TID 3902 Vascular Analysis

Contains either morphological or functional vascular measurement results of an analysis

Table TID 3902. Parameters

Parameter Name

Parameter Usage

$AnalysisPerformed

Analysis Performed


Type:

Extensible

Order:

Significant

Root:

No

Table TID 3902. Vascular Analysis

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (59776-5, LN, "Findings")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (111004, DCM, "Analysis Performed")

1

M

$AnalysisPerformed

3

>

CONTAINS

INCLUDE

DTID 3905 “Calcium Scoring Results”

1

UC

IFF the value of Row 2 is EV (122605, DCM, "Vascular Morphological Analysis")

4

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (281231009, SCT, "Blood Vessel of Head")

$SectionLaterality = EV (7771000, SCT, "Left")

$Anatomy = DCID 12105 “Intracranial Cerebral Vessel”

$AnalysisPerformed = $AnalysisPerformed

5

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (281231009, SCT, "Blood Vessel of Head")

$SectionLaterality = EV (24028007, SCT, "Right")

$Anatomy = DCID 12105 “Intracranial Cerebral Vessel”

$AnalysisPerformed = $AnalysisPerformed

6

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (281231009, SCT, "Blood Vessel of Head")

$SectionLaterality = EV (66459002, SCT, "Unilateral")

$Anatomy = DCID 12106 “Intracranial Cerebral Vessel (Unilateral)”

$AnalysisPerformed = $AnalysisPerformed

7

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (119568004, SCT, "Artery of Neck")

$SectionLaterality = EV (7771000, SCT, "Left")

$Anatomy = DCID 12104 “Extracranial Artery”

$AnalysisPerformed = $AnalysisPerformed

8

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (119568004, SCT, "Artery of Neck")

$SectionLaterality = EV (24028007, SCT, "Right")

$Anatomy = DCID 12104 “Extracranial Artery”

$AnalysisPerformed = $AnalysisPerformed

9

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (70791007, SCT, "Artery of Lower Extremity")

$SectionLaterality = EV (7771000, SCT, "Left")

$Anatomy = DCID 12109 “Lower Extremity Artery”

$AnalysisPerformed = $AnalysisPerformed

10

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (70791007, SCT, "Artery of Lower Extremity")

$SectionLaterality = EV (24028007, SCT, "Right")

$Anatomy = DCID 12109 “Lower Extremity Artery”

$AnalysisPerformed = $AnalysisPerformed

11

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (122774002, SCT, "Vein of Lower Extremity")

$SectionLaterality = EV (7771000, SCT, "Left")

$Anatomy = DCID 12110 “Lower Extremity Vein”

$AnalysisPerformed = $AnalysisPerformed

12

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (122774002, SCT, "Vein of Lower Extremity")

$SectionLaterality = EV (24028007, SCT, "Right")

$Anatomy = DCID 12110 “Lower Extremity Vein”

$AnalysisPerformed = $AnalysisPerformed

13

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (75531005, SCT, "Artery of Upper Extremity")

$SectionLaterality = EV (7771000, SCT, "Left")

$Anatomy = DCID 12107 “Upper Extremity Artery”

$AnalysisPerformed = $AnalysisPerformed

14

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (75531005, SCT, "Artery of Upper Extremity")

$SectionLaterality = EV (24028007, SCT, "Right")

$Anatomy = DCID 12107 “Upper Extremity Artery”

$AnalysisPerformed = $AnalysisPerformed

15

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (122775001, SCT, "Vein of Upper Extremity")

$SectionLaterality = EV (7771000, SCT, "Left")

$Anatomy = DCID 12108 “Upper Extremity Vein”

$AnalysisPerformed = $AnalysisPerformed

16

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (122775001, SCT, "Vein of Upper Extremity")

$SectionLaterality = EV (24028007, SCT, "Right")

$Anatomy = DCID 12108 “Upper Extremity Vein”

$AnalysisPerformed = $AnalysisPerformed

17

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (303402001, SCT, "Vascular Structure of Kidney")

$SectionLaterality = EV (7771000, SCT, "Left")

$Anatomy = DCID 12115 “Renal Vessel”

$AnalysisPerformed = $AnalysisPerformed

18

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (303402001, SCT, "Vascular Structure of Kidney")

$SectionLaterality = EV (24028007, SCT, "Right")

$Anatomy = DCID 12115 “Renal Vessel”

$AnalysisPerformed = $AnalysisPerformed

19

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (118634008, SCT, "Artery of Abdomen")

$SectionLaterality = EV (7771000, SCT, "Left")

$Anatomy = DCID 12111 “Abdominopelvic Artery (Paired)”

$AnalysisPerformed = $AnalysisPerformed

20

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (118634008, SCT, "Artery of Abdomen")

$SectionLaterality = EV (24028007, SCT, "Right")

$Anatomy = DCID 12111 “Abdominopelvic Artery (Paired)”

$AnalysisPerformed = $AnalysisPerformed

21

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (118634008, SCT, "Artery of Abdomen")

$SectionLaterality = EV (66459002, SCT, "Unilateral")

$Anatomy = DCID 12112 “Abdominopelvic Artery (Unpaired)”

$AnalysisPerformed = $AnalysisPerformed

22

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (84421000, SCT, "Vein of Abdomen")

$SectionLaterality = EV (7771000, SCT, "Left")

$Anatomy = DCID 12113 “Abdominopelvic Vein (Paired)”

$AnalysisPerformed = $AnalysisPerformed

23

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (84421000, SCT, "Vein of Abdomen")

$SectionLaterality = EV (24028007, SCT, "Right")

$Anatomy = DCID 12113 “Abdominopelvic Vein (Paired)”

$AnalysisPerformed = $AnalysisPerformed

24

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (84421000, SCT, "Vein of Abdomen")

$SectionLaterality = EV (66459002, SCT, "Unilateral")

$Anatomy = DCID 12114 “Abdominopelvic Vein (Unpaired)”

$AnalysisPerformed = $AnalysisPerformed

25

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (81040000, SCT, "Pulmonary Artery Structure")

$SectionLaterality = EV (66459002, SCT, "Unilateral")

$Anatomy = DCID 3829 “Pulmonary Artery”

$AnalysisPerformed = $AnalysisPerformed

26

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (41801008, SCT, "Coronary Artery Structure")

$Anatomy = DCID 3015 “Coronary Artery”

$AnalysisPerformed = $AnalysisPerformed

27

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (85439003, SCT, "Cardiac Vein Structure")

$Anatomy = DCID 3839 “Coronary Vein”

$AnalysisPerformed = $AnalysisPerformed

28

>

CONTAINS

INCLUDE

DTID 3906 “Vascular Section Measurements”

1-n

U

$VascularSection = DT (122972007, SCT, "Pulmonary Venous Structure")

$Anatomy = DCID 3840 “Pulmonary Vein”

$AnalysisPerformed = $AnalysisPerformed