The templates that comprise the Cardiac Catheterization Report are interconnected as shown in Figure A-7.
Figure A-7 shows the relationship of templates; it does not show the structural hierarchy of Content Items in the IOD.
The Cardiac Cath Report provides the overall clinical results of the catheterization procedure and interventions. In many cases, more detailed information is optionally available in other reports (Hemodynamic Measurements, Procedure Log, etc.). That information is collected and summarized here (and referenced when available).
When a Discharge Summary section (row 12) is included, this report template covers the full set of information required for submission to the ACC NCDR™ (version 2.0) registry.
The information required for such a submission must sometimes be reformatted from a single concept in these templates to two data elements for the registry, or vice versa.
This Template is expected to be used with the Basic Text SR or Enhanced SR IOD.3. This Cardiac Cath Report template is a baseline template within these SOP Classes that may be replaced; it is therefore in no sense binding for exchange of this type of report. It is solely an example of a possible encoding of the Cardiac Cath Report.
Extensible
Significant
Yes
Table TID 3800. Cardiac Catheterization Report Root
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
Root node |
||||
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
M |
|||
3 |
> |
HAS CONCEPT MOD |
INCLUDE |
1 |
M |
|||
4 |
> |
INCLUDE |
1 |
M |
||||
5 |
> |
CONTAINS |
INCLUDE |
1 |
U |
|||
6 |
> |
CONTAINS |
INCLUDE |
1 |
M |
|||
7 |
> |
CONTAINS |
INCLUDE |
1 |
M |
|||
8 |
> |
CONTAINS |
INCLUDE |
1 |
M |
|||
9 |
> |
CONTAINS |
INCLUDE |
1 |
U |
|||
10 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
|||
11 |
> |
CONTAINS |
INCLUDE |
1 |
M |
|||
12 |
> |
CONTAINS |
INCLUDE |
1 |
M |
|||
13 |
> |
CONTAINS |
INCLUDE |
1 |
U |
This template contains information about a cardiovascular patient's past medical history that is relevant for the interpretation of the SR document of which it is part. This information is considered to have some degree of "persistence" across different episodes of care.
Extensible
Significant
Table TID 3802. Cardiovascular Patient History
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
|||||
2 |
> |
CONTAINS |
TEXT |
1 |
U |
|||
3 |
> |
CONTAINS |
CONTAINER |
1 |
U |
|||
4 |
>> |
CONTAINS |
TEXT |
1-n |
U |
|||
5 |
>> |
CONTAINS |
INCLUDE |
1-n |
U |
$Problem = DCID 3756 “Cardiac Patient Risk Factors” |
||
6 |
>> |
CONTAINS |
INCLUDE |
1 |
U |
$Problem = EV (G-023F, SRT, "History of Diabetes mellitus") $Therapy = DCID 3722 “Diabetic Therapy” |
||
7 |
>> |
CONTAINS |
INCLUDE |
1 |
U |
$Problem = EV (G-0269, SRT, "History of Hypertension") $Therapy = DCID 3760 “Hypertension Therapy” |
||
8 |
>> |
CONTAINS |
INCLUDE |
1 |
U |
$Problem = EV (R-102B8, SRT, "History of Hypercholesterolemia") $Therapy = DCID 3761 “Antilipemic Agents” |
||
9 |
>> |
CONTAINS |
INCLUDE |
1 |
U |
$Problem = EV (D3-30000, SRT, "Arrhythmia") $Therapy = DCID 3762 “Antiarrhythmic Agents” |
||
10 |
>> |
CONTAINS |
INCLUDE |
1 |
U |
$Problem = EV (G-03AA, SRT, "History of Myocardial infarction") $ModType = DT (122170, DCM, "Type of Myocardial Infarction") $ModValue = DCID 3723 “MI Types” $Therapy = DCID 3764 “Myocardial Infarction Therapies” |
||
11 |
>> |
CONTAINS |
INCLUDE |
1 |
U |
$Problem = EV (G-0069, SRT, "History of Kidney disease") $Stage = DCID 3778 “Stages” |
