15.2 Conformance

An origin server conforming to the Modality Performed Procedure Step Service shall support the Retrieve Capabilities Transaction (see Section 8.9.1). Furthermore, it shall support the Transactions listed as Required in Table 15.2-1 and may support Transactions listed as Optional.

Table 15.2-1. Required and Optional Transactions

Transaction

Support

Section

Retrieve Capabilities

Required

Section 8.9

Create

Required

Section 15.4

Update

Required

Section 15.5

Retrieve

Optional

Section 15.6


Implementations shall specify in their Conformance Statement (see PS3.2) and the Retrieve Capabilities Transaction the supported Transactions and the implementations' role: origin server, user agent, or both.

In addition, for each supported Transaction they shall specify: