Order:

Significant

Table TID 3207. AA Main Results

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (121070, DCM, "Findings")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (G-C0E3, SRT, "Finding Site")

1

M

DCID 3462 “Chamber Identification”

3

>

CONTAINS

CODE

EV (122429, DCM, "Volume Method")

1

M

DCID 3453 “Cardiac Volume Methods”

4

>

CONTAINS

NUM

EV (122435, DCM, "Regression Volume Exponent")

1

U

Unit = DT (1, UCUM, "no units")

5

>

CONTAINS

NUM

EV (122431, DCM, "Regression Slope ED")

1

U

Unit = DT ({ratio}, UCUM, "ratio")

6

>

CONTAINS

NUM

EV (122432, DCM, "Regression Offset ED")

1

U

Unit = DT (ml, UCUM, "ml")

7

>

CONTAINS

NUM

EV (122433, DCM, "Regression Slope ES")

1

U

Unit = DT ({ratio}, UCUM, "ratio")

8

>

CONTAINS

NUM

EV (122434, DCM, "Regression Offset ES")

1

U

Unit = DT (ml, UCUM, "ml")

9

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = DCID 3468 “ED Volume”

$Unit = DT (ml, UCUM, "ml")

10

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = DCID 3469 “ES Volume”

$Unit = DT (ml, UCUM, "ml")

11

>

CONTAINS

IMAGE

No purpose of reference

1-n

U


Content Item Descriptions

Row 11

Secondary Capture image with ED and/or ES contours

TID 3208 Frame-to-Frame Results

The Frame-to-Frame Result Template consists of a CONTAINER providing measurements derived from the angiographic images on frame-by-frame basis.

Type:

Extensible

Order:

Significant

Table TID 3208. Frame-To-Frame Result

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (121070, DCM, "Findings")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (111004, DCM, "Analysis Performed")

1

M

EV (122499, DCM, "Frame to Frame Analysis")

3

>

CONTAINS

IMAGE

EV (121112, DCM, "Source of Measurement")

1-2

M

VM = 1: Single plane analysis, VM = 2: Biplane analysis

4

>

CONTAINS

CODE

EV (122429, DCM, "Volume Method")

1

M

DCID 3453 “Cardiac Volume Methods”

5

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

M

$Measurement = DCID 3471 “Estimated Volumes”

$TargetSite = DCID 3462 “Chamber Identification”

$Unit = DT (ml, UCUM, "ml")

6

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = EV (122445, DCM, "Wall Thickness")

$Unit = DT (mm, UCUM, "mm")

7

>

CONTAINS

IMAGE

No purpose of reference

1-n

U


Content Item Descriptions

Row 3

Identifies each frame analyzed, using the multi-valued Referenced Frame Number (0008,1160) attribute of the IMAGE content item.

Row 5, 6

Includes one measurement for each frame referenced in Row 3.

Row 7

Secondary Capture image with ventricular contours

TID 3209 Centerline Wall Motion

The Centerline Wall Motion Template consists of a CONTAINER providing measurements of the centerline wall motion.

Type:

Extensible

Order:

Significant

Table TID 3209. Centerline Wall Motion

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (121070, DCM, "Findings")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (111004, DCM, "Analysis Performed")

1

M

EV (122449, DCM, "Centerline Wall Motion Analysis")

3

>>

HAS CONCEPT MOD

CODE

EV (122410, DCM, "Contour Realignment")

1

M

DCID 3458 “Contour Realignment”

4

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

100

M

$Measurement = EV (122450, DCM, "Normalized Chord Length")

$Unit = DT (%, UCUM, "%")

5

>

CONTAINS

NUM

EV (122411, DCM, "Threshold Value")

1

M

Values shall be 1, 2 or 3

UNITS = EV ({sd}, UCUM, "Standard Deviations")

6

>

CONTAINS

CONTAINER

EV (122451, DCM, "Abnormal Region")

1-6

U

7

>>

CONTAINS

CODE

EV (F-32050, SRT, "Cardiac Wall Motion")

1

M

DCID 3703 “Wall Motion”

8

>>

CONTAINS

CODE

EV (R-404F0, SRT, "Circumferential Extent")

1

U

DCID 3460 “Circumferential Extent”

9

>>

CONTAINS

NUM

EV (122452, DCM, "First Chord of Abnormal Region")

1

M

Unit = DT (1, UCUM, "no unit")

10

>>

CONTAINS

NUM

EV (122453, DCM, "Last Chord of Abnormal Region")