||
12 |
> |
CONTAINS |
CONTAINER |
1 |
U |
|||
13 |
>> |
CONTAINS |
TEXT |
EV (F-02455, SRT, "Social History") |
1 |
U |
||
14 |
>> |
CONTAINS |
TEXT |
1-n |
U |
|||
15 |
>> |
CONTAINS |
CODE |
EV (F-93109, SRT, "Tobacco Smoking Behavior") |
1 |
U |
||
16 |
>> |
CONTAINS |
CODE |
DT (F-931D4, SRT, "Drug misuse behavior") |
1 |
U |
DT (D9-30400, SRT, "Cocaine Abuse") |
|
17 |
> |
CONTAINS |
CONTAINER |
1 |
U |
|||
18 |
>> |
CONTAINS |
TEXT |
1-n |
U |
|||
19 |
>> |
CONTAINS |
INCLUDE |
1-n |
U |
$ProcType = DT (P0-009C3, SRT, "Surgical Procedure") $Procedure = DCID 3721 “Cardiovascular Surgeries” |
||
20 |
> |
CONTAINS |
CONTAINER |
DT (30954-2, LN, "Relevant Diagnostic Tests and/or Laboratory Data") |
1 |
U |
||
21 |
>> |
CONTAINS |
TEXT |
1-n |
U |
|||
22 |
>> |
CONTAINS |
INCLUDE |
1-n |
U |
$ProcType = DT (P0-00002, SRT, "Diagnostic procedure") $Procedure = DCID 3757 “Cardiac Diagnostic Procedures” |
||
23 |
>> |
CONTAINS |
NUM |
1 |
U |
UNITS = EV (mg/dl, UCUM, "mg/dl") |
||
24 |
>> |
CONTAINS |
NUM |
1 |
U |
UNITS = EV (mg/dl, UCUM, "mg/dl") |
||
25 |
> |
CONTAINS |
CONTAINER |
1 |
U |
|||
26 |
>> |
CONTAINS |
TEXT |
1-n |
U |
|||
27 |
>>> |
HAS PROPERTIES |
CODE |
1 |
U |
|||
28 |
>> |
CONTAINS |
CODE |
1-n |
U |
|||
29 |
>>> |
HAS PROPERTIES |
NUM |
DT (G-C0B7, SRT, "Dosage") |
1 |
U |
||
30 |
>>> |
HAS PROPERTIES |
CODE |
1 |
U |
|||
31 |
> |
CONTAINS |
CONTAINER |
1 |
U |
|||
32 |
>> |
CONTAINS |
TEXT |
1-n |
U |
|||
33 |
>> |
CONTAINS |
CODE |
EV (G-03E5, SRT, "Family history of clinical finding") |
1-n |
U |
||
34 |
>>> |
HAS PROPERTIES |
CODE |
EV (G-C32E, SRT, "Subject relationship") |
1 |
M |
||
35 |
> |
CONTAINS |
CONTAINER |
1 |
U |
|||
36 |
>> |
CONTAINS |
TEXT |
1 |
U |
|||
37 |
>> |
CONTAINS |
INCLUDE |
1-n |
U |
$Device = DCID 3777 “Implanted Devices” |
This template describes the aspects of the patient that are specific to this clinical presentation (admission).
Extensible
Significant
Table TID 3803. Patient Presentation, Cath
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
|||||
2 |
> |
CONTAINS |
TEXT |
1 |
U |
|||
3 |
> |
CONTAINS |
DATETIME |
EV (15, NCDR [2.0b], "Admission Date/Time") |
1 |
U |
||
4 |
> |
CONTAINS |
CODE |
EV (17, NCDR [2.0b], "Admission Status") |
1 |
U |
||
5 |
> |
CONTAINS |
CODE |
EV (18, NCDR [2.0b], "Insurance Payor Type") |
1 |
U |
||
6 |
> |
CONTAINS |
CODE |
EV (46, NCDR [2.0b], "Congestive Heart Failure Prior to Procedure") |
1 |
U |
||
7 |
> |
CONTAINS |
CODE |
EV (47, NCDR [2.0b], "NYHA Classification") |
1 |
UC |
IFF Row 6 Value code meaning is <yes> |
|
8 |
> |
CONTAINS |
CODE |
EV (48, NCDR [2.0b], "Noninvasive Testing - Ischemia") |
1 |
U |
||
9 |
> |
CONTAINS |
CODE |
EV (49, NCDR [2.0b], "Pre-Cath Angina Type") |
1 |
U |
||
10 |
> |
CONTAINS |
CODE |
EV (50, NCDR [2.0b], "Pre-Cath Canadian Classification") |
1 |
U |
||
11 |
> |
CONTAINS |
CODE |
EV (51, NCDR [2.0b], "Acute Coronary Syndrome Time Period") |
1 |
UC |
IFF Row 9 Value code meaning is <ACS> |
|
12 |
> |
CONTAINS |
CONTAINER |
1 |
U |
|||
13 |
>> |
CONTAINS |
CODE |
1-n |
U |
|||
14 |
> |
CONTAINS |
NUM |
1-n |
U |
UNITS = EV (%, UCUM, "%") |
||
15 |
>> |
HAS CONCEPT MOD |
CODE |
EV (G-C036, SRT, "Measurement method") |
1 |
U |
||
16 |
>> |
INCLUDE |
1 |
U |
||||
17 |
> |
CONTAINS |
CODE |
1-n |
U |
|||
18 |
> |
CONTAINS |
TEXT |
1 |
U |
This template describes the patient-related information about this specific clinical encounter (catheterization).
