DICOM PS3.16 2026b - Content Mapping Resource

TID 3317 Stress Imaging Summary

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3317. Stress Imaging Summary

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

DT (122739, DCM, "Overall study quality")

1

M

BCID 3114 “Study Quality”

2

CODE

DT (113010, DCM, "Quality Issue")

1

U

BCID 3115 “Stress Imaging Quality Issue”

3

CODE

EV (121071, DCM, "Finding")

1

U

BCID 3116 “NM Extracardiac Finding”

4

INCLUDE

DTID 300 “Measurement”

1-2

U

$Measurement = DT (427751006, SCT, "Perfusion defect extent")

$ModType = EV (109054, DCM, "Patient State")

$ModValue = DCID 3102 “Rest/Stress State”

$Units = EV (%, UCUM, "%")

5

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = DT (429162008, SCT, "Stress ischemia extent")

$Units = EV (%, UCUM, "%")

6

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = DT (122762, DCM, "Number of diseased vessel territories")

$Units = EV ({territories}, UCUM, "territories")

7

CODE

EV (121071, DCM, "Finding")

1

U

EV (53741008, SCT, "Coronary artery disease")

8

>

HAS PROPERTIES

CODE

EV (363698007, SCT, "Finding Site")

1-n

M

BCID 3016 “Major Coronary Artery”

9

CODE

EV (251053005, SCT, "Myocardial perfusion")

1-2

U

BCID 3120 “Perfusion Finding”

10

>

HAS CONCEPT MOD

CODE

EV (363698007, SCT, "Finding Site")

1

M

BCID 3463 “Ventricle Identification”

11

>

HAS PROPERTIES

CODE

EV (116676008, SCT, "Associated Morphology")

1

U

BCID 3121 “Perfusion Morphology”

12

CODE

EV (121071, DCM, "Finding")

1

U

DT (429710001, SCT, "Transient cavitary dilatation")

13

>

INCLUDE

DTID 1350 “Negation Modifier, Presence of Finding”

1

U

14

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = DT (428832008, SCT, "Transient cavitary dilatation ratio")

$Units = EV ({ratio}, UCUM, "ratio")

15

INCLUDE

DTID 300 “Measurement”

1-2

U

$Measurement = EV (10230-1, LN, "LV Ejection Fraction")

$ModType = EV (109054, DCM, "Patient State")

$ModValue = DCID 3102 “Rest/Stress State”

$Units = EV (%, UCUM, "%")


Content Item Descriptions

Row 4

This row may be instantiated twice, once for resting state measurements, once for stress.

Row 15

The LVEF code specified in this row is defined in LOINC with method "imaging". LVEF measurement by ultrasound may also be encoded elsewhere in the Content Tree (e.g., in TID 3309 “Stress Echo Measurement Group”) with LOINC code 18043-0, which has method "ultrasound". It is recommended that such findings from the per-phase measurements be summarized here with the generic "LVEF by Imaging" concept code.

DICOM PS3.16 2026b - Content Mapping Resource