The "animationrate" parameter specifies the rate at which an animated 3D or MPR Volumetric Presentation is displayed.
The syntax of this parameter is:
%s"animationrate=" rt
This corresponds to Recommended Animation Rate (0070,1A03) in Section C.11.29 in PS3.3 and Section FF.2.4.2 in PS3.4.
Playback of the returned media on a client may or may not achieve the requested animation rate.
If "animationrate" is not present, and other animation parameters are present (e.g., "swivelrange", "animationstepsize", or "volumetriccurvepoint"), the origin server may apply a default animation rate, or alternatively, return 400 (Bad Request) and may include an appropriate Status Report.
If the Target Resource is a Volumetric Presentation State, the origin server shall return a 400 (Bad Request) and may include an appropriate Status Report.