Extensible
Significant
Table TID 3806. Cath Procedure
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
|||||
2 |
> |
CONTAINS |
DATETIME |
EV (52, NCDR [2.0b], "Procedure DateTime") |
1 |
M |
||
3 |
> |
CONTAINS |
TEXT |
EV (53, NCDR [2.0b], "Procedure Number in this admission") |
1 |
U |
Up to three numeric characters |
|
4 |
> |
CONTAINS |
TEXT |
1 |
U |
|||
5 |
> |
CONTAINS |
COMPOSITE |
1-n |
U |
|||
6 |
> |
CONTAINS |
NUM |
EV (55, NCDR [2.0b], "Fluoroscopy Time") |
1 |
U |
UNITS = DT (min, UCUM, "min") |
|
7 |
> |
CONTAINS |
NUM |
1 |
U |
UNITS = DT (mGy.cm2, UCUM, "mGy.cm2") |
||
8 |
> |
CONTAINS |
PNAME |
EV (76, NCDR [2.0b], "Catheterization Operator") |
1 |
M |
||
9 |
> |
CONTAINS |
PNAME |
1-n |
U |
|||
10 |
> |
CONTAINS |
PNAME |
1-n |
U |
|||
11 |
> |
CONTAINS |
CODE |
1 |
U |
|||
12 |
> |
CONTAINS |
CONTAINER |
EV (F-04460, SRT, "Medication Given") |
1 |
M |
||
13 |
>> |
CONTAINS |
CODE |
EV (57, NCDR [2.0b], "Thrombolytics") |
1 |
U |
||
14 |
>> |
CONTAINS |
CODE |
EV (58, NCDR [2.0b], "IIb/IIIa Blockade") |
1 |
U |
||
15 |
>> |
CONTAINS |
CODE |
EV (59, NCDR [2.0b], "Heparin") |
1-n |
U |
||
16 |
>> |
CONTAINS |
CODE |
EV (60, NCDR [2.0b], "Aspirin") |
1 |
U |
||
17 |
>> |
CONTAINS |
CODE |
EV (61, NCDR [2.0b], "Clopidogrel/Ticlopidine") |
1 |
U |
||
18 |
>> |
CONTAINS |
TEXT |
1-n |
U |
|||
19 |
> |
CONTAINS |
CODE |
1-n |
U |
|||
20 |
> |
CONTAINS |
CODE |
EV (74, NCDR [2.0b], "Percutaneous Entry Site") |
1 |
M |
||
21 |
> |
CONTAINS |
CODE |
EV (75, NCDR [2.0b], "Closure Device") |
1 |
U |
Content Item Descriptions
Row 3 |
Procedure Number (this admission) is specified as a numeric text string, and shall be treated as the ordinal of this catheterization procedure within the admission (i.e., "1" for the first catheterization, "2" for the second, etc.). |
This template describes the various aspects of a coronary intervention.
