This Template fully identifies a detection algorithm and the images and/or image regions on which it operated (see TID 4015 “CAD Detections Performed”).
Table TID 4017. CAD Detection Performed
|
Shall reference IMAGE Content Item(s) in the (111028, DCM, "Image Library") |
||||||||
|
Shall reference an IMAGE Content Item in the (111028, DCM, "Image Library") |
||||||||
|
If more than one detection algorithm has the same "Detection Performed" Code Value (e.g., CID 6014 “Mammography Single Image Finding”) then the "CAD Algorithm Identification" shall unambiguously distinguish between algorithms. |
|
|
Mammography CAD SR: When this Template is invoked for the Mammography CAD SR, the Image Library is mandatory, thus only Row 8 shall be present. Chest CAD SR: When this Template is invoked for the Chest CAD SR, the Image Library is optional, thus Row 7 or 8 may be present. Colon CAD SR: When this Template is invoked for the Colon CAD SR, the Image Library does not exist, thus only Row 7 may be present. |