| DICOM PS3.18 2026c - Web Services |
|---|
The "renderedvolumetricmetadata" parameter specifies that the response payload contains only a Rendered Volume Response Module of the parameters applied by the origin server to generate the volumetric rendering.
The syntax of this parameter is:
%s"renderedvolumetricmetadata =" "yes"
yes |
Indicates that only the Rendered Volume Response Module shall be present in the response payload. |
The origin server shall return a response payload containing a Rendered Volume Response Module as specified in Annex K.
If this parameter is not present, no Response Module is requested and the 2D representation of the rendered volume is returned as described in Section 10.4.1.1.7 and Section 10.4.1.1.8.
The Rendered Volume Response Module contains the complete set of query parameters, including both those specified by the user agent in the request and those determined by the origin server.
For any two requests where the query parameters, Target Resource, and Header Fields are identical, the query parameter values in the Rendered Volume Response Module returned by the origin server shall be identical.
When repeating a rendering request, the origin server is expected to deliver a result that is consistent. It is not mandatory for it to be precisely identical.
If the Target Resource is a Volumetric Presentation State, the origin server shall return a 400 (Bad Request) and may include an appropriate Status Report.
| DICOM PS3.18 2026c - Web Services |
|---|