Extensible
Significant
Table TID 3807. Percutaneous Coronary Intervention Procedure
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
|||||
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
M |
|||
3 |
> |
CONTAINS |
PNAME |
1 |
M |
|||
4 |
> |
CONTAINS |
PNAME |
1-n |
U |
|||
5 |
> |
CONTAINS |
PNAME |
1-n |
U |
|||
6 |
> |
CONTAINS |
CODE |
EV (G-C09C, SRT, "Procedure Priority") |
1 |
M |
||
7 |
> |
CONTAINS |
CONTAINER |
1 |
U |
|||
8 |
>> |
CONTAINS |
CODE |
1-n |
U |
|||
9 |
>> |
CONTAINS |
CODE |
1 |
U |
|||
10 |
>>> |
HAS PROPERTIES |
CODE |
1 |
U |
|||
11 |
>>> |
HAS PROPERTIES |
DATETIME |
1 |
U |
|||
12 |
>> |
CONTAINS |
TEXT |
1-n |
U |
|||
13 |
> |
CONTAINS |
NUM |
EV (122175, DCM, "Number of lesion interventions attempted") |
1 |
M |
UNITS = EV (1, UCUM, "units") |
|
14 |
> |
CONTAINS |
NUM |
EV (122176, DCM, "Number of lesion interventions successful") |
1 |
M |
UNITS = EV (1, UCUM, "units") |
|
15 |
> |
CONTAINS |
CODE |
1 |
M |
|||
16 |
> |
CONTAINS |
TEXT |
1 |
U |
|||
17 |
> |
CONTAINS |
INCLUDE |
1-n |
M |
Content Item Descriptions
Rows 8 and 12 |
Allow the recording of findings as either codes or as text; the same finding shall not be recorded as both. |
Rows 15 and 16 |
Allow the recording of procedure results as either codes or as text, but not as both. |
Extensible
Significant
Table TID 3808. Lesion Intervention Information
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
|||||
2 |
> |
CONTAINS |
INCLUDE |
1 |
M |
|||
3 |
> |
CONTAINS |
CODE |
EV (108, NDR [2.0b], "Previous Dilation") |
1 |
U |
||
4 |
> |
CONTAINS |
CODE |
EV (103, NDR [2.0b], "Guidewire crossing lesion") |
1 |
U |
||
5 |
> |
CONTAINS |
CODE |
EV (G-C50A, SRT, "Uses Equipment") |
1-n |
M |
||
6 |
>> |
HAS CONCEPT MOD |
CODE |
1 |
MC |
IF Device is Primary for this Lesion |
||
7 |
>> |
HAS PROPERTIES |
TEXT |
1 |
U |
|||
8 |
>> |
HAS PROPERTIES |
NUM |
1-n |
U |
|||
9 |
>> |
HAS PROPERTIES |
NUM |
1-n |
U |
|||
10 |
>> |
HAS PROPERTIES |
DATETIME |
1 |
U |
|||
11 |
>> |
HAS PROPERTIES |
NUM |
1 |
U |
UNITS = EV (s, UCUM, "s") |
||
12 |
> |
CONTAINS |
NUM |
EV (R-101BB, SRT, "Lumen Diameter Stenosis") |
1 |
M |
UNITS = EV (%, UCUM, "%") |
|
13 |
>> |
HAS CONCEPT MOD |
CODE |
EV (G-72BB, SRT, "Catheterization Procedure Phase") |
1 |
M |
EV (G-7298, SRT, "Post-intervention Phase") |
|
14 |
>> |
HAS CONCEPT MOD |
CODE |
1 |
U |
|||
15 |
>> |
INCLUDE |
1 |
U |
||||
16 |
> |
CONTAINS |
CODE |
1 |
U |
|||
17 |
> |
CONTAINS |
CODE |
EV (115, NCDR [2.0b], "Dissection in segment observed") |
1 |
U |
||
18 |
> |
CONTAINS |
CODE |
EV (116, NCDR [2.0b], "Acute closure observed") |
1 |
U |
||
19 |
> |
CONTAINS |
CODE |
EV (117, NCDR [2.0b], "Acute closure re-opened") |
1 |
UC |
IFF Row 18 value is <yes> |
|
20 |
> |
CONTAINS |
CODE |
EV (118, NCDR [2.0b], "Perforation occurred") |
1 |
U |
||
21 |
> |
CONTAINS |
IMAGE |
DT (121080, DCM, "Best illustration of finding") |
1 |
U |
||
22 |
> |
CONTAINS |
TEXT |
1 |
U |
Extensible
Significant
Table TID 3809. Other Interventional Procedures
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
|||||
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
M |
|||
3 |
> |
CONTAINS |
TEXT |
1 |
U |
|||
4 |
> |
CONTAINS |
CODE |
1 |
U |
No BCID |
||
5 |
> |
CONTAINS |
TEXT |
1 |
U |
Content Item Descriptions
Rows 3 and 4 |
Allow the recording of procedure description as either code or as text; the same description shall not be recorded as both. |
The Cardiac Catheterization Findings Template provides the structure for the diagnostic findings of the cath procedure, organized into sub-sections based on type of sub-procedure. It also provides for top-level summary findings and diagnoses.
