This template forms the top of a content tree that allows an ultrasound device to describe the results of an adult echocardiography imaging procedure. It is instantiated at the root node. It can also be included in other templates that need to incorporate echocardiography findings into another report as quoted evidence.
Extensible
Significant
Yes
Table TID 5200. Echocardiography Procedure Report
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
Root node |
||||
2 |
> |
HAS CONCEPT MOD |
INCLUDE |
1 |
U |
|||
3 |
> |
HAS OBS CONTEXT |
INCLUDE |
1 |
M |
|||
4 |
> |
CONTAINS |
CONTAINER |
1 |
U |
|||
5 |
>> |
CONTAINS |
CODE |
1-n |
M |
|||
6 |
> |
CONTAINS |
INCLUDE |
1 |
U |
|||
7 |
> |
CONTAINS |
CONTAINER |
1 |
U |
|||
8 |
>> |
CONTAINS |
IMAGE |
No purpose of reference |
1-n |
M |
||
9 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$SectionSubject = EV (T-32600, SRT, "Left Ventricle") $MeasType = DCID 12200 “Echocardiography Left Ventricle” |
||
10 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$SectionSubject = EV (T-32500, SRT, "Right Ventricle") $MeasType = DCID 12204 “Echocardiography Right Ventricle” |
||
11 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$SectionSubject = EV (T-32300, SRT, "Left Atrium") $MeasType = DCID 12205 “Echocardiography Left Atrium” |
||
12 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$SectionSubject = EV (T-32200, SRT, "Right Atrium") $MeasType = DCID 12206 “Echocardiography Right Atrium” |
||
13 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$SectionSubject = EV (T-35400, SRT, "Aortic Valve") $MeasType = DCID 12211 “Echocardiography Aortic Valve” |
||
14 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$SectionSubject = EV (T-35300, SRT, "Mitral Valve") $MeasType = DCID 12207 “Echocardiography Mitral Valve” |
||
15 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$SectionSubject = EV (T-35200, SRT, "Pulmonic Valve") $MeasType = DCID 12209 “Echocardiography Pulmonic Valve” |
||
16 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$SectionSubject = EV (T-35100, SRT, "Tricuspid Valve") $MeasType = DCID 12208 “Echocardiography Tricuspid Valve” |
||
17 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$SectionSubject = EV (T-42000, SRT, "Aorta") $MeasType = DCID 12212 “Echocardiography Aorta” |
||
18 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$SectionSubject = EV (T-44000, SRT, "Pulmonary artery") $MeasType = DCID 12210 “Echocardiography Pulmonary Artery” |
||
19 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$SectionSubject = EV (T-48600, SRT, "Vena Cava") $MeasType = DCID 12215 “Echocardiography Vena Cavae” |
||
20 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$SectionSubject = EV (T-48581, SRT, "Pulmonary Venous Structure") $MeasType = DCID 12214 “Echocardiography Pulmonary Veins” |
||
21 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$SectionSubject = EV (T-39050, SRT, "Pericardial cavity") $MeasType = DCID 12250 “Cardiac Ultrasound Common Linear Measurements” |
||
22 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$SectionSubject = EV (P5-30031, SRT, "Cardiac Shunt Study") $MeasType = DCID 12217 “Echocardiography Cardiac Shunt” |
||
23 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$SectionSubject = EV (D4-30000, SRT, "Congenital Anomaly of Cardiovascular System") $MeasType = DCID 12218 “Echocardiography Congenital” |
||
24 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
$Procedure = DT (P5-B3121, SRT, "Echocardiography for Determining Ventricular Contraction") |
Content Item Descriptions
Row 24 |
The wall motion findings of stress stage. There may be multiple Template instances to report wall motion findings of multiple stages. |
Patient Characteristic concepts in this template, which may replicate attributes in the Patient Study Module, are included here as possible targets of by-reference relationships from other content items in the SR tree.
Several of the concepts in this template duplicate concepts in TID 1007 “Subject Context, Patient”. The difference in use is that this template has those concepts as primary observations of the patient, while in TID 1007 “Subject Context, Patient” the concepts are used to set (or reset) the context for other observations.
Extensible
Significant
Table TID 5201. Echocardiography Patient Characteristics
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
|||||
2 |
> |
CONTAINS |
NUM |
1 |
U |
UNITS = DCID 7456 “Units of Measure for Age” |
||
3 |
> |
CONTAINS |
CODE |
1 |
U |
|||
4 |
> |
CONTAINS |
NUM |
1 |
U |
|||
5 |
> |
CONTAINS |
NUM |
EV (F-008EC, SRT, "Systolic Blood Pressure") |
1 |
U |
||
6 |
> |
CONTAINS |
NUM |
EV (F-008ED, SRT, "Diastolic Blood Pressure") |
1 |
U |
||
7 |
> |
CONTAINS |
NUM |
1 |
M |
|||
8 |
>> |
INFERRED FROM |
CODE |
1 |
U |
This is a generic section heading Template for any of the anatomical headings. Measurements within a section heading appear as groups (by image mode, acquisition protocol, and/or protocol stage).
Table TID 5202. Parameters
Parameter Name |
Parameter Usage |
---|---|
$SectionSubject |
The subject modifier of the section heading container |
$MeasType |
The concept name of the measurement |
Extensible
Significant
Table TID 5202. Echo Section
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
|||||
2 |
> |
HAS CONCEPT MOD |
CODE |
EV (G-C0E3, SRT, "Finding Site") |
1 |
M |
$SectionSubject |
|
3 |
> |
CONTAINS |
CONTAINER |
1-n |
M |
|||
4 |
>> |
HAS CONCEPT MOD |
CODE |
EV (G-0373, SRT, "Image Mode") |
1 |
U |
||
5 |
>> |
HAS CONCEPT MOD |
CODE |
1 |
U |
|||
6 |
>> |
HAS CONCEPT MOD |
TEXT |
1 |
U |
|||
7 |
>> |
HAS ACQ CONTEXT |
CODE |
1 |
U |
|||
8 |
>> |
CONTAINS |
INCLUDE |
1-n |
M |
$Measurement = $MeasType |
Content Item Descriptions
Rows 4, 5 |
Type of measurement group. May be grouped by image mode, or acquisition protocol, or some other user or manufacturer designated classification |
Row 7 |
For measurements acquired in a staged protocol, all measurements in a measurement group are acquired at the identified stage. |
Table TID 5203. Parameters
Parameter Name |
Parameter Usage |
---|---|
$Measurement |
Coded term or Context Group for Concept Name of measurement |
$Method |
Value for Measurement Method |
Extensible
Significant
Table TID 5203. Echo Measurement
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
INCLUDE |
1 |
M |
$Measurement = $Measurement $Method = $Method $TargetSite = BCID 12236 “Echo Anatomic Sites” $TargetSiteMod = BCID 12237 “Echocardiography Anatomic Site Modifiers” |
||||
2 |
> |
HAS CONCEPT MOD |
CODE |
EV (G-C048, SRT, "Flow Direction") |
1 |
U |
||
3 |
> |
HAS CONCEPT MOD |
CODE |
EV (R-40899, SRT, "Respiratory Cycle Point") |
1 |
U |
||
4 |
> |
HAS CONCEPT MOD |
CODE |
EV (R-4089A, SRT, "Cardiac Cycle Point") |
1 |
U |
||
5 |
> |
HAS ACQ CONTEXT |
CODE |
EV (G-0373, SRT, "Image Mode") |
1 |
U |
||
6 |
> |
HAS ACQ CONTEXT |
CODE |
1 |
U |
|||
7 |
> |
HAS ACQ CONTEXT |
CODE |
1 |
U |
Content Item Descriptions
Row 1 |
TID 300 specifies an "Equivalent Meaning of Concept Name" that allows the creating application to specify the preferred composed concept name representing the measurement and the associated post-coordination Concept Modifiers (e.g., the ASE terminology described in Section N.3 “Illustrative Mapping to ASE Concepts” in PS3.17 ). |
The Wall Motion Analysis Template is used to document wall motion scoring for any imaging modality.
