DICOM PS3.16 2026c - Content Mapping Resource

TID 4009 Mammography CAD Individual Calcification

This Template provides the detail specific to an individual calcification.

Type:

Non-Extensible

Order:

Significant

Root:

No

Table TID 4009. Mammography CAD Individual Calcification

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

EV (111009, DCM, "Calcification Type")

1-n

MC

At least one of Rows 1, 2, 3 shall be present

DCID 6010 “Mammography Calcification Type”

2

INCLUDE

DTID 1400 “Linear Measurement”

1-n

MC

At least one of Rows 1, 2, 3 shall be present

If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID 1400 “Linear Measurement” shall be used.

3

INCLUDE

DTID 1401 “Area Measurement”

1-n

MC

At least one of Rows 1, 2, 3 shall be present

If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID 1401 “Area Measurement” shall be used.

4

INCLUDE

DTID 1402 “Volume Measurement”

1-n

U

If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID 1402 “Volume Measurement” shall be used.


DICOM PS3.16 2026c - Content Mapping Resource