Extensible
Significant
Table TID 3810. Cardiac Catheterization Findings
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINS |
CONTAINER |
1 |
M |
||||
2 |
> |
CONTAINS |
INCLUDE |
1 |
U |
|||
3 |
> |
CONTAINS |
INCLUDE |
1 |
U |
|||
4 |
> |
CONTAINS |
INCLUDE |
1 |
U |
|||
5 |
> |
CONTAINS |
INCLUDE |
1 |
U |
|||
6 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
|||
7 |
> |
CONTAINS |
CODE |
1-n |
U |
|||
8 |
>> |
HAS PROPERTIES |
CODE |
EV (G-C197, SRT, "Severity") |
1 |
U |
||
9 |
> |
CONTAINS |
TEXT |
1-n |
U |
Content Item Descriptions
Rows 7 and 9 |
Allow the recording of findings as either codes or as text; the same finding shall not be recorded as both. |
Extensible
Significant
Table TID 3812. Hemodynamic Findings
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINS |
CONTAINER |
1 |
M |
||||
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
M |
DT (PA-50030, SRT, "Hemodynamic measurements") |
||
3 |
> |
CONTAINS |
CODE |
1-n |
U |
|||
4 |
>> |
INCLUDE |
1 |
U |
||||
5 |
>> |
HAS PROPERTIES |
CODE |
EV (G-C197, SRT, "Severity") |
1 |
U |
||
6 |
> |
CONTAINS |
CODE |
1-n |
U |
|||
7 |
>> |
HAS PROPERTIES |
CODE |
EV (G-C0B2, SRT, "Condition") |
1 |
U |
EV (R-102B9, SRT, "Large v-wave") |
|
8 |
> |
CONTAINS |
CODE |
1 |
U |
EV (R-102BA, SRT, "Diastolic pressure equalization") |
||
9 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
$Report = DT (122120, DCM, "Hemodynamics Report") |
Content Item Descriptions
Row 4 |
(Through DTID 3819 Common Findings Template) may be used to encode any significant hemodynamic numeric measurements. For reference, see Templates 3550 Pressure Waveform Measurements, and 3560 Derived Hemodynamic Measurements. |
The information contained here about the left ventricle is relatively qualitative in nature. It is a high-level summary of the more detailed information that may be contained in an optional Quantitative Ventricular Analysis report. This template addresses findings about any ventricular septal defect (Row 7), the myocardial wall (Row 11), and about the aortic root (Row 16).