Table TID 5204. Parameters
Parameter Name |
Parameter Usage |
---|---|
$Procedure |
The imaging procedure used for wall motion analysis. |
Extensible
Significant
Table TID 5204. Wall Motion Analysis
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
|||||
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
M |
$Procedure |
||
3 |
> |
HAS ACQ CONTEXT |
CODE |
1 |
U |
|||
4 |
> |
CONTAINS |
IMAGE |
1 |
U |
|||
5 |
> |
CONTAINS |
TEXT |
1 |
U |
|||
6 |
> |
CONTAINS |
NUM |
1 |
U |
|||
7 |
>> |
HAS CONCEPT MOD |
CODE |
EV (G-E048, SRT, "Assessment Scale") |
1 |
M |
||
8 |
> |
CONTAINS |
CONTAINER |
1 |
U |
|||
9 |
>> |
HAS CONCEPT MOD |
CODE |
EV (G-C0E3, SRT, "Finding Site") |
1 |
M |
DT (T-D0772, SRT, "Myocardial Wall") |
|
10 |
>> |
CONTAINS |
CODE |
1-n |
M |
|||
11 |
>>> |
HAS PROPERTIES |
CODE |
EV (F-32050, SRT, "Cardiac Wall Motion") |
1 |
MC |
IF row 12 is absent |
|
12 |
>>> |
HAS PROPERTIES |
CODE |
EV (G-C504, SRT, "Associated Morphology") |
1 |
MC |
IF row 11 is absent |
|
13 |
>>> |
HAS PROPERTIES |
NUM |
DT (G-C1E3, SRT, "Score") |
1 |
U |
||
14 |
>>> |
HAS PROPERTIES |
NUM |
EV (122624, DCM, "Wall Thickness Ratio end-systolic to end-diastolic") |
1 |
U |
UNITS = DT (%, UCUM, "%") |
Content Item Descriptions
Row 3 |
The stage of the protocol at which these findings were scored. This row may be absent if this is a generic, non-staged scoring. |
Row 4 |
Image that graphically depicts the segments and their scores. |
Row 5 |
Text narration accompanying this stage. |
Row 6 |
The composite score computed from the average of the scored segments |
Row 7 |
The type of scoring scheme used to score this exam. |
Row 8 |
A container of all of the individual segment findings for this stage. The container shall be present if the observer makes an assessment, including the assessment of Not Visualized. It shall not be present if no evaluation was made. |
Rows 11, 12 |
Scar/thinning (in Row 12) may accompany akinesis and dyskinesis (in Row 11). |
Row 13 |
A numeric designation for the score. Score ranges vary, typically 0-4 or 0-5. Numeric scores may depend on wall motion findings as well as morphology findings. See Table 5204-1 for conventional numeric assignment schemes. The UCUM annotation code enables specifying the numeric range, ({L:N}, UCUM, "scale L:N"), where L and N are the lower and upper ends of the range. |
A description of the scoring schemes described in Table 5204-1 is available in Recommendations for Quantitation of the Left Ventricle by Two-Dimensional Echocardiography, Journal of the American Society of Echocardiography, Vol 2, No 5 358-367, Oct 1989.
Table 5204-1. Numeric Score Assignment for Segmental Findings
Conventional Numeric Assignment |
Wall Motion Finding or Morphology Finding |
||
---|---|---|---|
4 Point |
5 Point |
5 Point with Graded Hypokinesis |
|
-1 |
(F-32050, SRT, "Cardiac Wall Motion") = (R-0030D SRT, "Hyperkinesis") |
(F-32050, SRT, "Cardiac Wall Motion") = (R-0030D, SRT, "Hyperkinesis") |
(F-32050, SRT, "Cardiac Wall Motion") = (R-0030D, SRT, "Hyperkinesis") |
0 |
(F-32050, SRT, "Cardiac Wall Motion") = (122288, DCM, "Not Visualized") |
(F-32050, SRT, "Cardiac Wall Motion") = (122288, DCM, "Not Visualized") |