1

M

Unit = DT (1, UCUM, "no unit")

11

>

CONTAINS

CONTAINER

EV (122417, DCM, "Regional Abnormal Wall Motion ")

1-4

U

12

>>

HAS CONCEPT MOD

CODE

EV (G-C03E, SRT, "Finding Site")

1

M

DCID 3461 “Regional Extent”

13

>>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (122459, DCM, "Territory Region Severity")

$ModType = EV (F-32050, SRT, "Cardiac Wall Motion")

$ModValue = DCID 3703 “Wall Motion”

$Unit = DT ({sd}, UCUM, "Standard Deviations")

14

>>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (122461, DCM, "Opposite Region Severity")

$ModType = EV (F-32050, SRT, "Cardiac Wall Motion")

$ModValue = DCID 3703 “Wall Motion”

$Unit = DT ({sd}, UCUM, "Standard Deviations")

15

>

CONTAINS

IMAGE

No purpose of reference

1

U


Content Item Descriptions

Row 4

Normalized lengths of the chords determined between ED and ES contour. The measurement template allows the specification of the statistical properties of the normal population and of the chord measurement relative to the population.

Row 8

If the Circumferential Extent is not specified no limitations to the boundaries for regions are assumed.

Row 11

The Regional Abnormal Wall Motion container holds the information on the severity of the decreased or increased wall motion of the 4 predefined regions as described in [Sheehan, 1986].

Row 12

The name of the region with an abnormal ventricular wall motion as described in [Sheehan, 1986].

Row 13

The severity of the wall motion abnormality expressed in Standard Deviations above or below normal in the territory region as described in [Sheehan, 1986].

Row 14

The severity of the wall motion abnormality expressed in Standard Deviations above or below normal in the opposite region as described in [Sheehan, 1986].

Row 15

Secondary Capture image with centerline analysis result

TID 3210 Radial Based Wall Motion

The Radial Based Wall Motion Template consists of a CONTAINER providing measurements of the radial based wall motion.

Type:

Extensible

Order:

Significant

Table TID 3210. Radial Based Wall Motion

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (121070, DCM, "Findings")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (111004, DCM, "Analysis Performed")

1

M

EV (122493, DCM, "Radial Based Wall Motion Analysis")

3

>>

HAS CONCEPT MOD

CODE

EV (122410, DCM, "Contour Realignment")

1

M

DCID 3458 “Contour Realignment”

4

>

CONTAINS

CONTAINER

EV (121070, DCM, "Findings")

1-n

M

5

>>

HAS CONCEPT MOD

CODE

EV (G-C0E3, SRT, "Finding Site")

1

M

DCID 3718 “Myocardial Wall Segments in Projection”

6

>>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (122495, DCM, "Regional Contribution to Ejection Fraction")

$Unit = DT (%, UCUM, "%")

7

>>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = EV (122496, DCM, "Radial Shortening")

$Unit = DT (%, UCUM, "%")

8

>

CONTAINS

IMAGE

No purpose of reference

1

U


Content Item Descriptions

Row 6

The CREF values of the 6 regions determined for the radial based wall motion

Row 7

The shortening of the measured radials within the region

Row 8

Secondary Capture image with radial based analysis result

TID 3211 Landmark Based Wall Motion

The Landmark Based Wall Motion Template consists of a CONTAINER providing measurements of the landmark based wall motion.

Type:

Extensible

Order:

Significant

Table TID 3211. Landmark Based Wall Motion

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (121070, DCM, "Findings")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (111004, DCM, "Analysis Performed")

1

M

EV (122497, DCM, "Landmark Based Wall Motion Analysis")

3

>>

HAS CONCEPT MOD

CODE

EV (122410, DCM, "Contour Realignment")

1

M

DCID 3458 “Contour Realignment”

4

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

M

$Measurement = EV (122498, DCM, "Slice Contribution to Ejection Fraction")

$Unit = DT (%, UCUM, "%")

5

>

CONTAINS

CONTAINER

EV (121070, DCM, "Findings")

5

M

6

>>

HAS CONCEPT MOD

CODE

EV (G-C0E3, SRT, "Finding Sites")

1

M

DCID 3718 “Myocardial Wall Segments in Projection”

7

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (122495, DCM, "Regional Contribution to Ejection Fraction")

$Unit = DT (%, UCUM, "%")

8

>

CONTAINS

IMAGE

No purpose of reference

1

U


Content Item Descriptions

Row 8

Secondary Capture image with Landmark Based Analysis result