Extensible
Significant
Table TID 3814. Left Ventriculography Findings
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINS |
CONTAINER |
1 |
M |
||||
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
M |
DT (P5-30041, SRT, "Left Ventriculography") |
||
3 |
> |
CONTAINS |
CODE |
EV (F-30117, SRT, "Left Ventricular Function - Finding") |
1 |
M |
||
4 |
> |
INCLUDE |
1 |
M |
$Measurement = EV (10230-1, LN, "LV Ejection Fraction") $Units = EV (%, UCUM, "%") $Method = DCID 3748 “Angiographic EF Testing Method” $Derivation = DCID 3745 “Calculation Method” |
|||
5 |
> |
CONTAINS |
CODE |
EV (F-0224E, SRT, "Left Ventricular Cavity Size") |
1 |
U |
||
6 |
> |
CONTAINS |
CODE |
EV (F-02225, SRT, "Left Ventricular Contractility") |
1 |
U |
||
7 |
> |
CONTAINS |
CODE |
1 |
U |
EV (D4-31159, SRT, "Ventricular Septal Defect") |
||
8 |
>> |
HAS PROPERTIES |
CODE |
EV (G-C504, SRT, "Associated Morphology") |
1 |
U |
||
9 |
> |
CONTAINS |
INCLUDE |
1 |
U |
|||
10 |
> |
CONTAINS |
CONTAINER |
1 |
U |
|||
11 |
>> |
HAS CONCEPT MOD |
CODE |
EV (G-C0E3, SRT, "Finding Site") |
1 |
M |
DT (T-D0772, SRT, "Myocardial Wall") |
|
12 |
>> |
CONTAINS |
CODE |
1-n |
M |
|||
13 |
>>> |
HAS PROPERTIES |
CODE |
EV (F-32050, SRT, "Cardiac Wall Motion") |
1 |
M |
||
14 |
>>> |
HAS PROPERTIES |
CODE |
EV (G-C504, SRT, "Associated Morphology") |
1 |
U |
||
15 |
>>> |
HAS PROPERTIES |
NUM |
DT (G-C1E3, SRT, "Score") |
1 |
U |
||
16 |
> |
CONTAINS |
CONTAINER |
1 |
U |
|||
17 |
>> |
HAS CONCEPT MOD |
CODE |
EV (G-C0E3, SRT, "Finding Site") |
1 |
M |
DT (F-04403, SRT, "Aortic Root") |
|
18 |
>> |
CONTAINS |
CODE |
1-n |
M |
The information contained here about right ventricle is relatively qualitative in nature. It is a high-level summary of the more detailed information that may be contained in an optional Quantitative Ventricular Analysis report.
Extensible
Significant
Table TID 3815. Right Ventriculography Findings
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINS |
CONTAINER |
1 |
M |
||||
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
M |
DT (P5-3003F, SRT, "Right Ventriculography") |
||
3 |
> |
INCLUDE |
1 |
M |
$Measurement = EV (10231-9, LN, "RV Ejection Fraction")) $Units = EV (%, UCUM, "%") $Method = DCID 3748 “Angiographic EF Testing Method” $Derivation = DCID 3745 “Calculation Method” |
|||
4 |
> |
CONTAINS |
CODE |
EV (F-022A1, SRT, "Right Ventricular Cavity Size") |
1 |
U |
||
5 |
> |
CONTAINS |
CODE |
EV (F-0227A, SRT, "Right Ventricular Contractility") |
1 |
U |
||
6 |
> |
CONTAINS |
INCLUDE |
1 |
U |
Extensible
Significant
Table TID 3816. Ventricular Assessment
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINS |
CONTAINER |
1-n |
U |
||||
2 |
> |
HAS CONCEPT MOD |
CODE |
EV (G-C0E3, SRT, "Finding Site") |
1 |
M |
||
3 |
> |
CONTAINS |
CODE |
1-n |
M |
|||
4 |
>> |
HAS PROPERTIES |
CODE |
EV (G-C197, SRT, "Severity") |
1 |
U |
||
5 |
>> |
HAS PROPERTIES |
NUM |
DT (G-C1E3, SRT, "Score") |
1 |
U |
||
6 |
CONTAINS |
INCLUDE |
1-n |
U |
$Report = DT (122292, DCM, "Quantitative Ventriculography Report") |
The information contained here about with regards to coronary artery lesions is relatively qualitative in nature. It is a high-level summary of the more detailed information that may be contained in an optional Quantitative Coronary Arteriography report. This template addresses findings about the individual arteries (Row 4), and about individual lesions (Row 9).