(F-32050, SRT, "Cardiac Wall Motion") = (122288, DCM, "Not Visualized") |
1 |
(F-32050, SRT, "Cardiac Wall Motion") = (R-00344, SRT, "Normal Wall Motion") |
(F-32050, SRT, "Cardiac Wall Motion") = (R-00344, SRT, "Normal Wall Motion") |
(F-32050, SRT, "Cardiac Wall Motion") = (R-00344, SRT, "Normal Wall Motion") |
1.5 |
(F-32050, SRT, "Cardiac Wall Motion") = (R-00327, SRT, "Mild Hypokinesis") |
||
2 |
(F-32050, SRT, "Cardiac Wall Motion") = (F-32056, SRT, "Hypokinesis") |
(F-32050, SRT, "Cardiac Wall Motion") = (F-32056, SRT, "Hypokinesis") |
(F-32050, SRT, "Cardiac Wall Motion") = (R-0032F, SRT, "Moderate Hypokinesis") |
2.5 |
(F-32050, SRT, "Cardiac Wall Motion") = (R-00370, SRT, "Severe Hypokinesis") |
||
3 |
(F-32050, SRT, "Cardiac Wall Motion") = (F-30004, SRT, "Akinesis") |
(F-32050, SRT, "Cardiac Wall Motion") = (F-30004, SRT, "Akinesis") |
(F-32050, SRT, "Cardiac Wall Motion") = (F-30004, SRT, "Akinesis") |
4 |
(F-32050, SRT, "Cardiac Wall Motion") = (F-32052, SRT, "Dyskinesis") |
(F-32050, SRT, "Cardiac Wall Motion") = (F-32052, SRT, "Dyskinesis") |
(F-32050, SRT, "Cardiac Wall Motion") = (F-32052, SRT, "Dyskinesis") |
5 |
(G-C504, SRT, "Associated Morphology") = (D3-10510, SRT, "Ventricular Aneurysm") |
(G-C504, SRT, "Associated Morphology") = (D3-10510, SRT, "Ventricular Aneurysm") |
This template forms the top of a content tree that allows an ultrasound application to describe the results of a Cardiac Ultrasound imaging procedure. It is instantiated at the root node.
Extensible
Significant
Yes
Table TID 5220. Pediatric, Fetal and Congenital Cardiac Ultrasound Reports
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
Root node |
||||
2 |
> |
HAS CONCEPT MOD |
INCLUDE |
1 |
M |
|||
3 |
> |
HAS OBS CONTEXT |
INCLUDE |
1 |
M |
|||
4 |
> |
CONTAINS |
CONTAINER |
1 |
U |
|||
5 |
>> |
CONTAINS |
CODE |
1-n |
U |
|||
6 |
>> |
CONTAINS |
TEXT |
1 |
U |
|||
7 |
> |
CONTAINS |
INCLUDE |
1 |
U |
|||
8 |
> |
CONTAINS |
INCLUDE |
1 |
U |
|||
9 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
No more than one inclusion per fetus |
||
10 |
> |
CONTAINS |
INCLUDE |
1 |
U |
|||
11 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
No more than one inclusion per fetus |
||
12 |
> |
CONTAINS |
CONTAINER |
1 |
U |
|||
13 |
>> |
CONTAINS |
IMAGE |
No purpose of reference |
1-n |
M |
||
14 |
> |
CONTAINS |
INCLUDE |
DTID 5221 “Cardiac Ultrasound Pediatric Echo Measurement Section” |
1 |
U |
||
15 |
> |
CONTAINS |
INCLUDE |
1-n |
UC |
For Fetal Report only. |
No more than one inclusion per fetus |
Content Item Descriptions
Row 3 |
For Fetal Report, this row establishes the subject context of the mother. |
Row 7 |
For Fetal Report, this row will be the patient history of the mother. |
Row 8 |
For Fetal Report, this row will be the Patient Characteristics for the mother. |
Row 10 |
For Fetal Report, this row will be the Summary Section for the mother. |
Extensible
Significant
Table TID 5221. Cardiac Ultrasound Pediatric Echo Measurement Section
This is a generic section heading Template for any of the anatomical headings. Measurements within a section heading appear as groups (by image mode or acquisition protocol).
