| DICOM PS3.16 2026b - Content Mapping Resource |
|---|
Each instance of the Stress Test Measurement Group represents a group of data elements acquired at approximately the same instant, and conventionally rendered as row in a tabular presentation. It is typically generated during the Stress study whenever a time interval elapses (for example, every minute of the phase), when a technician observes data worth capturing, or when measurements exceed a given range.
Table TID 3304. Stress Test Measurement Group
|
UNITS = DCID 3212 “Treadmill Speed” |
||||||||
|
$Measurement = DT (122706, DCM, "Rating of Perceived Exertion") $Method = BCID 3239 “Perceived Exertion Scale” |
||||||||
|
UNITS = DCID 3500 “Pressure Unit” |
||||||||
|
UNITS = DCID 3500 “Pressure Unit” |
||||||||
|
$Measurement = DT (164931005, SCT, "ST Elevation") $TargetSite = DCID 3001 “ECG Lead” |
||||||||
|
$Measurement = DT (429622005, SCT, "ST Depression") $TargetSite = DCID 3001 “ECG Lead” |
||||||||
|
$Measurement = DCID 3228 “ECG Timing Measurement” $TargetSite = DCID 3001 “ECG Lead” |
||||||||
|
$Measurement = DCID 3227 “QTc Measurement” $TargetSite = DCID 3001 “ECG Lead” $Equation = DCID 3678 “QT Correction Algorithm” |
||||||||
|
$Measurement = DCID 3229 “ECG Axis Measurement” |
||||||||
|
$Measurement = DCID 3526 “Blood Gas Saturation” |
||||||||
|
The Container shall have a specific Content Item Observation DateTime (0040,A032) Attribute to indicate the time at which the measurements were made. |
|
|
Included TID 3106 “Drugs/Contrast Administered” allows the recording of test medications other than the Pharmacological Stress Agent identified in TID 3301 “Stress Test Procedure Description”. |
|
|
ECG ST elevation/depression is measured in units of mV, but is conventionally reported in units of mm, based on strip recordings with scaling of 100 uV/mm. The display application should render these measurements in units meaningful to the user. |
|
|
Note that the MDC codes for "per lead" measurements specified in CID 3228 “ECG Timing Measurement” are base codes for post-coordination with lead identifiers conveyed in the Target Site modifier in TID 300 “Measurement”. MDC also defines pre-coordinated codes that include both the measurement and the lead, which may be used in this row. |
|
|
Note that the MDC code for "QTc interval per lead" specified in CID 3227 “QTc Measurement” is a base code for post-coordination with lead identifiers conveyed in the Target Site modifier in TID 300 “Measurement”. MDC also defines pre-coordinated codes that include both the measurement and the lead, which may be used in this row. Note that TID 300 “Measurement” enables the encoding of a non-standard correction algorithm, either as a local code, or as a TEXT Method Citation (see TID 300 “Measurement” Row 12). |
|
|
Recommended range for ECG axis measurements is -90° to +270° |
| DICOM PS3.16 2026b - Content Mapping Resource |
|---|