Extensible
Significant
Table TID 3817. Coronary Arteriography Findings
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINS |
CONTAINER |
1 |
M |
||||
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
M |
DT (P5-30100, SRT, "Coronary Arteriography") |
||
3 |
> |
CONTAINS |
CODE |
EV (F-04404, SRT, "Coronary artery feature ") |
1 |
U |
||
4 |
> |
CONTAINS |
CONTAINER |
1-n |
U |
|||
5 |
>> |
HAS CONCEPT MOD |
CODE |
EV (G-C0E3, SRT, "Finding Site") |
1 |
M |
||
6 |
>>> |
HAS CONCEPT MOD |
CODE |
EV (G-A1F8, SRT, "Topographical modifier") |
1 |
U |
||
7 |
>> |
CONTAINS |
CODE |
1-n |
U |
|||
8 |
>> |
CONTAINS |
INCLUDE |
1-n |
U |
|||
9 |
> |
CONTAINS |
CONTAINER |
1-n |
U |
|||
10 |
>> |
HAS CONCEPT MOD |
CODE |
EV (G-C0E3, SRT, "Finding Site") |
1 |
M |
DT (R-002EF, SRT, "Coronary artery lesion (culprit)") |
|
11 |
>> |
CONTAINS |
INCLUDE |
1 |
M |
|||
12 |
>> |
CONTAINS |
INCLUDE |
1-n |
U |
$Report = DT (122291, DCM, "Quantitative Arteriography Report") |
Extensible
Significant
Table TID 3818. Other Cardiographic Findings
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINS |
CONTAINER |
1 |
M |
||||
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
M |
|||
3 |
> |
HAS CONCEPT MOD |
CODE |
EV (G-C0E3, SRT, "Finding Site") |
1 |
U |
||
4 |
> |
CONTAINS |
INCLUDE |
1-n |
M |
Table TID 3819. Parameters
Parameter Name |
Parameter Usage |
---|---|
$Report |
Title of composite object (evidence document) referenced |
Extensible
Significant
Table TID 3819. Common Findings
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
TEXT |
1 |
U |
|||||
2 |
IMAGE |
1 |
U |
|||||
3 |
NUM |
No BCID |
1 |
U |
||||
4 |
> |
INCLUDE |
1 |
U |
||||
5 |
COMPOSITE |
1 |
U |
|||||
6 |
> |
HAS PROPERTIES |
CODE |
1 |
U |
$Report |
Content Item Descriptions
Row 3 |
May be used to encode any significant image- or waveform-based numeric measurements, with post-coordination of the Concept Name using child Content Items (with relationship HAS CONCEPT MOD), as permitted by Section 6.2.4. The source of the measurement may be noted using the Quotation Template in Row 4. |
Extensible
Significant
Table TID 3820. Adverse Outcomes, Cath
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
|||||
2 |
> |
CONTAINS |
CODE |
EV (DD-60002, SRT, "Complication of Procedure") |
1-n |
U |
||
3 |
>> |
HAS CONCEPT MOD |
CODE |
EV (G-D709, SRT, "Relative Time") |
1 |
U |
DCID 12102 “Temporal Periods Relating to Procedure or Therapy” |
|
4 |
> |
CONTAINS |
CODE |
EV (DD-60002, SRT, "Complication of Procedure") |
1-n |
U |
||
5 |
>> |
HAS CONCEPT MOD |
CODE |
EV (G-D709, SRT, "Relative Time") |
1 |
U |
DCID 12102 “Temporal Periods Relating to Procedure or Therapy” |
|
6 |
> |
CONTAINS |
TEXT |
EV (DD-60002, SRT, "Complication of Procedure") |
1-n |
U |
||
7 |
> |
CONTAINS |
CODE |
1 |
U |
|||
8 |
>> |
INFERRED FROM |
NUM |
1 |
U |
UNITS = EV ([iU], UCUM, "International unit") |
||
9 |
>>> |
HAS PROPERTIES |
NUM |
EV (R-0038B, SRT, "Normal Range Upper Limit") |
1 |
M |
UNITS = EV ([iU], UCUM, "International unit") |
|
10 |
>> |
INFERRED FROM |
NUM |
1 |
M |
UNITS = EV ([iU], UCUM, "International unit") |
||
11 |
> |
CONTAINS |
IMAGE |
DT (121080, DCM, "Best illustration of finding") |
1-n |
U |
Content Item Descriptions
Rows 2, 3 and 4 |
Allow the recording of outcomes as either codes or as text; the same outcome shall not be recorded as both. |
Extensible
Non-Significant
Table TID 3824. Summary, Cath
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINS |
CONTAINER |
1 |
M |
||||
2 |
> |
CONTAINS |
CODE |
1-n |
U |
|||
3 |
>> |
HAS PROPERTIES |
CODE |
EV (G-C197, SRT, "Severity") |
1 |
U |
||
4 |
> |
CONTAINS |
TEXT |
1 |
U |
|||
5 |
> |
CONTAINS |
TEXT |
1-n |
U |
Extensible
Significant
Table TID 3828. Discharge Summary, Cath
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINS |
CONTAINER |
1 |
M |
||||
2 |
> |
CONTAINS |
DATETIME |
1 |
U |
|||
3 |
> |
CONTAINS |
CODE |
EV (122164, DCM, "Coronary Artery Bypass During This Admission") |
1 |
U |
||
4 |
>> |
HAS PROPERTIES |
CODE |
EV (G-C09C, SRT, "Procedure Priority") |
1 |
U |
||
5 |
>> |
HAS PROPERTIES |
DATETIME |
1 |
U |
|||
6 |
> |
CONTAINS |
CODE |
1 |
U |
|||
7 |
>> |
HAS PROPERTIES |
DATETIME |
1 |
U |
|||
8 |
>> |
HAS PROPERTIES |
CODE |
EV (25, NCDR [2.0b], "Primary Cause of Death") |
1 |
U |
||
9 |
>> |
HAS PROPERTIES |
CODE |
1 |
U |
|||
10 |
> |
CONTAINS |
TEXT |
1 |
U |
Table TID 3829. Parameters
Parameter Name |
Parameter Usage |
---|---|
$Problem |
Coded Value or Context Group for problem |
$ModType |
Modifier Name for Concept Name of problem |
$ModValue |
Modifier Value for Concept Name of problem |
$Therapy |
Coded Value or Context Group for therapy received for problem |
$Stage |
Coded Value or Context Group for problem or disease stage |
Extensible
Significant
Table TID 3829. Problem Properties
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
|||||
2 |
> |
CONTAINS |
DATETIME |
1 |
U |
|||
3 |
> |
CONTAINS |
DATETIME |
1 |
U |
|||
4 |
> |
CONTAINS |
CODE |
1 |
M |
$Problem |
||
5 |
>> |
HAS CONCEPT MOD |
CODE |
$ModType |
1-n |
U |
$ModValue |
|
6 |
>> |
HAS PROPERTIES |
DATETIME |
1 |
U |
|||
7 |
>> |
HAS PROPERTIES |
DATETIME |
1 |
U |
|||
8 |
>> |
HAS PROPERTIES |
CODE |
1 |
U |
|||
9 |
>> |
HAS PROPERTIES |
CODE |
DT (G-C197, SRT, "Severity") |
1 |
U |
||
10 |
>> |
HAS PROPERTIES |
CODE |
DT (G-C16B, SRT, "Stage") |
1 |
U |
$Stage |
|
11 |
> |
CONTAINS |
CODE |
1 |
U |
|||
12 |
> |
CONTAINS |
CODE |
EV (P0-0000E, SRT, "Therapy") |
1-n |
U |
$Therapy |
|
13 |
>> |
HAS PROPERTIES |
CODE |
1 |
U |
|||
14 |
> |
CONTAINS |
TEXT |
1 |
U |
Table TID 3830. Parameters
Parameter Name |
Parameter Usage |
---|---|
$ProcType |
Coded Value for class of procedure |
$Procedure |
Coded Value or Context Group for procedure |
$ModType |
Modifier Name for Concept Name of procedure |
$ModValue |
Modifier Value for Concept Name of procedure |
Extensible
Significant
Table TID 3830. Procedure Properties
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CODE |
$ProcType |
1 |
M |
$Procedure |
|||
2 |
> |
HAS CONCEPT MOD |
CODE |
$ModType |
1-n |
U |
$ModValue |
|
3 |
> |
HAS PROPERTIES |
DATETIME |
1 |
U |
|||
4 |
> |
HAS PROPERTIES |
COMPOSITE |
EV (R-42B89, SRT, "Clinical Report") |
1-n |
U |
||
5 |
>> |
HAS PROPERTIES |
CODE |
1 |
U |
|||
6 |
> |
HAS PROPERTIES |
TEXT |
EV (R-42B89, SRT, "Clinical Report") |
1-n |
U |
Description of report with URL or other reference for report |
|
7 |
> |
HAS PROPERTIES |
TEXT |
1 |
U |
|||
8 |
> |
HAS PROPERTIES |
PNAME |
1 |
UC |
IF Service Performer is a person |
||
9 |
> |
HAS PROPERTIES |
TEXT |
1 |
UC |
IF Service Performer is an organization |
||
10 |
> |
HAS PROPERTIES |
TEXT |
1 |
U |