Table TID 5222. Parameters
Parameter Name |
Parameter Usage |
---|---|
$SectionSubject |
The subject modifier of the section heading container |
$MeasType |
The concept name of the measurement |
Extensible
Significant
Table TID 5222. Pediatric, Fetal and Congenital Cardiac Ultrasound Section
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
|||||
2 |
> |
HAS CONCEPT MOD |
CODE |
EV (G-C0E3, SRT, "Finding Site") |
1 |
M |
$SectionSubject |
|
3 |
>> |
HAS CONCEPT MOD |
CODE |
EV (P1-32006, SRT, "Heart valve replacement - prosthesis") |
1 |
U |
||
4 |
> |
CONTAINS |
CONTAINER |
1-n |
M |
|||
5 |
>> |
HAS CONCEPT MOD |
CODE |
EV (G-0373, SRT, "Image Mode") |
1 |
U |
||
6 |
>> |
HAS CONCEPT MOD |
TEXT |
1 |
U |
|||
7 |
>> |
CONTAINS |
INCLUDE |
DTID 5223 “Pediatric, Fetal and Congenital Cardiac Ultrasound Measurement” |
1-n |
M |
$Measurement = $MeasType |
This Template provides for the post-coordination of a measurement with a variety of concept modifiers and acquisition context observations. When invoked from TID 5222 “Pediatric, Fetal and Congenital Cardiac Ultrasound Section”, the measurement concept is implicitly post-coordinated with the concept modifiers of the Measurement Group (TID 5222 “Pediatric, Fetal and Congenital Cardiac Ultrasound Section” Rows 5 and 6), and with the Finding Site of the report section (TID 5222 “Pediatric, Fetal and Congenital Cardiac Ultrasound Section” Row 2). The finding site may be further specified within this Template by the Target Site and Target Site Modifiers (CID 12280 “Cardiac Ultrasound Target Sites” and CID 12281 “Cardiac Ultrasound Target Site Modifiers”).
The implicit finding site inherited from TID 5222 “Pediatric, Fetal and Congenital Cardiac Ultrasound Section” can be made explicit by using the same finding site concept in the Target Site (the measurement concept modifier), rather than a term from CID 12280 “Cardiac Ultrasound Target Sites”. This explicit post-coordination allows the use of one of the modifiers of CID 12281 “Cardiac Ultrasound Target Site Modifiers” to that finding site, as the Target Site Modifier requires an explicit Target Site in the measurement structure (TID 300 “Measurement” Rows 5 and 7). In fact, any child concept of the finding site in the SNOMED hierarchy may be used as the measurement Target Site.
The finding or target site may be identified by a concept from the SNOMED "clinical finding" or "morphological anomaly" hierarchies (e.g., D4-31220 "Atrial Septal Defect", or M-36700 "Effusion"), rather than the "anatomical structure" hierarchy. In this case, the meaning is inferred as "the anatomic location of the clinical finding or morphological anomaly, within the constraints of other implicit or explicit post-coordinated finding site concepts."
Thus when TID 5221 “Cardiac Ultrasound Pediatric Echo Measurement Section” Row 14 invokes TID 5222 “Pediatric, Fetal and Congenital Cardiac Ultrasound Section” with the section finding site concept (T-39000, SRT, "Pericardium"), and TID 5223 “Pediatric, Fetal and Congenital Cardiac Ultrasound Measurement” Row 1 applies the target site (M-36700, SRT, "Effusion"), the effective finding site is "pericardial effusion".
Table TID 5223. Parameters
Parameter Name |
Parameter Usage |
---|---|
$Measurement |
Coded term or Context Group for Concept Name of measurement |
$Method |
Value for Measurement Method |
Extensible
Significant
Table TID 5223. Pediatric, Fetal and Congenital Cardiac Ultrasound Measurement
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
INCLUDE |
1 |
M |
$Measurement = $Measurement $Method = $Method $TargetSite = BCID 12280 “Cardiac Ultrasound Target Sites” $TargetSiteMod = BCID 12281 “Cardiac Ultrasound Target Site Modifiers” $Derivation = DCID 3838 “Diameter Derivation” |
||||
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
U |
|||
3 |
> |
HAS CONCEPT MOD |
CODE |
EV (G-C048, SRT, "Flow Direction") |
1 |
U |
||
4 |
> |
HAS CONCEPT MOD |
CODE |
EV (R-40899, SRT, "Respiratory Cycle Point") |
1 |
U |
||
5 |
> |
HAS CONCEPT MOD |
CODE |
EV (R-4089A, SRT, "Cardiac Cycle Point") |
1 |
U |
||
6 |
> |
HAS ACQ CONTEXT |
CODE |
EV (G-0373, SRT, "Image Mode") |
1 |
U |
||
7 |
> |
HAS ACQ CONTEXT |
CODE |
1 |
U |
Content Item Descriptions
Row 1 |
For an index type of measurement, the concept name of this row 1 will still be the original measurement concept name; it is row 2 that gives the indication that row 1 is actually an index type of measurement. When this happens, the measurement value of row 1 should be a value after being indexed and the measurement unit of row 1 should be an index type of unit. For example, to insert a "Stroke Volume Index" measurement to this SR object, the concept name of row 1 will be "Stroke Volume", its numerical value will be the calculation result of "Stroke Volume /BSA" and its units are "ml/cm2". |
Row 2 |
When this row is available, the row 1 is an index calculation of the object. |
Contains a list of Fetus Specific characteristics.
