DICOM PS3.18 2026d - Web Services

10.6.3.3 Response Payload

A success response shall contain a list of matching results in an Acceptable Media Type. See Section 8.7.5.

If (results = 0) then there are no matches, and a 200 (OK) response shall be returned with an empty payload. The empty payload shall be formatted as specified in Section 8.3.4.4.1.

A failure response payload may contain a Status Report describing any failures, warnings, or other useful information.

10.6.3.3.1 Study Resource

For each matching Study, the origin server response shall contain Attributes in accordance with Table 10.6.3-3. The "Type" column in the table below refers to the Query/Retrieve Attribute Types defined in Section C.2.2.1 “Attribute Types” in PS3.4. The unique key for a Study resource Search response is the Study Instance UID (0020,000D).

Table 10.6.3-3. Study Resource Search Response Payload

Attribute Name

Tag

Type

Condition

Study Date

(0008,0020)

R

Study Time

(0008,0030)

R

Accession Number

(0008,0050)

R

Instance Availability

(0008,0056)

C

Shall be present if known

Modalities in Study

(0008,0061)

R

Referring Physician's Name

(0008,0090)

R

Timezone Offset From UTC

(0008,0201)

C

Shall be present if known

Retrieve URL

(0008,1190)

C

Shall be present if the Instance is retrievable by the Retrieve Transaction

Patient's Name

(0010,0010)

R

Patient ID

(0010,0020)

R

Patient's Birth Date

(0010,0030)

R

Patient's Sex

(0010,0040)

R

Study Instance UID

(0020,000D)

U

Study ID

(0020,0010)

R

Number of Study Related Series

(0020,1206)

R

Number of Study Related Instances

(0020,1208)

R


Note

  1. While some of the above Attributes are optional in Table C.6-1 “Patient Level Attributes for the Patient Root Query/Retrieve Information Model” in PS3.4, Table C.6-2 “Study Level Keys for the Patient Root Query/Retrieve Information Model” in PS3.4, and Table C.6-5 “Study Level Keys for the Study Root Query/Retrieve Information Model” in PS3.4, they are consistent with those required in [IHE RAD TF-2] Table 4.14-1.

  2. If either Prior Record Key (0008,041C) or Record Key (0008,041B) is present in the request and the Origin Server supports these attributes, then the responses are ordered as defined in Section C.6.4 “Repository Query SOP Class” in PS3.4.

In addition, the response shall contain:

Series or Instance Level Attributes contained in includefield parameters shall not be returned.