Extensible
Significant
Table TID 5225. Cardiac Ultrasound Fetal Characteristics
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
|||||
2 |
> |
HAS OBS CONTEXT |
INCLUDE |
1 |
MC |
IF this template is invoked more than once to describe more than one fetus. |
||
3 |
> |
CONTAINS |
NUM |
1 |
U |
UNITS = DCID 7456 “Units of Measure for Age” |
||
4 |
> |
CONTAINS |
DATE |
1 |
U |
|||
5 |
> |
CONTAINS |
NUM |
1 |
U |
Comments and observations of the procedure of immediate clinical interest.
Extensible
Significant
Table TID 5226. Cardiac Ultrasound Summary Section
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
|||||
2 |
> |
CONTAINS |
CODE |
1-n |
MC |
IF row 3 does not exist |
||
3 |
> |
CONTAINS |
TEXT |
1-n |
MC |
IF row 2 does not exist |
||
4 |
> |
CONTAINS |
CODE |
EV (P0-009C3, SRT, "Surgical Procedure") |
1-n |
U |
BCID 12247 “Pediatric, Fetal and Congenital Cardiac Surgical Interventions” |
Comments and observations of the procedure of immediate clinical interest.
Extensible
Significant
Table TID 5227. Cardiac Ultrasound Fetal Summary Section
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
|||||
2 |
> |
HAS OBS CONTEXT |
INCLUDE |
1 |
MC |
IF this template is invoked more than once to describe more than one fetus |
||
3 |
> |
CONTAINS |
CODE |
1-n |
MC |
IF row 4 does not exist |
||
4 |
> |
CONTAINS |
TEXT |
1-n |
MC |
IF row 3 does not exist |
||
5 |
> |
CONTAINS |
CODE |
EV (P0-009C3, SRT, "Surgical Procedure") |
1-n |
U |
BCID 12247 “Pediatric, Fetal and Congenital Cardiac Surgical Interventions” |
Extensible
Significant
Table TID 5228. Cardiac Ultrasound Fetal Measurement Section
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
|||||
2 |
> |
HAS OBS CONTEXT |
INCLUDE |
1 |
MC |
IF this template is invoked more than once to describe more than one fetus. |
||
3 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
$Measurement = DCID 12279 “Cardiac Ultrasound Fetal General Measurements” |
||
4 |
> |
CONTAINS |
INCLUDE |
DTID 5222 “Pediatric, Fetal and Congenital Cardiac Ultrasound Section” |
1-n |
U |
$SectionSubject = EV (T-F6845, SRT, "Ductus arteriosus") $MeasType = DCID 12218 “Echocardiography Congenital” |
|
5 |
> |
CONTAINS |
INCLUDE |
DTID 5222 “Pediatric, Fetal and Congenital Cardiac Ultrasound Section” |
1-n |
U |
$SectionSubject = EV (T-F6806, SRT, "Ductus venosus") $MeasType = DCID 12218 “Echocardiography Congenital” |
|
6 |
> |
CONTAINS |
INCLUDE |
DTID 5222 “Pediatric, Fetal and Congenital Cardiac Ultrasound Section” |
1-n |
U |
$SectionSubject = EV (T-F1810, SRT, "Umbilical artery") $MeasType = DCID 12218 “Echocardiography Congenital” |
|
7 |
> |
CONTAINS |
INCLUDE |
DTID 5222 “Pediatric, Fetal and Congenital Cardiac Ultrasound Section” |
1-n |
U |
$SectionSubject = EV (T-D03B4, SRT, "Umbilical vein") $MeasType = DCID 12218 “Echocardiography Congenital” |
|
8 |
> |
CONTAINS |
INCLUDE |
DTID 5222 “Pediatric, Fetal and Congenital Cardiac Ultrasound Section” |
1-n |
U |
$SectionSubject = EV (T-45600, SRT, "Middle cerebral artery") $MeasType = DCID 12218 “Echocardiography Congenital” |